Hydrogel-coating improves the in-vivo stability of electrochemical aptamer-based biosensors

Shaoguang Li, Jun Dai, Man Zhu, Netzahualcóyotl Arroyo-Currás, Hongxing Li, Yuanyuan Wang, Quan Wang, Xiaoding Lou, Tod E. Kippin, Shixuan Wang, Kevin W. Plaxco, Hui Li, Fan Xia
doi: https://doi.org/10.1101/2020.11.15.383992
Shaoguang Li
aEngineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Jun Dai
bDepartment of Obstetrics and Gynecology, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Man Zhu
aEngineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Netzahualcóyotl Arroyo-Currás
cDepartment of Pharmacology and Molecular Sciences, Johns Hopkins University School of Medicine, Baltimore, Maryland 21205, United States
dDepartment of Chemical and Biomolecular Engineering, Whiting School of Engineering, Johns Hopkins University, Baltimore, Maryland 21218, United States
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Hongxing Li
aEngineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Yuanyuan Wang
aEngineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Quan Wang
aEngineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Xiaoding Lou
aEngineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Tod E. Kippin
dDepartment of Chemical and Biomolecular Engineering, Whiting School of Engineering, Johns Hopkins University, Baltimore, Maryland 21218, United States
eDepartment of Psychological and Brain Sciences, University of California, Santa Barbara, CA 93106
fThe Neuroscience Research Institute, University of California, Santa Barbara, CA 93106
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Shixuan Wang
bDepartment of Obstetrics and Gynecology, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Kevin W. Plaxco
gDepartment of Molecular Cellular and Developmental Biology, University of California, Santa Barbara, CA 93106
hDepartment of Chemistry and Biochemistry, University of California, Santa Barbara, CA 93106
iCenter for Bioengineering, University of California, Santa Barbara, CA 93106
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • For correspondence: kwp@chem.ucsb.edu
Hui Li
aEngineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Fan Xia
aEngineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Abstract
  • Full Text
  • Info/History
  • Metrics
  • Supplementary material
  • Preview PDF
Loading

Abstract

The ability to track the levels of specific molecules, such as drugs, metabolites, and biomarkers, in the living body, in real time and for long durations would improve our understanding of health and our ability to diagnose, treat and monitor disease. To this end, we are developing electrochemical aptamer-based (E-AB) biosensors, a general platform supporting high-frequency, real-time molecular measurements in the living body. Here we report that the addition of an agarose hydrogel protective layer to E-AB sensors significantly improves their baseline stability when deployed in the complex, highly time-varying environments found in vivo. The improved stability is sufficient that these hydrogel-protected sensors achieved good baseline stability when deployed in situ in the veins, muscles, bladder, or tumors of living rats without the use of the drift correction approaches traditionally required in such placements. Finally, this improved stability is achieved without any significant, associated “costs” in terms of detection limits, response times, or biocompatibility.

Introduction

The ability to track the levels of specific molecules, such as drugs, metabolites, or biomarkers continuously and in real time in the living body would vastly improve our knowledge of physiology, pharmacokinetics, and toxicology and would pave the way for truly high-precision personalized medicine. Such a technology, for example, would improve our understanding of many time-dependent physiological events, including the distribution phases of therapeutic drugs, the pulsatile release of hormones, and the hemostatic control of key metabolites.[1–3] In the clinic such a technology could likewise provide the high-precision, patient-specific pharmacokinetics required to provide high precision dosing, and even support real-time, feedback-controlled drug delivery of unprecedented accuracy.[4–5] The development of such a technology, however, faces significant hurdles.[6–7] Specifically, such a technology must: (1) achieve clinically-relevant sensitivity and specificity; (2) must be reversible, so that it can follow rising and falling concentrations; (3) it must operate continuously, or at least at a frequency that is high relative to physiological timescales (it thus cannot rely on batch processing, such as separations or the addition of exogenous reagents); and (4) it must remain stable in the complex, fluctuating environments found within the body. Faced with these hurdles, continuous, in-vivo sensing had, until recently only been reported for a short list of metabolites, physiological molecules (e.g., glucose,[8] lactate,[9] and blood oxygen[10]) and neurotransmitters (e.g., dopamine, acetylcholine).[11–12] Moreover, the sensors for each of these targets are critically reliant on the chemical or enzymatic reactivity of their targets, and thus they are not generalizable to the detection of other, arbitrary targets.

Against this background, we[6, 12–14], followed by others[15–17], have developed electrochemical aptamer-based (E-AB) biosensors, the first platform technology supporting high-frequency, in-vivo molecular measurement that does not rely on the intrinsic chemical or enzymatic reactivity of its targets. To achieve this, E-AB sensors employ a target binding-induced conformational change to generate an electrochemical signal (Fig. 1a). Specifically, E-AB sensors are comprised of redox-reporter-modified DNA or RNA aptamer, a class of functional oligonucleotides that binds a specific analyte and can be artificially selected via high-throughput, in-vitro methodologies, that are covalently attached to an interrogating electrode. The binding of an analyte to these recognition elements alters the efficiency with which the redox reporters transfers electrons to or from the electrode, producing an easily measurable signal change when the sensor is interrogated electrochemically.[12] Because this signaling mechanism recapitulates the conformation-linked signaling typically employed by naturally occurring chemo-perception systems, E-AB sensors are selective enough to deploy directly in complex sample matrices.[18–19]

Fig. 1
  • Download figure
  • Open in new tab
Fig. 1

Here we have demonstrated the ability of a protective hydrogel coating layer to improve the baseline stability of electrochemical aptamer-based (E-AB) biosensors when they are deployed in situ in the living body. The signaling mechanism of E-AB sensors, which recapitulates the conformation-linked signaling employed by naturally occurring chemo-perception systems, renders them selective enough to be deployed in complex matrix, such as blood serum.[18–19] (a) However, they suffer from significant signaling baseline drift when they are deployed directly in whole blood for continuous, real-time measurements both in vitro and in vivo without any drift corrections (bottom panels shown are real data collected in vivo). Historically we have corrected this drift using algorithms such as kinetic differential measurements or chronoamperometry.[23–24] (b) Here, in contrast, we demonstrate the good baseline stability of uncorrected, gel-protected sensors (bottom panels shown are real data collected in vivo).

While E-AB sensors are reasonably stable in blood and serum in vitro, they exhibit significant drift when deployed directly in the living body[20–22], which presumably arises due to degradation of the target-recognizing aptamer and the non-specific adsorption of cells and other blood components to the sensor surface. To circumvent this, we have historically employed drift correction methods, such as kinetic differential measurements (KDM)[23], dual-reporter approach[12] or chronoamperometry.[24] The former two employs a secondary signal from the sensor to correct for the drift, while the latter monitors the electron transfer time constant to quantify the target, an observation that is independent of the number of aptamer probes and thus rather insensitive to sensor degradation. Using these approaches, we have achieved good measurement accuracy and returns to baseline for in-vivo runs up to several hours. Ultimately, however, these approaches fail when the sensor degradation becomes so great that its signal-to-noise ratios become unacceptably low. Thus motivated, here we explore a complementary technology: the use of a hydrogel matrix to protect E-AB sensors from non-specific adsorptions and sensor degradations by decreasing the access of high molecular weight, biological components (e.g., proteins, blood cells) to the sensor surface (Fig. 1b), reducing their ability to degrade sensor performance.

Results

Gel-coating improves E-AB sensor performance

The utility of gel coating to solve the baseline drift of E-AB sensors relies on the differential kinetics of molecular diffusion through a dense hydrogel. To characterize this diffusion, we first used both modeling and experimental approaches to estimate the size-dependence of molecular diffusion kinetics through a representative hydrogel. Specifically, using Amsden’s theoretical model, [25], we found that the time required for diffusion through a certain length of hydrogel is greatly dependent on both molecular size (here, we employed a variety of proteins in whole blood) and gel porosity (i.e., gel wt% concentrations, Fig. 2a and 2b, Supplementary Fig. 1). Consistent with this, when we incubated a 5 mg/mL solution of fibrinogen (~25 nm in diameter) and 5 mg/mL solution of BSA (~3 nm in diameter) on top of a gel for 24 h and monitored the two proteins’ diffusion into the gel mass, we found that, at a depth of ~25 μm (the thickness of the gel on our protected sensors), the concentration of the former only reached 10% of the applied concentration (Fig. 2c and 2d). In contrast, the latter, lower-molecular weight protein was homogeneously distributed throughout the entire ~ 400 μm thickness of the gel slice.[26–27] Not surprisingly, the presence of a hydrogel also prevents the agglomeration of blood cells onto a sensor surface (Fig. 2e), as these are far too large to penetrate the gel network.

Fig. 2
  • Download figure
  • Open in new tab
Fig. 2

Hydrogel greatly reduces the diffusion of high-molecular weight proteins and effectively eliminates the diffusion of cells. (a) Using Amsden’s theoretical model, we modeled the extent to which the presence of a hydrogel reduces the diffusion constant of various proteins (Dg/D0 reflects the diffusion constant in the gel relative to that seen in the free solution). As expected, the gel significantly reduces the diffusion of high-molecular weight proteins, while have less of an effect on lower molecular weight proteins. (b) The time duration required for blood components to diffuse through a specific depth (here, ~25 μm) is likewise greatly dependent on their molecular weight and the density of the gel. (c-d) Using confocal microscopy we have monitored the diffusion of BSA (~3 nm in diameter) and fibrinogen (~25 nm) through a gel slice fabricated from 3 wt% gel solution over the course of 24 h. Under these circumstances the concentration of fibrinogen at a depth of 25 μm reaches only ~10% of the applied concentration. In contrast, after the same 24 h incubation the concentration of BSA seen in the gel is homogeneously distributed across its entire 400 μm thickness of the gel. (e) Not surprisingly, while red blood cells adhere to an unprotected electrode (i.e., no gel coating) none are seen on a gel-protected electrode (The red blood cells on the unprotected sensors are outlined with black circles).

Baseline stability in-vitro

Motivated by the ability of a hydrogel to greatly reduce the diffusion of blood cells and higher molecular weight proteins, we next investigated their ability to protect E-AB sensors in vitro in (a) undiluted whole blood and (b) excised, solid tissue, using a sensor against the antibiotic kanamycin[14] as our test bed. Using a standard dip-coating protocol (Supplementary Fig. 2), we first fabricated a gel-protected E-AB sensors and then interrogated these sensors in vitro both in buffer and whole blood using a three-electrode system (Fig. 3a). As expected, we observed similar stability when we challenge unprotected and gel-protected versions of this sensor in phosphate buffered saline (Supplementary Fig. 3). In contrast, when the two are challenged in vitro in undiluted whole blood, the performance of the gel-protected sensor is notably improved over that of unprotected sensors. For example, over 10 h the peak currents of unprotected sensors fall by 70% in undiluted whole blood and over 10 h under these conditions, those of gel-protected sensors fall by less than 5% (Fig. 3b, c, d). Perhaps not coincidentally, after interrogating for 10 h in whole blood, we observed that, while unprotected sensor was covered with biocomponents, gel-protected sensor exhibited no adsorptions to such components (Supplementary Fig. 4). The improved stability of gel-protected sensors likewise holds for sensors inserted into excised solid tissue (Fig. 3e) and for sensors employing aptamers against other than kanamycin (Supplementary Fig. 5).

Fig. 3
  • Download figure
  • Open in new tab
Fig. 3

Gel-protected sensors achieve excellent stability while maintaining rapid response time and low limits of detection. (a) Here we employed a three-electrode system to interrogate our sensor, including a gold working electrode that serves as the sensor, a Ag/AgCl reference electrode, and a platinum counter electrode. The gold working electrode was used as is or protected by gel layer, i.e., unprotected or gel-protected sensors. (b) When deployed in whole blood, unprotected sensors exhibited a significant signal loss shown as the current obtained via square wave voltammetry decaying dramatically. (c) Under the same conditions, the signal obtained from gel-protected sensors is quite stable. (d) For example, while unprotected sensors exhibit 80% signal loss over 10 h in whole blood, gel-protected sensors exhibit no signal loss under the same conditions. After interrogating for 10 h in whole blood, while unprotected sensor was fully covered with biocomponents, gel-protected sensor exhibited no adsorptions to these components (Supplementary Fig. 4). Gel-protected sensors likewise remain far more stable (e) when inserted into a tissue sample (here fresh pork) and (f) in plasma. (g) When deployed in PBS buffer in the presence of a 5 mM solution of DNases-I, unprotected sensors exhibit 20% signal loss over 1.5 h, while for gel-protected sensors the loss is less than 5%. The (h) response time and (i) limits of detection of gel-protected sensors remain quite similar to those of unprotected sensors.

To determine which blood components contribute most significantly to the observed drift we next challenged gel-protected and unprotected sensors for 10 h in either plasma, which is the liquid fraction of whole blood (Fig. 3f), or formed elements (Supplementary Fig. 6), which is the cellular fraction. When incubated for 10 h, gel-protected sensors exhibited excellent stability under both conditions. In contrast, unprotected sensors lost 80% of their signal in formed elements and 20% in plasma, suggesting that the agglomeration of blood’s cellular components on the sensor or the reaction of their components with the DNA aptamer (e.g., DNAses liberated from ruptured cells, Fig. 3g) are a larger source of baseline drift than that associated with the proteins in plasma.

The drift protection provided by the gel coating comes at relatively little cost in sensor equilibration time, detection limits. For example, while the presence of the protecting gel slows sensor response times, this effect is small relative to the timescales of the physiological processes E-AB sensors have been used to investigate. Specifically, a kanamycin-detecting, gel-protected sensor reaches 90% of its maximum signal change within less than one minute (Fig. 3h), a timescale far faster than the tens of minutes elimination rate of this drug.[14] The detection limit of a gel-protected sensor is likewise effectively indistinguishable from that of the equivalent, unprotected sensor (Fig. 3i). Consistent with this, the electron transfer kinetics of gel-protected sensors are closely comparable to those observed for unprotected sensors (Supplementary Fig. 7), suggesting their gain and detection limits should be similar.

Continuous, real-time, in vivo molecular measurements

Motivated by the improved stability of gel-protected sensors exhibit in vitro, we next tested them in a variety of in vivo scenarios. As the first of these tests we emplaced a kanamycin-detecting sensor in the external jugular veins of an anesthetized Sprague-Dawley rats, and injected the drug into the opposite external jugular vein. Under these conditions, an unprotected (non-drift-corrected) sensor exhibits ~40% loss in signal over the course of 2 h (Fig. 4a and 4b). Gel protection reduced this loss to less than 5% over the same period. Following two sequential intravenous boluses of kanamycin, the gel-protected sensor recorded consecutive concentration spikes corresponding to each bolus, with maximum kanamycin concentrations (Cmax) of ~200 μM and the effective clearance of 90% of the drug from the circulatory system within 50 min (Fig. 4c), values that are fully consistent with prior studies of the pharmacokinetics of this drug.[6] Repeating these measurements with the gel-protected sensor in multiple rats we observed reproducibility consistent with the known pharmacokinetic variability of the aminoglycosides[28] (Supplementary Fig. 8-9). Finally, we used gel-protected sensors to follow monotonically increasing intravenous kanamycin doses spanning the 10-30 mg/kg therapeutic ranges used in humans[29] (Fig. 4d). The sensor responded rapidly to each injection, and measured peak drug concentrations in good accordance with the relevant delivered dose (Fig. 4e).

Fig. 4
  • Download figure
  • Open in new tab
Fig. 4

Gel-protected E-AB sensors support continuous, real-time molecular measurements in situ in the living body without the use of the drift-correction algorithms previously used in such deployments.[23] (a) To show this, we first emplaced kanamycin-detecting sensors in the external jugular veins of anesthetized Sprague-Dawley rats, which we then challenged via intravenous injection of the drug into the opposite external jugular vein. (b) Unprotected sensors exhibited significant drift over the course of these few-hour experiments, rendering them incapable of determining the target concentrations without employing drift-correction methods. (c) In contrast, gel-protected sensors are much more stable, achieving precise, continuous molecular measurement. (d) We then followed monotonically increasing intravenous doses of kanamycin spanning the 10-30 mg/kg therapeutic ranges used in humans. (e) The sensor responded rapidly to each injection, measuring maximum concentrations between 34 and 100 μM in good accordance with the delivered dose.

Gel-protected sensors achieve clinically relevant accuracy in vivo without employing the drift correction mechanism we have previously employed. To see this, we performed simultaneous in-vivo and ex-vivo measurements using, respectively, a gel-protected E-AB sensor and blood draws followed by high-performance liquid chromatography (Fig. 5). Specifically, we implanted one kanamycin-detecting, gel-protected sensor in the jugular vein of a living rat and performed the real-time, continuous E-AB measurements while drawing blood samples at 20 to 30 min intervals from the opposite jugular vein. Performing HPLC-ELSD measurements immediately after in-vivo test (to maintain the freshness of the sample) using a standard protocol for blood pre-treatment (detailed protocols see supplementary materials, Supplementary Figs. 10-11), the ex-vivo measured concentrations for each sample were in close accordance to the values obtained from E-AB sensors (Fig. 5b, c). For example, measurements taken by the two approaches during the elimination phase of the drug’s pharmacokinetics were within 30% of one another, a level of accuracy similar to that of commercial glucose sensors.[30–31]

Fig. 5
  • Download figure
  • Open in new tab
Fig. 5

Gel-protected E-AB sensors achieved clinically relevant accuracy without employing drift-correction approaches.[23–24] (a) and (b) To see this, we performed parallel pharmacokinetics studies using ex-vivo analysis via high-performance liquid chromatography as our gold standard. Specifically, used a gel-protected E-AB sensor in the jugular vein of a live rat to perform real-time measurements of kanamycin, while simultaneously collecting blood samples at an interval of 20 to 30 minutes for subsequent bench-top analysis. (c) The results of the two approaches are in good accordance throughout the experiment. The precision of E-AB sensors as defined by the standard deviation observed during the pre-challenge baseline is on the order of ~30 μM. The error bars on the E-AB data are derived from the error in the calibration curves obtained via in-vitro titrations. The error bars on HPLC data were derived from replicate measurements performed on each sample. In both cases the error bars reflect 95% confidence intervals.

Simultaneous, multi-compartment measurements

Gel-protected E-AB sensors also achieve good baseline stability when placed in other bodily compartments, suggesting that they can be used, for example, to monitor time-varying molecular concentrations throughout the body[32–33] To see this, we first implanted three E-AB sensors in a single rat: one in a jugular vein, a second in a leg muscle, and a third in the bladder (Fig. 6a). Upon intravenous challenge with 40 mg/kg kanamycin, we again observed a rapid increase in plasma drug level peaking at ~200 μM. As expected, the sensors placed in the tissue and bladder also measured rising drug levels, albeit with a delay arising due to the slow transport of the drug into the tissues and the slow excretion of the drug via the kidneys (Fig. 6b). Following on this, we also implanted a sensor against the chemotherapeutic doxorubicin into a solid tumor (Fig. 6c, Supplementary materials, video 1). Once again, when we implanted a doxorubicin-detecting gel-protected sensor in BALB/C nude mice, we achieve a micromolar precision upon intravenous challenge with the drug (Fig. 6d) at therapeutically relevant concentrations.

Fig. 6
  • Download figure
  • Open in new tab
Fig. 6

Gel-protected E-AB sensors exhibit good baseline stability when placed in a variety of bodily compartments. (a) To show this we simultaneously implanted kanamycin-detecting sensors in the jugular vein, leg muscle, and bladder of a single rat. (b) Upon intravenous challenge with 40 mg/kg kanamycin we once again observed an immediate rise of drug in the vein, peaking at ~200 μM. The sensors in the muscle and bladder exhibited slower drug level rise, indicating a slower uptake into the solid tissues and slower elimination of the drug from the blood. (c) We have also explored the use of E-AB sensors to perform continuous, real-time molecular tracking in situ in a tumor. (d) Specifically, a gel-protected, doxorubicin-detecting sensor placed inside of a xenograph tumor achieved micromolar precision upon an intravenous challenge of the drug at 40 mg/kg, a dose that is therapeutically relevant.

Biocompatibility

We also performed preliminary exploration of the biocompatibility of both unprotected and gel-protected E-AB sensors, a critical issue for the development of long-duration implantable devices.[7, 34–37] To do so we employed five groups of rats (15 rats per group): those implanted with (1) gel-protected and (2) unprotected sensors in the tail vein; those implanted with (3) gel-protected and (4) unprotected sensors in the muscle of the left hind limb; and (5) a control group that was not implanted with any sensors, but was subjected to the same surgical procedures as a combination of both vein and muscle groups. In each case, the sensors remained in the animal for one week. During this time observed no significant difference among the five groups in terms of body weight, food/water consumption, or behavior (Fig. 7a-c). A week post-implantation we performed blood examination and morphology studies. Upon visual inspection, we observed what appeared to be complete recovery of the implantation site and no inflammation of the skin (Supplementary Figs. 12-17). Consistent with this, we did not observe any significant differences in blood vessel morphology, and only slight differences in muscle tissue between implanted sensors (either protected or unprotected) and controls (Fig. 7d). Likewise, we did not detect any hematological changes in terms of white blood cells, red blood cells, hemoglobin, and platelets counts (Fig. 7f. To test for the presence of thrombosis in intravenous placements,[38] we collected the vessels from the rats one week after tail-vein implantation. In neither case did we observe thrombosis for either gel-protected or unprotected implantations (Fig. 7g, Supplementary Fig. 18). Neither did we detect any significant differences in endothelin, D-dimer and thrombomodulin between the five groups (Fig. 7h), further suggesting that sensor implantation does not lead to thrombosis. Finally, to evaluate the immune response provoked by our sensors (which can be an obstacle to practical implantable devices[39–40]), we monitored the inflammatory biomarkers C-reactive protein, tumor necrosis factor α, interleukin 12 and interleukin 6, observing no statistical difference (p >0.05) between any of the five groups (Fig. 7i).

Fig. 7
  • Download figure
  • Open in new tab
Fig. 7

Both gel-protected and unprotected sensors exhibit good biocompatibility. To determine this we employed five groups of 15 rats each: (1) gel-protected and (2) unprotected sensors (denoted here as “no gel”) implanted in the tail vein; (3) gel-protected and (4) unprotected sensors implanted in the muscle of the left hind limb; and (5) a control group that was not implanted with any sensors, but was subjected to the same surgical procedures (both vein and muscle) as the former groups (Supplementary Figs.12-17). (We observed no significant differences in (a) water consumption, (b) food consumption or (c) animal behavior between implanted and control animals. (d) Likewise, we observed no significant difference in the morphology of the tail vein after one-week implantations. Here, the black arrows denote the tail vein. (e) Implantations of both protected and unprotected in the muscle cause mild inflammation, as indicated by black dotted lines. (f) We observe no significant differences in blood counts between any of these five groups. Likewise, we observed (g) no thrombosis, and (h) no significant change in the coagulation markers endothelin, D-dimer, or thrombomodulin. (i) Finally, we observed no sigificant differences in the plasma levels of the inflammatory factors C-reactive protein, interleukin 12, interleukin 6 and tumor necrosis factor α.

Conclusions

Here we demonstrate that the application of an argarose hydrogel coating significantly reduces drift when E-AB sensors are deployed both in vitro in undiluted whole blood and in situ in the veins, the bladder, solid healthy tissue, or solid neoplastic tissues of live rats. We believe this improved stability is due to the dependent differential kinetics of molecule diffusion across the gel, which allows low-molecular weight target molecules to diffuse rapidly to the sensor surface while largely blocking the approach of high-molecular weight components, minimizing non-specific adsorption to the sensor and enzymatic degradation of its aptamer. Moreover, gel-protected sensors achieve these improvements without significant reductions in time resolution, limits of detection, precision, or accuracy. Finally, we have demonstrated that gel-protected sensors retain the same biocompatibility as unprotected E-AB sensors when implanted in blood vessels or muscle tissue, as neither causes any detectable impact on animal behavior or blood properties, and only minimal changes in tissue morphology.

Author contributions

Dr. S. L. and Dr. H. L. conceived of the project. Dr. S. L., Dr. H. L., Dr. K.W. P., Dr. J. D., Dr. X. L. and Dr. F. X. designed the experiments. M. Z., H. L fabricated the sensors and tested the sensors, conducted the HPLC measurements. Dr. S. L., Dr. J. D., Dr. H. L., Q. W. and Dr. N. A.-C. directed the animal studies, performed the controller simulations and analyzed the data. Dr. H. L., Dr. K.W. P., Dr. S. L. and Dr. J. D. wrote and edited the manuscript. All authors discussed the results and commented on the manuscript.

Additional information

Supplementary information is available for this paper.

Acknowledgements

This work was supported by the National Natural Science Foundation of China (21525523, 21804121, 21874121, 21801231), the Fundamental Research Funds (CUG170665, CUG170668) for the Central Universities, China University of Geosciences (Wuhan) and the US National Institutes of Health (EB022015).

Reference

  1. 1.↵
    Undevia, S. D., Gomez-Abuin, G. & Ratain, M. J. Pharmacokinetic variability of anticancer agents. Nat. Rev. Cancer 5, 447–458 (2005).
    OpenUrlCrossRefPubMedWeb of Science
  2. 2.
    Kanto, J. & Gepts, E. Pharmacokinetic implications for the clinical use of propofol. Clin. Pharmacokinet. 17, 308–326 (1989).
    OpenUrlCrossRefPubMedWeb of Science
  3. 3.↵
    Muller, P. Y. & Milton, M. N. The determination and interpretation of the therapeutic index in drug development. Nat. Rev. Drug Discov. 11, 751–761 (2012).
    OpenUrlCrossRefPubMed
  4. 4.↵
    Hamburg, M. A. & Collins, F. S. The path to personalized medicine. New Engl. J. Med. 363, 301–304 (2010).
    OpenUrlCrossRefPubMedWeb of Science
  5. 5.↵
    Mage, P. L. et al. Closed-loop control of circulating drug levels in live animals. Nat. Biomed. Eng. 1, 0070 (2017).
    OpenUrl
  6. 6.↵
    Arroyo-currás, N. et al. Real-time measurement of small molecules directly in awake, ambulatory animals. Proc. Natl. Acad. Sci. U. S. A. 114, 645–650 (2017).
    OpenUrlAbstract/FREE Full Text
  7. 7.↵
    Xie, X. et al. Reduction of measurement noise in a continuous glucose monitor by coating the sensor with a zwitterionic polymer. Nat. Biomed. Eng. 2, 894–906 (2018).
    OpenUrl
  8. 8.↵
    Hovorka, R. Continuous glucose monitoring and closed-loop systems. Diabet. Med. 23, 1–12 (2006).
    OpenUrlCrossRefPubMedWeb of Science
  9. 9.↵
    Baker, D. A. & Gough, D. A. A Continuous, implantable lactate sensor. Anal. Chem. 67, 1536–1540 (1995).
    OpenUrlCrossRef
  10. 10.↵
    Khan, Y. et al. A flexible organic reflectance oximeter array. PNAS. 115, 11015–11024 (2018).
    OpenUrl
  11. 11.↵
    Liu, X. et al. Ultrathin cell-membrane-mimic phosphorylcholine polymer film coating enables large improvements for In vivo electrochemical detection. Angew. Chem. Int. Edit. 56, 11802–11806 (2017).
    OpenUrl
  12. 12.↵
    Li, H., Arroyo-currás, N., Kang, D., Ricci, F. & Plaxco, K. W. Dual-reporter drift correction to enhance the performance of electrochemical aptamer-based sensors in whole blood. J. Am. Chem. Soc. 138, 15809–15812 (2016).
    OpenUrlCrossRef
  13. 13.
    Cash, K. J., Ricci, F. & Plaxco, K. W. An electrochemical sensor for the detection of protein-small molecule Interactions directly in serum and other complex matrices. J. Am. Chem. Soc. 131, 6955–6957 (2009).
    OpenUrlCrossRefPubMed
  14. 14.↵
    Li, H., Dauphin-Ducharme, P., Ortega, G. & Plaxco, K. W. Calibration-free electrochemical biosensors supporting accurate molecular measurements directly in undiluted whole blood. J. Am. Chem. Soc. 139, 11207–11213 (2017).
    OpenUrlCrossRef
  15. 15.↵
    Hou, H. et al. A generalizable and noncovalent strategy of interfacing aptamers with microelectrode for selective sensing of neurotransmitter In vivo. Angew. Chem. Int. Ed. 59, 18996–19000 (2020).
    OpenUrl
  16. 16.
    Chan, D. et al. Combinatorial polyacrylamide hydrogels for preventing biofouling on implantable biosensors. bioRxiv. https://doi.org/10.1101/2020.05.25.115675 (2020).
  17. 17.↵
    Nakatsuka, N. et al. Aptamer-field-effect transistors overcome Debye length limitations for small-molecule sensing. Science. 362, 319–324 (2018).
    OpenUrlAbstract/FREE Full Text
  18. 18.↵
    Li, H. et al. A biomimetic phosphatidylcholine-terminated monolayer greatly improves the in vivo performance of electrochemical aptamer-based sensors. Angew. Chem. Int. Ed. 56, 7492–7495 (2017).
    OpenUrl
  19. 19.↵
    Vallee-Belísle, A., Ricci, F., Uzawa, T., Xia, F. & Plaxco, K. W. Bioelectrochemical switches for the quantitative detection of antibodies directly in whole blood. J. Am. Chem. Soc. 134, 15197–15200 (2012).
    OpenUrlPubMed
  20. 20.↵
    Idili, A. et al. Seconds-resolved pharmacokinetic measurements of the chemotherapeutic irinotecan in situ in the living body. Chem. Sci. 10, 8164–8170 (2019).
    OpenUrl
  21. 21.
    Dauphin-Ducharme, P. et al. Electrochemical aptamer-based sensors for improved therapeutic drug monitoring and high-precision, feedback-controlled drug delivery. ACS Sensors 4, 2832–2837 (2019).
    OpenUrl
  22. 22.↵
    Li, H. et al. High frequency, calibration-free molecular measurements in situ in the living body. Chem. Sci. 10, 10843–10848 (2019).
    OpenUrl
  23. 23.↵
    Ferguson, B. S. et al. Real-time, aptamer-based tracking of circulating therapeutic agents in living animals. Sci. Transl. Med. 5, 213ra165 (2013).
    OpenUrlAbstract/FREE Full Text
  24. 24.↵
    Arroyo-currás, N. et al. Subsecond-resolved molecular measurements in the living body using chronoamperometrically interrogated aptamer-based sensors. ACS Sensors 3, 360–366 (2018).
    OpenUrl
  25. 25.↵
    Liang, S. et al. Protein diffusion in agarose hydrogel in situ measured by improved refractive index method. J. Control. Release 115, 189–196 (2006).
    OpenUrlCrossRefPubMedWeb of Science
  26. 26.↵
    Maciel, C. et al. Evans blue as a simple method to discriminate mosquitoes’ feeding choice on small laboratory animals. PLOS ONE 9, 110551–110562 (2014).
    OpenUrl
  27. 27.↵
    Liu, P., Zhu, Z., Zeng, C. & Nie, G. Specific absorption spectra of hemoglobin at different PO2 levels: potential noninvasive method to detect PO2 in tissues. J. Biomed. Opt. 17, 125002–125008 (2012).
    OpenUrlPubMed
  28. 28.↵
    Vieira, P.A. et al. Ultra-high-precision, in-vivo pharmacokinetic measurements highlight the need for and a route toward more highly personalized medicine. Front Mol Biosci 6, 69 (2019).
    OpenUrl
  29. 29.↵
    Centers for Disease Control and Prevention (CDC) Treatment of tuberculosis. MMWR Morb Mortal Wkly Rep 52(RR-11):27–28 (2003).
    OpenUrl
  30. 30.↵
    Klonoff, D. C. Point-of-care blood glucose meter accuracy in the hospital setting. Diabetes Spectr. 27, 174–179 (2014).
    OpenUrlAbstract/FREE Full Text
  31. 31.↵
    https://www.dexcom.com/getstartedcgm?utm_source=adwords&utm_campaign=b&sfc=70133000001LlpgAAC&gclid=CO3fwIK419MCFRSDfgod2nwEiw.
  32. 32.↵
    Gao, X. et al. Dietary methionine influences therapy in mouse cancer models and alters human metabolism. Nature 572, 397–401 (2019).
    OpenUrlCrossRefPubMed
  33. 33.↵
    Dodd, D. et al. A gut bacterial pathway metabolizes aromatic amino acids into nine circulating metabolites. Nature 551, 648–652 (2017).
    OpenUrlCrossRefPubMed
  34. 34.↵
    Sun, K. H. et al. Evaluation of in vitro and in vivo biocompatibility of a myo-inositol hexakisphosphate gelated polyaniline hydrogel in a rat model. Sci. Rep. 6, 23931 (2016).
    OpenUrl
  35. 35.
    Edlund, U., Albertsson, A. C., Singh, S. K., Fogelberg, I. & Lundgren, B. O. Sterilization, storage stability and in vivo biocompatibility of poly(trimethylene carbonate)/poly(adipic anhydride) blends. Biomaterials 21, 945–955 (2000).
    OpenUrlPubMed
  36. 36.
    Junge, K. et al. Mesh biocompatibility: effects of cellular inflammation and tissue remodelling. Langenbecks Arch. Surg. 397, 255–270 (2012).
    OpenUrlCrossRefPubMed
  37. 37.↵
    1. Saltzman, W. M.
    Saltzman, W. M. in Biomedical Engineering: bridging medicine and technology (eds Saltzman, W. M.) 627 (The United States of America by Cambridge University Press, New York, 2009).
  38. 38.↵
    Wang, D. et al. Expanded poly(tetrafluoroethylene) blood vessel grafts with embedded reactive oxygen species (ROS)-responsive antithrombogenic drug for elimination of thrombosis. ACS Appl. Mater. Interfaces 12, 29844–29853 (2020).
    OpenUrl
  39. 39.↵
    Veiseh, O. et al. Size-and shape-dependent foreign body immune response to materials implanted in rodents and non-human primates. Nat. Mater. 14, 643–651 (2015).
    OpenUrlCrossRefPubMed
  40. 40.↵
    Choi, C., Lee, Y., Cho, K. W., Koo, J. H. & Kim, D. H. Wearable and implantable soft bioelectronics using two-dimensional materials. Accounts Chem. Res. 52, 73–81 (2019).
    OpenUrl
Back to top
PreviousNext
Posted November 17, 2020.
Download PDF

Supplementary Material

Email

Thank you for your interest in spreading the word about bioRxiv.

NOTE: Your email address is requested solely to identify you as the sender of this article.

Enter multiple addresses on separate lines or separate them with commas.
Hydrogel-coating improves the in-vivo stability of electrochemical aptamer-based biosensors
(Your Name) has forwarded a page to you from bioRxiv
(Your Name) thought you would like to see this page from the bioRxiv website.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Share
Hydrogel-coating improves the in-vivo stability of electrochemical aptamer-based biosensors
Shaoguang Li, Jun Dai, Man Zhu, Netzahualcóyotl Arroyo-Currás, Hongxing Li, Yuanyuan Wang, Quan Wang, Xiaoding Lou, Tod E. Kippin, Shixuan Wang, Kevin W. Plaxco, Hui Li, Fan Xia
bioRxiv 2020.11.15.383992; doi: https://doi.org/10.1101/2020.11.15.383992
Reddit logo Twitter logo Facebook logo LinkedIn logo Mendeley logo
Citation Tools
Hydrogel-coating improves the in-vivo stability of electrochemical aptamer-based biosensors
Shaoguang Li, Jun Dai, Man Zhu, Netzahualcóyotl Arroyo-Currás, Hongxing Li, Yuanyuan Wang, Quan Wang, Xiaoding Lou, Tod E. Kippin, Shixuan Wang, Kevin W. Plaxco, Hui Li, Fan Xia
bioRxiv 2020.11.15.383992; doi: https://doi.org/10.1101/2020.11.15.383992

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
  • Tweet Widget
  • Facebook Like
  • Google Plus One

Subject Area

  • Bioengineering
Subject Areas
All Articles
  • Animal Behavior and Cognition (4845)
  • Biochemistry (10777)
  • Bioengineering (8034)
  • Bioinformatics (27256)
  • Biophysics (13964)
  • Cancer Biology (11110)
  • Cell Biology (16026)
  • Clinical Trials (138)
  • Developmental Biology (8772)
  • Ecology (13269)
  • Epidemiology (2067)
  • Evolutionary Biology (17342)
  • Genetics (11679)
  • Genomics (15904)
  • Immunology (11013)
  • Microbiology (26044)
  • Molecular Biology (10627)
  • Neuroscience (56464)
  • Paleontology (417)
  • Pathology (1729)
  • Pharmacology and Toxicology (2999)
  • Physiology (4539)
  • Plant Biology (9614)
  • Scientific Communication and Education (1613)
  • Synthetic Biology (2683)
  • Systems Biology (6968)
  • Zoology (1508)