Abstract
The SARS-CoV-2 pandemic urges for cheap, reliable and rapid technologies for disinfection and decontamination. We here evaluated the efficiency of UV-C irradiation to inactivate surface dried SARS-CoV-2. Drying for two hours did not have a major impact on the infectivity of SARS-CoV-2, indicating that exhaled virus in droplets or aerosols stays infectious on surfaces at least for a certain amount of time. Strikingly, short exposure of high titer surface dried virus (3*10^6 IU/ml) with UV-C light (0.66 mJ/cm2) resulted in a total reduction of SARS-CoV-2 infectivity. Together, our results demonstrate that SARS-CoV-2 is rapidly inactivated by relatively low doses of UV-C irradiation. Hence, UV-C treatment is an effective non-chemical possibility to decontaminate surfaces from high-titer infectious SARS-CoV-2.
Introduction
SARS-CoV-2 has spread globally and there is an urgent need for rapid, highly efficient, environmentally friendly, and non-chemical disinfection procedures. Application of UV-C light is an established technology for decontamination of surfaces and aerosols (1–3). This procedure has proven effective to inactivate SARS-CoV-1 (4–6), several other enveloped and non-enveloped viruses as well as bacteria (7). Recently, it has also been shown that SARS-CoV-2 is sensitive to inactivation by UV-C irradiation (8–10). However, doses and exposure times necessary for total inactivation of SARS-CoV-2 were in a range precluding efficient application of UV-based methods to be employed for large-scale decontamination of surfaces and aerosols (10). We hence conducted a “real-life” application approach simulating the inactivation of dried surface residing infectious SARS-CoV-2 by a mobile handheld UV-C emitting device and an UV-C box designed to decontaminate medium-size objects. Our data shows that surface dried SARS-CoV-2 retains infectivity for at least two hours. Short exposure of high-titer surface dried SARS-CoV-2 to UV-C light lead to a total reduction of infectivity. Hence, UV-C irradiation is a rapid and cost-effective technology to decontaminate surfaces from high-titer SARS-CoV-2.
Material and Methods
Cell culture
Caco-2 (Human Colorectal adenocarcinoma) cells were cultured at 37 °C with 5% CO2 in DMEM containing 10% FCS, with 2 mM l-glutamine, 100 μg/ml penicillin-streptomycin and 1% NEAA.
Viruses
The recombinant SCoV2 expressing mNeonGreen (icSARS-CoV-2-mNG) (11) was obtained from the World Reference Center for Emerging Viruses and Arboviruses (WRCEVA) at the UTMB (University of Texas Medical Branch). To generate icSARS-CoV-2-mNG stocks, 200,000 Caco-2 cells were infected with 50 μl of virus stock in a 6-well plate, the supernatant was harvested 48 hpi, centrifuged, and stored at −80°C.
For MOI determination, a titration using serial dilutions of the virus stock was conducted. The number of infectious virus particles per ml was calculated as the (MOI × cell number)/(infection volume), where MOI = −ln(1 - infection rate).
UV-C light inactivation treatment
35 uL of virus stock, corresponding to ~4*106 infectious units (IU) of icSARS-CoV-2-mNG were spotted (in triplicates) in 6-well plates and dried for two hours at RT. 6-well plates spotted with dried virus were treated with UV-C-light using the Soluva® pro UV Disinfection Chamber (Heraeus) for 60 seconds or the Soluva® pro UV Disinfection Handheld (Heraeus) for 2 seconds in a fix regime at 5 and 20 cm plate distance. In addition, a moving regime using a slow (3.75 cm/s) and fast (12 cm/s) speed at 20 cm distance was tested. As control, 6-well plates were spotted with the virus and dried, but not UV-treated. After UV-treatment, the spotted virus was reconstituted using 1 mL of infection media (culture media with 5% FCS). As control, 35 uL of the original virus stock were diluted to 1 ml with infection media and used as virus stock infection control.
Evaluation of UV-treatment
For infection experiments, 1 ×104 Caco-2 cells/well were seeded in 96-well plates the day before infection. Cells were incubated with the SARS-CoV-2 strain icSARS-CoV-2-mNG at a MOI=1.1 (stock) or the UV-treated and reconstituted virus in serial two-fold dilutions from 1:200 up to 1:51200. 48 hpi cells were fixed with 2% PFA and stained with Hoechst33342 (1 μg/mL final concentration) for 10 minutes at 37°C. The staining solution was removed and exchanged for PBS. For quantification of infection rates, images were taken with the Cytation3 (Biotek) and Hoechst+ and mNG+ cells were automatically counted by the Gen5 Software (Biotek).
Viral titers (number of infectious virus particles per ml) were calculated as the (MOI × cell number)/(infection volume), where MOI = −ln(1 - infection rate). Infection rates lower than 0.01 were used as a cutoff and set to 0 in order to avoid false positive calculations.
Software and statistical analysis
GraphPad Prism 8.0 was used for statistical analyses and to generate graphs. Figures were generated with CorelDrawX7. Other software used included Gen5 v.3.10.
Results
We set up an experimental approach to evaluate the effect of UV-C treatment on the stability of SARS-CoV-2. Simulating the situation that exhaled droplets or aerosols from infected individuals contaminate surfaces, we produced a high-titer SARS-CoV-2 infectious stock and dried 35μL of this stock corresponding to ~4*10^6 IU/ml in each well of a 6-well plate. The plates were then either non-treated or exposed to five UV-C regimens (Fig. 1a). These include inactivation for 60 s in a box designed to disinfect medium-size objects, 2 s exposure at 5 cm or 20 cm distance with a handheld UV-C disinfection device and finally an approach simulating decontamination of surfaces via the handheld UV-C device. For this, we performed slow and fast-moving at a distance of ~20 cm, with “slow” corresponding to a speed of ~3.75 cm/s (supplemental movie 1) and “fast” at ~12 cm/s (supplemental movie 2). UV-C irradiance (254 nm) in the box with an exposure time of 60 seconds corresponds to an irradiation dose of 800 mJ/cm2; for the handheld (HH) at 5 cm the UV-C dose at two second irradiation time is 80 mJ/cm2 and at 20 cm is 16 mJ/cm2. From the speed of the “slow” and “fast” moving regimens we calculate a UV-C dose of 2.13 mJ/cm2 (slow) and 0.66 mJ/cm2 (fast).
Subsequently, dried virus was reconstituted with 1 mL infection media and used to inoculate naïve Caco-2 cells at serial dilutions to calculate viral titers. Taking advantage of an infectious SARS-CoV-2 strain expressing the chromophore mNeonGreen (11), we quantified infected (mNG+) and total (Hoechst+) cells by single-cell counting with an imaging multiplate reader. Of note, even short UV-C treatment of the dried virus in the context of the moving “fast” regimen completely inactivated SARS-CoV-2, as no infected cells were detected based on fluorescence protein expression (Fig. 1b). Titration of two-fold series dilutions of the UV-treated and non-treated control samples, as well as the freshly thawed strain as reference, revealed that (i) drying for two hours does not have a major impact on the infectivity of SARS-CoV-2 and (ii) all five UV-C treatment regimens effectively inactivate SARS-CoV-2 (Fig. 1c). Calculation of viral titers based on the titration of the reconstituted virus stocks revealed a loss of titer due to drying from ~4*10^6 to ~3*10^6 IU/ml and effective 6-log titer reduction of SARS-CoV-2 by all employed UV-C treatment regimens (Fig. 1d). Altogether, our data demonstrate that UV-C regimens that expose high-titer SARS-CoV-2 to doses down to 0.66 mJ/cm2 are sufficient to achieve complete inactivation of the virus.
Discussion
Disinfection of surfaces and aerosols by UV-C irradiation is an established, safe and non-chemical procedure used for the environmental control of pathogens (1–3, 12). UV-C treatment has proven effective against several viruses including SARS-CoV-1 (4–6) and other coronaviruses i.e. Canine coronaviruses (13). Hence, as recently demonstrated by others (8–10) and now confirmed by our study it was expected that SARS-CoV-2 is permissive for inactivation by UV-C treatment. One critical question is the suitability of this technology in a “real-life” setting in which the exposure time of surfaces or aerosols should be kept as short as possible to allow for a realistic application, for example in rooms that need to be used frequently as operating rooms or lecture halls. Furthermore, in such a setting, we assume that the virus is exhaled from an infected person by droplets and aerosols, dries on surfaces and hence represents a threat to non-infected individuals. We simulated such a situation and first evaluated if surface dried SARS-CoV-2 is infectious. Drying for two hours, in agreement with previous work (14), did not result in a significant reduction of viral infectivity indicating smear-infections could indeed play a role in the transmission of SARS-CoV-2 (Fig. 1). On the other hand, our virus-preparations are dried in cell culture pH-buffered medium containing FCS, which might stabilize viral particles. Hence, even though this is not the scope of the current study, it will be interesting to evaluate if longer drying or virus-preparations in PBS affect the environmental stability of SARS-CoV-2. Irrespective of the latter, UV-C-exposure of dried high-titer SARS-CoV-2 preparations containing ~3*10^6 IU/ml resulted in a complete reduction of viral infectivity (Fig. 1). In this context, it is noteworthy that we achieved a 6-log virus-titer reduction in a setting simulating surface disinfection with a moving handheld device. With the “fast”-moving protocol (see supplemental video 1) we were exposing surfaces at a distance of 20 cm with a speed of 12.5 cm/s resulting in a calculated UV-C dose of 0.66 mJ/cm2 at 254 nm. This is substantially less than the previously reported 1048 mJ/cm2 necessary to achieve a 6-log reduction in virus titers when exposing aqueous SARS-CoV-2 to UV-C (10). In another study, using a 222 nm UV-LED source, 3 mJ/cm2 lead to a 4-log reduction of infectious SARS-CoV-2 (9) and 20 s deep-ultraviolet treatment at 280 nm corresponding to a dose of 75 mJ/cm2 reduced 3-logs (8).
Comparing these values to other pathogens, SARS-CoV-2 seems particularly sensitive towards UV-C light. To achieve a 3-log titer reduction, 75-130 mJ/cm2 are necessary for adenovirus, 11-28 mJ/cm2 for poliovirus, and bacteria as for instance Bacillus subtilis require 18-61 mJ/cm2 (7). This is in-line with susceptibility of SARS-CoV towards UV-C in aerosols at 2.6 mJ/cm2, whereas adenovirus or MS2-bacteriophages were resistant to such a treatment (1).
Altogether, we establish the effectiveness of UV-C treatment against SARS-CoV-2 in a setting designed to simulate realistic conditions of decontamination. The easy, rapid, chemical-free, and high efficacy of UV-C treatment to inactivate SARS-CoV-2 demonstrates the applicability of this technology in a broad range of possible settings.
Author contributions
NR and MS designed the experiments; NR performed the experiments with support from RB; NR, RB and MS analyzed the data; NR and MS drafted the figures and wrote the manuscript; MS developed the manuscript to its final form; MS planned and supervised the study; all authors read, edited, and approved the final manuscript.
Conflict of interest
The authors declare no conflict of interest
Funding
This work was supported by grants to MS from the Baden-Württemberg foundation (BW-Stiftung), the Deutsche Forschungsgemeinschaft, the MWK Baden-Württemberg as well as by basic funding provided to MS by the University Hospital Tübingen. Heraeus provided the Soluva UV-disinfection chamber and the UV-handheld disinfection device and contributed funding for consumables. The funders had no role in study design, data analysis or decision to publish the data.
Supplemental Movie 1. UV-irradiation using the Handheld device, slow-moving regime. SARS-CoV-2-mNG was spotted in a 6-well plate, dried for two hs and UV-irradiated as shown in the video. Speed is calculated at approx. 3.75 cm/s.
Supplemental Movie 2. UV-irradiation using the Handheld device, fast-moving regime. SARS-CoV-2-mNG was spotted in a 6-well plate, dried for two hs and UV-irradiated as shown in the video. Speed is calculated at approx. 12.5 cm/s.