ABSTRACT
To battle the COVID-19 pandemic, widespread testing for the presence of the SARS-CoV-2 virus is worldwide being employed by specific real-time RT-PCR (rRT-PCR) of viral RNA. The CDC has issued a recommended panel of PCR-based test sets that entail several primer/probe sets that target the SARS-CoV-2 N-gene, but also one that targets the human RNase P gene (h-RP) as a positive control for RNA extraction and/or reverse-transcription (RT) efficacy.
We discovered that the CDC-recommended h-RP primer/probe set has a faulty design, because both PCR primers are located in the same exon, which allows for unwanted PCR-amplification of background genomic DNA (gDNA). By removing RNA from nose-swab samples by an RNase treatment, we showed that the presence of gDNA in samples resulted in false-positive signals for the h-RP test control. This is rather serious, because it could lead to false-negative test outcomes, since the CDC interpretation of an absent SARS-CoV-2 rRT-PCR signal plus a positive h-RP rRT-PCR signal is interpreted as “2019-nCoV not detected”, whereas a false-positive h-RP rRT-PCR signal resulting from amplification of gDNA should be interpreted as “Invalid Result” and the procedure should be repeated.
In order to overhaul the faulty h-RP rRT-PCR primer/probe set with minimal modification, we designed and tested several new h-RP reverse primers. Replacement of the CDC-recommended PCR reverse primer with our selected exon-exon junction reverse primer corrected the problem of false-positive results with this important SARS-CoV-2 RT-PCR test control and thus eliminated the problem of potential false-negative COVID-19 diagnoses.
Competing Interest Statement
The authors have declared no competing interest.
Footnotes
Relevant study discovered during submission, small paragraph concerning this study was added in the concluding remarks and the related reference added to the references.