A 3D Reconstruction Algorithm for Real-time Simultaneous Multi-Source EIT Imaging for Lung Function Monitoring

Tzu-Jen Kao, Bruce Amm, David Isaacson, Jonathan Newell, Gary Saulnier, Jennifer L. Mueller
doi: https://doi.org/10.1101/2020.05.29.124222
Tzu-Jen Kao
1GE Research, Niskayuna NY 12309, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Bruce Amm
1GE Research, Niskayuna NY 12309, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
David Isaacson
2Rensselaer Polytechnic Institute, Troy NY 12180, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Jonathan Newell
2Rensselaer Polytechnic Institute, Troy NY 12180, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Gary Saulnier
3University at Albany, SUNY, Albany NY 12222, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Jennifer L. Mueller
4Colorado State University, Fort Collins CO 80523, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • For correspondence: mueller@math.colostate.edu
  • Abstract
  • Full Text
  • Info/History
  • Metrics
  • Preview PDF
Loading

Abstract

Monitoring regional pulmonary ventilation and pulsatile perfusion changes in a 3D region of interest (ROI) of the lung is a promising application for electrical impedance tomography (EIT). This paper describes a 3D analytical reconstruction algorithm that was embedded in a prototype EIT system to enable a real-time image reconstruction at nearly 20 frames per second for monitoring impedance changes in the chest in real-time. The derivation and results of the 3D analytical forward solution and inverse solution and details of the real-time reconstruction algorithm are given. The algorithm and EIT system are validated with simulated data, in-vitro phantoms, and finally shown to be capable of imaging ventilation and pulsatile perfusion in human subjects. The human subject data was obtained using a high-precision, high-speed and simultaneous multiple current source (SMS-EIT) developed by GE Research. Data was collected using four rows of 8 electrodes for a healthy adult male subject and 2 rows of 16 electrodes for six healthy human female subjects, with one row placed above the breasts and a second row placed at the infra-mammary fold. Each of the female subjects performed a breathing maneuver with a volumetric incentive spirometer, and the volume of air inhaled was calculated from the EIT images. Pulsatile perfusion images were computed from this data, and regional lung filling was also analyzed.

1 Introduction

Electrical Impedance Tomography (EIT) is an emerging medical imaging technology that estimates the internal electrical properties of the region of interest (ROI) based upon voltage measurements arising from the application of low-amplitude, low-frequency AC current applied to electrodes that are placed on the surface [Barber 1984, Newell 1988, Brown 2003, Holder 2005]. The most developed and widespread use of EIT is for pulmonary imaging, for which the literature is extensive, and we refer to the review articles [Frerichs 2017, Martins 2019, Nguyen 2012, Frerichs 2000] for more information. EIT is especially suitable for bedside pulmonary monitoring or for patients with chronic respiratory disease because it is noninvasive, non-ionizing, and an image can be produced in real time and obtained simultaneously with pulmonary function tests [Muller 2018].

The EIT reconstruction problem is to solve for the electrical conductivity and permittivity distributions inside the body based on the measured voltages and the known excitation currents, in 2D or 3D. It has long been recognized that 3D reconstruction algorithms increase accuracy and decrease artefacts caused by out-of-plane objects [Blue 2000, Eyuboglu 1989, Guardo 1991, Rabbani 1991] and are necessary to image the full lung and to obtain accurate lung volume estimates. In this work, we report the derivation of a real-time 3D reconstruction algorithm based on the ToDLeR (Three-Dimensional Linearized Reconstruction) algorithm [Blue 1997, Blue 2000], which was a simplification and a faster version of the inversion algorithm of Goble [Goble 1990]. The algorithm is implemented in the GENESIS SMS-EIT system, and the reconstructed conductivity images are produced at 20 frames per second on a 32-electrode configuration. Preliminary studies of healthy human subjects with this system have been reported [Amm 2014, Kao 2014]. In this paper we present the details of the solutions and the gradient computations that facilitate the real-time implementation. We further demonstrate the capabilities of the system and algorithm for estimating lung volumes on female subjects. A multiple electrode row 3-D reconstruction algorithm for lung imaging must be effective on both male and female subjects, and this small study demonstrates that lung volumes can be estimated from the images, regional lung filling can be seen, and images of pulsatile pulmonary perfusion can be computed from the data. This is not intended to be a comprehensive validation study.

Lung volume estimates and regional ventilation maps are important clinical measures that can be obtained from electrical impedance tomography images. Clinical validation against CT images has shown that EIT is effective for obtaining images of regional ventilation distribution [Frerichs 2002a, Frerichs 2002, Costa 2009, Smit 2004, Victorino 2004]. However, most EIT systems use a single row of electrodes placed around the chest, and very few studies have been performed on female subjects. In this work, data was collected on six healthy human female subjects with one row of 16 electrodes placed above the breasts and a second row of 16 electrodes placed at the infra-mammary fold. A breathing maneuver was performed with a volumetric incentive spirometer, and 3-D images were computed using the real-time 3D algorithm described in Section II.B. The volume of air inspired was then calculated from the 3-D EIT images using the method derived in [Muller 2015]. The results are compared to the inspiration volumes recorded by the volumetric incentive spirometer during the maneuver. Regional lung filling is also analyzed.

The paper is organized as follows. Section II.A. describes the forward problem and the analytic solution to the homogenous case. In Section II.B. we report the detailed derivation of the solution to the inverse problem based on the ToDLeR algorithm for real-time implementation. The validation on simulated data is in Section III.A. and on phantom data in Section III.B. Results on human subjects are in Section IV, with the study of lung volume estimation on the female subjects in Section IV.B.

2 Methods

The reconstruction algorithm is based on a perturbation approximation. It assumes that the conductivity change distribution differs only slightly from a uniform conductivity distribution. To reconstruct the conductivity, a forward solution and the Jacobian of the forward model with respect to a small perturbation in conductivity will be needed.

1.1 The Forward Model for EIT

The electromagnetic field introduced by the applied current on the surface is governed by Maxwell’s equations. At low current and frequency, it can be simplified to the generalized Laplace equation: Embedded Image where Ω is a bounded region with boundary ∂Ω, Embedded Image is the conductivity distribution, u is the electrical potential, Embedded Image is the externally applied current density, and Embedded Image is the outward-pointing normal vector to the boundary. The forward problem in EIT is to solve for the resulting voltages on the surface of a body from the given current applied on the surface and the internal conductivity distribution.

In the homogeneous case, the first equation in (1) can be reduced to a second order partial differential equation (PDE) Embedded Image

A variety of methods can be used to solve the PDE in (2), such as finite element method (FEM), which can be used to solve for complex geometries. In our targeted application of lung function monitoring, we simplified the geometry of a human torso as a cylinder, with multiple rings of electrodes placed on the side wall. With this simplification, (2) can be solved analytically using separation of variables. We include the derivation here since explicit form of the Fourier coefficients is relevant to our method. See also [Pidcock 1995, Evans 2010] for analytic solutions in this geometry.

In a cylindrical coordinate system, with subscript notation, the boundary condition in (1) can be rewritten as Embedded Image where h is the height of the cylinder and r0 is the radius of the cylinder, and uz = ∂u/∂z. The solution to (2) with the first two boundary conditions in (3), is then Embedded Image where In is the modified Bessel function of order n: Embedded Image

The Fourier coefficients an, bn, Anm, and Bnm are to be determined by using the third boundary condition in (3). In this study, we applied the ‘ave-gap model’ as our electrode model. In the ‘ave-gap model’, we assume that the current density is evenly distributed over the surface of each electrode and that the potential of each electrode is the average value of the potential over the area covered by the electrode. This electrode model does not take into account the shunting effect and the contact impedance between the electrodes and object [Cheng 1989]. By using this electrode model, the current density on the electrode can be approximated as Embedded Image where L is the total number of electrodes; el is the electrode l; Il is the current applied to electrode el; Q is the surface area of each electrode, assuming all electrodes have the same shape, and Embedded Image

In this study, all of the electrodes have the same shape and dimension and are attached to the surface of the body. In the analytical solver, we assume the electrode has a rectangular shape with angular span of Δθ and height of Eh. The electrodes can all be placed in one layer or be split into multiple layers depending on the tradeoff between the area to be covered and the in-plane spatial resolution to be achieved. The area covered by the electrode l is Embedded Image where (r0, θl, zl) are the center of electrode l.

Let Ea denote the area of an electrode. Substituting the current density (6) and potential (4) into the third boundary condition in (3), the Fourier coefficients an, bn, Anm, and Bnm can be computed as Embedded Image where In′ is the first derivative of the modified Bessel function In(x).

2.2 The 3D Reconstruction Algorithm

The reconstruction algorithm used in this study is based on a perturbation approximation which assumes that the conductivity distribution for which we are trying to solve differs only slightly from a uniform conductivity distribution by a perturbation η. Embedded Image

For a uniform conductivity, (1) becomes Embedded Image where k1 is the applied current pattern.

For the conductivity distribution σ in (10) and applied current pattern indexed by k2, linearizing the solution Embedded Image to (1) about u0 yields Embedded Image where Embedded Image is the potential due to σ0, and Embedded Image is O(η). Then equation (1) becomes Embedded Image

Combining (11) and (12), applying the divergence theorem, and neglecting the second and higher order terms, results in Embedded Image where Embedded Image is the measured voltage on electrode l with conductivity distribution σ and current pattern k2, Embedded Image is the predicted voltage on electrode l with uniform conductivity σ0 and current pattern k1. K is the total number of current patterns used. Denoting the area of an electrode by Q, where all electrodes are assumed to be uniform in size, using the ‘Avegap model’ [Cheng, 1989], we can make the approximations Embedded Image

Discretizing the volume, (13) can be written as Embedded Image where i is the voxel index of the discretized volume, N is the total number of voxels, ηi is the perturbation in the i-th voxel, and Embedded Image

Thus, (15) can be written in a vector-matrix form Embedded Image Embedded Image

Here d is the data term, and the K2 by N matrix A is the Jacobian of the forward model with respect to a small perturbation in conductivity. The data term d can be calculated from the known current pattern, the measured voltage and the predicted voltage with a uniform conductivity, which is known or can be estimated as described in section 3. In time difference imaging, voltage taken at reference time t0 will be used as Embedded Image.

As shown in (17), if A were non-singular, one could simply invert it and multiply by the data term to obtain η. Unfortunately, A is ill conditioned, and so it is necessary to regularize it to improve its conditioning in order to obtain a stable solution. The regularization of A will be described in section III. In this section the focus is to describe in detail how to compute the matrix A analytically.

The gradient of u can be computed from (4) and (9) as Embedded Image where Embedded Image

Adopting the notation Embedded Image each entry of the matrix A is given by Embedded Image

Solving the integrals results in Embedded Image where Embedded Image Embedded Image where S is the number of Fourier terms used in the approximation. The calculation of FR121, FR122, FR131, FR132, FR14 and FR3 must be done numerically using numerical quadrature. The others can all be solved analytically.

3 Reconstruction

The SMS-EIT GENESIS system is a real-time, 32-channel, 3D EIT system capable of running up to 20 frames per second at 10 kHz [Ashe 2014]. 31 trigonometric current patterns are applied simultaneously to all of the channels and the resulting voltages are measured on the electrodes. In the following experimental study, 32 electrodes were placed in 4 circular equally spaced rings. The current patterns and electrode configuration are presented in Figure 1 and Figure 2.

Figure 1:
  • Download figure
  • Open in new tab
Figure 1:

The 31 Applied Trigonometric Current Patterns.

Figure 2:
  • Download figure
  • Open in new tab
Figure 2:

Electrode configurations.

A time difference imaging method is implemented in real-time on the prototype system. The voltage signal collected at a reference time t0 is used as a reference, which represents Embedded Image in (18) to calculate the data term d. The reconstructed image is the change in conductivity and permittivity compared to the reference time t0. The time difference imaging approach serves to reduce error caused by mismatch between geometry, electrode shape, location, and contact impedance in the model.

Note that a best fit constant-conductivity approximation to the measured data can be computed from Embedded Image where Embedded Image is the measured voltage signal at t0 and Embedded Image is the predicted voltage with a uniform conductivity of 1. One could then approximate the value of the conductivity by Embedded Image

For a system with L electrodes and L-1 current patterns, there are L(L-1)/2 independent measurements, since the current-to-voltage operator is self-adjoint. In this work, the reconstructions are computed on the Joshua tree mesh [Cheney 1990], as shown in Figure 3, with 496 voxels, matching the number of degrees of freedom for 32 electrodes and 31 current patterns. Since the 32 electrodes were placed in 4 equally spaced rings, we used a 4-layer Joshua tree mesh with 124 voxels in each layer. The height of each layer is determined by the ring-to-ring distance between electrodes and the distance from the bottom / top of the cylinder to its closest ring of electrodes. For 2 rings of electrodes with 16 electrodes in each ring, a 2-layer Joshua tree mesh with 248 voxels in each layer is used.

Figure 3.
  • Download figure
  • Open in new tab
Figure 3.

2D Joshua tree mesh with 496 voxels

The coefficient matrix A is pre-computed and saved to facilitate the real-time reconstruction. The matrix A is regularized using the same approach as in the NOSER algorithm [Cheney 1990]. That is, we solve the regularized system Embedded Image where ε is the regularization parameter, and R is a diagonal matrix with the diagonal terms of ATA.

The implementation of the reconstruction algorithm can be summarized as follows.

  1. Pre-compute matrix A by equation (21);

  2. Apply currents (total number of current patterns K = 31), and measure the resulting voltage on electrodes (total number of electrodes L = 32) at reference time t0 and at time t;

  3. For each set of measured voltages, subtract a constant to ensure the voltages over all electrodes sum to 0;

  4. Compute the data term d using (18), taking the measured voltage at reference time t0 as Embedded Image;

  5. Compute η using (25);

4 Validation of the Reconstruction Algorithm

4.1 Validation with Simulated Data

The forward solver from the open-source software package EIDORS [23] was used to generate data simulating a cylinder containing saline with conductivity 150 mS/m for the reference data and the cylinder filled with saline plus a spherical target placed near the third ring of electrodes from the top at the 7-8 o’clock position. The conductivity ratio between the spherical target and saline is 10:1. See Figure 4 for the target location and cylinder geometry.

Figure 4:
  • Download figure
  • Open in new tab
Figure 4:

Simulation model used in EIDORS and the reconstruction image (Left: simulation top view, Middle: simulation side view, Right: 3D reconstructed images presented in 2D slides.)

At right in Figure 4 is the reconstructed time difference image at each layer in the mesh. Conductivity values higher than that of the reference image are displayed in red and lower values are displayed in blue. Conductivity units are arbitrary but consistent for a given electrode configuration and reconstruction. The image indicates a higher conductivity target at the 7-8 o’clock position in the 3rd layer from the top, which is consistent with the ground truth in both relative conductivity and position.

4.2 Validation with in-vitro phantom data

To validate the reconstruction algorithm with in-vitro phantom data, two sets of data were collected with the SMS-EIT system. Reference voltage data (a homogenous dataset) was collected on a cylinder tank filled with saline (150 mS/m). The inhomogenous dataset was collected after placing a small cylinder target in the saline tank, between the third and fourth ring of electrodes from the top, at the 2 o’clock position as shown in the left and middle figure of Figure 5. The 3D reconstructed conductivity image is displayed in 2D slices with 4-layer Joshua tree mesh which representing the electrodes layers. The reconstruction image indicates a higher conductivity target in the 2 o’clock position between the 3rd and 4th layer from the top.

Figure 5:
  • Download figure
  • Open in new tab
Figure 5:

Phantom validation with a saline tank and a cylinder stainless steel target. (Left: side view, Middle: Top view, Right: 3D reconstructed images presented in 2D slides.)

4.3 Validation with Human Subject Data

The SMS-EIT system was validated under IRB approval to collect data and analyze system performance for healthy human subjects at GE Research. The system passed safety test to meet for IEC60601-1 safety requirement for medical device.

Commercially available ECG electrodes (Intelesens, Belfast, Ireland) with 3 cm conductive diameter were arranged in 4 layers and attached to the male subject’s torso as shown in Figure 6. The center-to-center distance between two neighboring electrode layers is 6 cm. The top ring was placed at the 3rd intercostal space and the bottom ring was placed at the bottom of the rib cage. A safety reference ground electrode was placed on the subject’s shoulder and connected to the system’s safety circuit to detect any unexpected leakage current. The system will switch off automatically if the leakage current is greater than the threshold of 0.450 mARMS. A pulse oximeter, based on the photoplethysmograph, was placed on the subject’s index finger on the left hand, and data was collected while the subject breathed through a mask connected to a pneumotachograph to verify the impedance result of respiratory activity. The real-time 3D re-constructed image is shown in Figure 7, and the update speed is about 20 frame/second when displayed in 2D slices format.

Figure 6:
  • Download figure
  • Open in new tab
Figure 6:

Experimental setup for human subject study.

Figure 7:
  • Download figure
  • Open in new tab
Figure 7:

The real-time reconstructed image in a volunteer healthy subject.

The subject performed 1 minute of normal breathing, 1 minute of deep breathing, a 30 second breath hold, and 30 seconds of normal breathing in one 3-minute sequence, as shown in Figure 8. The waveform comparison between global conductivity, pneumotachograph, and photoplethysmograph has been previously reported in [Kao 2014].

Figure 8:
  • Download figure
  • Open in new tab
Figure 8:

Detected ventilation signal and pulsatile cardiac activity signal on a healthy human subject study.

5 Lung Volume Estimates on Female Subjects

5.1 Study Design

This study was conducted in accordance with the amended Declaration of Helsinki. Data were collected on 6 female subjects in the EIT lab at Colorado State University, Fort Collins, CO under the approval of the IRB. Informed written parental consent and informed assent were obtained from subjects under age 18, and informed written consent from subjects age 18 and up were obtained prior to participation. Subjects ranged in age from 17 to 49 years, chest perimeters at the intra-mammary fold ranged from 29 to 34 inches, and bra cup sizes ranged from A to D.

5.2 Protocol

A breathing maneuver using a 2500 ml volumetric incentive spirometer (VIS) (Air-life 001904A by CareFusion) was used to record inspiratory volumes using the following protocol:

  1. Subject exhales fully and places mouth tightly on the VIS

  2. Slow inhale to a repeatedly comfortable level, desirably between 750 and 1000 ml

  3. Breath-hold for about 3 s while the blue measurement guide on the VIS drops to the bottom

  4. Inhale another additional 750 – 1000 ml

  5. Breath-hold 3s for about 3 s while the blue measurement guide on the VIS drops to the bottom

  6. Slowly exhale

Subjects were instructed in the procedure and asked to practice several times until they felt comfortable with the procedure. Two rows of 16 Philips 13951C pediatric ECG electrodes were applied around the circumference of the subject’s chest. Electrodes 1 - 16 were placed at the level of the intra-mammary fold, or vertebra T9, and electrodes 17 - 32 were placed above the breasts, at the level of vertebra T5. The circumference of the subject was measured at each row, and the average was used for the circumference of the cylindrical model. Biopac was used for 3-lead EKG measurements during data collection. 3D reconstructions were computed using the algorithm described in Section II.B.

5.3 Computation of lung volume estimates

Key time points in the maneuver were identified using a principal component analysis (PCA) of the data. From the PCA, a reference image was chosen representing a point of maximal exhalation prior to the start of the first inhalation maneuver. Frames corresponding to the end of the first inhalation, start of the second inhalation, and end of the second inhalation were also identified. These frames are denoted by

  • SFI = start of first inhalation

  • EFI = end of first inhalation

  • SSI = start of second inhalation

  • ESI = end of second inhalation.

Since the lungs become more resistive as they fill with air, a voxel was chosen to be in the lung region if it became more resistive from frame SFI to frame EFI or from frame SSI to ESI. Since the subjects in this study had healthy lungs, this criterion is not likely to incorrectly omit lung voxels but may include voxels corresponding to artifacts. From observations of the images, the outer 4 rows of voxels in the lower ring and the outer 6 voxels in the upper ring were omitted from the lung region. For most subjects this segmentation resulted in nearly all voxels in the upper layer except for the outer six voxel rings meeting the criterion to be included in the lung region. However, anatomically, this is reasonable since in most females, a cross-section at the T5 vertebrate level is above the heart but may include the aorta.

Ventilation volumes were computed using measures derived in [Muller 2015]. For each voxel p in the segmented lung region, an approximation to the local volume fraction of air, fa(p,t), is computed at time t. Denoting the reconstructed conductivity in voxel p at time t by σa(p,t), and assuming that the conductivity changes in each voxel in the ventilation image sequence are caused by changes in volume of air in that voxel, the conductivity, σa(p,t), is decomposed as Embedded Image where σaM and σam are the maximum and minimum values, respectively, of the conductivity over the data collection period. From this decomposition, the volume fractions of air can be computed from the conductivities by Embedded Image

Denoting the volume of p by vol(p), the volume of air in voxel p at time t is estimated by Embedded Image

The cross-sectional area of a voxel was computed from the area of a pixel in the Joshua tree mesh, and the height of the voxel was taken to be one half the distance between the upper and lower rows of electrodes. Total lung volumes were then computed by summing over all voxels in the mesh.

5.4 Results

Table 1 contains the inhalation volumes recorded with the VIS. In the subsequent images, we show the results from Subjects 1 and 2 for brevity. The total lung volumes for each frame in the reconstruction computed from the segmented EIT images are plotted in Figures 9 and 10 for subjects 1 and 8, respectively with the key time points SFI, EFI, SSI, and ESI annotated in the figures. Plots for the remaining subjects are found in Figure 11.

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 1:

Inhalation volumes recorded using the volumetric incentive spirometer (VIS). Note that the subject held her breath between each inhalation while the VIS returned to 0 ml. Thus, the second inhalation represents volume inhaled in addition to the initial inhalation.

Figure 9:
  • Download figure
  • Open in new tab
Figure 9:

Lung volume per frame for Subject 1 computed from the segmented EIT images with key time points annotated.

Figure 10:
  • Download figure
  • Open in new tab
Figure 10:

Lung volume per frame for Subject 2 computed from the segmented EIT images with key time points annotated.

Figure 11:
  • Download figure
  • Open in new tab
Figure 11:

Lung volume per frame for subjects 3, 4, 5, and 6.

Figures 12 and 13 show the conductivity change in the segmented lung region from frame SFI to EFI (left), frame SSI to ESI (center), the overall change from frame SFI to ESI (right). Each image is shown on its own scale so that the changes are evident since the overall change has the highest dynamic range. All plots are in DICOM orientation. In Figure 12 one sees that the upper lung filled more than the lower lung during the first inhalation, and the lower lung was filled more significantly during the second inhalation, while in Figure 13, the lower lung fills more than the upper lung. This demonstrates the resolution of the algorithm along the cranio-caudal axis in addition to the in-plane resolution that is evident in the images.

Figure 12:
  • Download figure
  • Open in new tab
Figure 12:

Ventilation difference images for Subject 1. Left: EFI minus SFI, Center: ESI minus SSI, Right: ESI minus SFI. The lower row of images depicts conductivity changes in the voxels centered on the intra-mammary fold, while the upper row depicts changes in the voxels centered on the ring of electrodes placed above the breasts.

Figure 13:
  • Download figure
  • Open in new tab
Figure 13:

Ventilation difference images for Subject 2. Left: EFI minus SFI, Center: ESI minus SSI, Right: ESI minus SFI. The lower row of images depicts conductivity changes in the voxels centered on the intra-mammary fold, while the upper row depicts changes in the voxels centered on the ring of electrodes placed above the breasts.

Reconstructions of pulsatile pulmonary perfusion were computed for the periods of time during which the subject held her breath. Reconstructions from frames 190 to 243 for Subject 1 and frames 34 to 81 for Subject 2 were computed with frames 189 and 33 as a reference, for Subjects 1 and 2, respectively. Figures 14 and 15 show a single frame in the sequence of reconstructed images that captures the systolic heart. In each figure the time trace of the conductivity heart voxel is shown in red, and time traces of the conductivity of a voxel from the left lung and right are shown in blue and green, respectively.

Figure 14:
  • Download figure
  • Open in new tab
Figure 14:

Time difference image for one frame in the pulsatile pulmonary perfusion sequence for Subject 1. The frame is indicated by a red star in the lower plot which depicts the time traces of the reconstructed conductivity in a voxel from the heart region (red line), left lung (blue line), and right lung (green line). Plots are in DICOM orientation.

Figure 15:
  • Download figure
  • Open in new tab
Figure 15:

Time difference image for one frame in the pulsatile pulmonary perfusion sequence for Subject 2. The frame is indicated by a red star in the lower plot which depicts the time traces of the reconstructed conductivity in a voxel from the heart region (red line), left lung (blue line), and right lung (green line). Plots are in DICOM orientation.

6 Conclusion

This paper provides a detailed derivation of the analytical forward solution and the Jacobian matrix for a real-time implementation of a 3-D reconstruction algorithm based on the ToDLeR algorithm. The algorithm is validated with simulated data from EIDORS and in-vitro data collected from a saline tank. Two human subject studies are presented, demonstrating that the prototype SMS-EIT system has the ability to detect pulmonary ventilation and pulsatile cardiac activity signals in real-time.

6 Acknowledgment

We thank Dave Davenport, Jeff Ashe, Jim Sabatini, David Shoudy, Gregory Boverman and Xing Wang from GE Research who helped to build the EIT prototype system.

Footnotes

  • lncs@springer.com

References

  1. [1].↵
    D. C. Barber and B. H. Brown, “Applied potential tomography,” J Phys Eng Sci Instrum. vol. 17, pp. 723–33, 1984.
    OpenUrl
  2. [2].↵
    J. C. Newell, D. G. Gisser, and D. Isaacson, “An electric current tomograph,” IEEE Trans Biomed Eng.vol. 35, no. 10, pp. 828–33, 1988.
    OpenUrlPubMed
  3. [3].↵
    B. H. Brown, “Electrical impedance tomography (EIT): a review,” J. Med. Eng. Technol. vol. 27, pp. 97–108, 2003.
    OpenUrlCrossRefPubMed
  4. [4].↵
    D. S. Holder, “Introduction to biomedical electrical impedance tomography Electrical Impedance To-mography Methods, History and Applications. Bristol,” Institute of Physics, pp. 423–49, 2005.
  5. [5].↵
    I. Frerichs, M. B. Amato, A. H. van Kaam, D. G. Tingay, Z. Zhao, B. Grychtol, M. Bodenstein, H. Gagnon, S. H. Böhm, E. Teschner, O. Stenqvist, T. Mauri, V. Torsani, L. Camporota, A. Schibler, G. K. Wolf, D. Gommers, S. Leonhardt, A. Adler & TREND study group, “Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group,” Thorax, vol. 72, no. 1, pp. 83–93, 2017.
    OpenUrlAbstract/FREE Full Text
  6. [6].↵
    T. de Castro Martins, A. K. Sato, F. S. de Moura, E. D. L. B. de Camargo, O. L. Silva, T. B. R. Santos, Z. Zhao, K. Müeller, M. B. P. Amato, J. L. Mueller, R. G. Lima, and M. de Sales Guerra Tsuzuki. “A review of electrical impedance tomography in lung applications: Theory and algorithms for absolute images,” Annual Reviews in Control, vol. 48, pp. 442–471, May 2019.
    OpenUrl
  7. [7].↵
    D.T. Nguyen, C. Jin, A. Thiagalingam, and A. L. McEwan, “A review on electrical impedance tomography for pulmonary perfusion imaging,” Physiol. Meas. vol. 33: 695. April 2012.
    OpenUrlPubMed
  8. [8].↵
    I. Frerichs, “Electrical impedance tomography (EIT) in applications related to lung and ventilation: a review of experimental and clinical activities,” Physiol Meas. vol. 21, no. 2, pp. R1–21. May 2000
    OpenUrlCrossRefPubMedWeb of Science
  9. [9].↵
    P. A Muller, J. L. Mueller, M. Mellenthin, R. Murthy, M. Capps, B. D. Wagner, M. Alsaker, R. Deterding, S. D. Sagel, and J. Hoppe, “Evaluation of surrogate measures of pulmonary function derived from electrical impedance tomography data in children with cystic fibrosis”, Physiol. Meas., vol. 39, no. 4, pp. 045008, 2018.
    OpenUrl
  10. [10].↵
    M K Pidcock, M. Kuzuglu, and K. Leblebicioglu, “Analytic and semi-analytic solutions in electrical impedance tomography: Three-dimensional problems”, Physiol. Meas., vol. 16, no. 2, pp. 77–90, 1995.
    OpenUrlPubMed
  11. [11].↵
    Evans, L. C., Partial differential equations, American Mathematical Society, 2010.
  12. [12].↵
    R. S. Blue, D. Isaacson, and J. C. Newell, “Real-time three-dimensional electrical impedance imaging,” Physiol. Meas. vol. 21, pp. 1–12, 2000.
    OpenUrlPubMed
  13. [13].↵
    B. M. Eyuboglu, B. H. Brown and D. C. Barber, “Limitations to SV determination from APT images,” Images of the Twenty-First Century. Proceedings of the Annual International Engineering in Medicine and Biology Society, Seattle, WA, USA, 1989, pp. 442–443 vol.2.
  14. [14].↵
    R. Guardo, C. Boulay, B. Murray and M. Bertrand, “An experimental study in electrical impedance tomography using backprojection reconstruction,” in IEEE Transactions on Biomedical Engineering, vol. 38, no. 7, pp. 617–627, July 1991.
    OpenUrlCrossRefPubMedWeb of Science
  15. [15].↵
    K. S. Rabbani and A. M. B. H. Kabir, “Studies on the effect of the third dimension on a two-dimensional electrical impedance tomography system,” Clin. Phys. Physiol. Meas. vol. 12, no. 4, pp. 393. 1991.
    OpenUrlCrossRefPubMedWeb of Science
  16. [16].↵
    R. S. Blue, “Real-time three dimensional electrical impedance tomography,” PhD thesis, Rensselaer Polytechnic Institute, Troy, NY, 1997.
  17. [17].↵
    J. C. Goble, “The three-dimensional inverse problem in electric current computed tomography,” PhD thesis, Rensselaer Polytechnic Institute, Troy, NY, 1990.
  18. [18].↵
    B. Amm, T-J. Kao, X. Wang, G. Boverman, D. Shoudy, J. Sabatini, J. Ashe, J. C. Newell, G. Saulnier, D. Isaacson, D. Davenport, “Real-time 3D electrical impedance imaging for ventilation monitoring of the lung: Pilot study,” Conf Proc IEEE Eng Med Biol Soc. 2014;2014:6064–7. doi: 10.1109/EMBC.2014.6945012, PMID: 25571380
    OpenUrlCrossRefPubMed
  19. [19].↵
    T-J. Kao, B. Amm B. X. Wang, G. Boverman, D. Shoudy, J. Sabatini, J. Ashe, J. C. Newell, G. Saulnier, D. Isaacson, D. Davenport,“Real-time 3D electrical impedance imaging for ventilation and perfusion of the lung in lateral decubitus position,” 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Chicago, IL, 2014, pp. 1135–1138.
  20. [20].↵
    I. Frerichs, J. Hinz, P. Herrmann, G. Weisser, G. Hahn, T. Dudykevych, M. Quintel and G. Hellige, “Detection of local lung air content by electrical impedance tomography compared with electron beam CT,” Journal of Applied Physiology, vol. 93, no. 2, p. 660, 2002.
    OpenUrlCrossRefPubMedWeb of Science
  21. [21].↵
    I. Frerichs, J. Hinz, P. Herrmann, G. Weisser, G. Hahn, M. Quintel and G. Hellige, “Regional lung perfusion as determined by electrical impedance tomography in comparison with electron beam CT imaging,” IEEE Transactions on Medical Imaging, vol. 21, pp. 646–652, 2002.
    OpenUrlCrossRefPubMedWeb of Science
  22. [22].↵
    E. Costa, R. Lima and M. Amato, “Electrical impedance tomography,” Current Opinion in Critical Care, vol. 15, pp. 18–24, 2009.
    OpenUrlCrossRefPubMedWeb of Science
  23. [23].↵
    H. Smit, A. V. Noordegraaf, J. Marcus, A. Boonstra, P. Vries and P. Postmus, “Determinants of pulmonary perfusion measured by electrical impedance tomography,” European journal of applied physiology, vol. 92, no. 1, pp. 45–49, 2004.
    OpenUrlCrossRefPubMedWeb of Science
  24. [24].↵
    J. Victorino, J. Borges, V. Okamoto, G. Matos, M. Tucci, M. Caramez, H. Tanaka, F. Sipmann, D. Santos and e. a. C. Barbas, “Imbalances in regional lung ventilation: a validation study on electrical impedance tomography,” American journal of respiratory and critical care medicine, vol. 169, no. 7, pp. 791–800, 2004.
    OpenUrlCrossRefPubMedWeb of Science
  25. [25].↵
    P. A. Muller, T. Li, D. Isaacson, J. C. Newell, G. J. Saulnier, T-J. Kao, J. Ashe, “Estimating a regional ventilation-perfusion index,” Physiol Meas. vol. 36, no. 6, pp. 1283–95, Jun 2015.
    OpenUrl
  26. [26].↵
    K. Cheng, D. Isaacson, J. C. Newell and D. G. Gisser, “Electrode models for electric current computed tomography,” IEEE Transactions on Biomedical Engineering, vol. 36, no. 9, pp. 918–924, Sept. 1989.
    OpenUrlCrossRefPubMed
  27. [27].↵
    J. Ashe, D. Shoudy, G. Boverman, J. Sabatini, T-J. Kao, B. Amm, “A high precision parallel current drive experimental EIT system,” 15th International Conference on Biomedical Applications of Electrical Impedance Tomography, EIT 2014 Proceedings, Gananoque, Ontario, Canada, 24–26, April 2014.
  28. [28].↵
    M. Cheney, D. Isaacson, J. Newell, S. Simske, J. Goble, J. C. Newell, J. C. Goble, “NOSER: an algorithm for solving the inverse conductivity problem,” Int. J. Imaging Syst. Technol. vol. 2, pp. 66–75, 1990.
    OpenUrl
  29. [29].
    A. Adler and W. R. B. Lionheart, “Uses and abuses of EIDORS: An extensible software base for EIT,” Physiol. Meas. vol. 27, pp. S25–S42, 2006.
    OpenUrlCrossRefPubMedWeb of Science
Back to top
PreviousNext
Posted May 31, 2020.
Download PDF
Email

Thank you for your interest in spreading the word about bioRxiv.

NOTE: Your email address is requested solely to identify you as the sender of this article.

Enter multiple addresses on separate lines or separate them with commas.
A 3D Reconstruction Algorithm for Real-time Simultaneous Multi-Source EIT Imaging for Lung Function Monitoring
(Your Name) has forwarded a page to you from bioRxiv
(Your Name) thought you would like to see this page from the bioRxiv website.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Share
A 3D Reconstruction Algorithm for Real-time Simultaneous Multi-Source EIT Imaging for Lung Function Monitoring
Tzu-Jen Kao, Bruce Amm, David Isaacson, Jonathan Newell, Gary Saulnier, Jennifer L. Mueller
bioRxiv 2020.05.29.124222; doi: https://doi.org/10.1101/2020.05.29.124222
Digg logo Reddit logo Twitter logo Facebook logo Google logo LinkedIn logo Mendeley logo
Citation Tools
A 3D Reconstruction Algorithm for Real-time Simultaneous Multi-Source EIT Imaging for Lung Function Monitoring
Tzu-Jen Kao, Bruce Amm, David Isaacson, Jonathan Newell, Gary Saulnier, Jennifer L. Mueller
bioRxiv 2020.05.29.124222; doi: https://doi.org/10.1101/2020.05.29.124222

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
  • Tweet Widget
  • Facebook Like
  • Google Plus One

Subject Area

  • Bioengineering
Subject Areas
All Articles
  • Animal Behavior and Cognition (3703)
  • Biochemistry (7828)
  • Bioengineering (5705)
  • Bioinformatics (21362)
  • Biophysics (10611)
  • Cancer Biology (8217)
  • Cell Biology (11982)
  • Clinical Trials (138)
  • Developmental Biology (6792)
  • Ecology (10430)
  • Epidemiology (2065)
  • Evolutionary Biology (13916)
  • Genetics (9734)
  • Genomics (13116)
  • Immunology (8181)
  • Microbiology (20081)
  • Molecular Biology (7882)
  • Neuroscience (43201)
  • Paleontology (321)
  • Pathology (1285)
  • Pharmacology and Toxicology (2270)
  • Physiology (3367)
  • Plant Biology (7262)
  • Scientific Communication and Education (1317)
  • Synthetic Biology (2012)
  • Systems Biology (5551)
  • Zoology (1135)