Abstract
Vagus nerve stimulation (VNS) is used as therapy in epilepsy and depression and is tested as a potential treatment for several chronic disorders. Typically, VNS is delivered at increasing stimulus intensity until a response is observed (threshold intensity). Factors that affect threshold intensities for engagement of different fiber types and concomitant physiological responses have not been studied. We determined neural and physiological responses to increasing stimulus intensities of VNS in anesthetized and awake animals, and examined the effect of implant- and anesthesia-related factors on threshold intensities in a rodent model of VNS. In acute and long-term cervical vagus nerve implants (53 and 14 rats, respectively) VNS was delivered under isoflurane, ketamine-xylazine, or awake at different intensities. Stimulus-evoked compound action potentials (eCAPs) were recorded, elicited physiological responses were registered, including changes heart rate (HR), breathing, and blood pressure (BP), and threshold intensities were determined. The intensity that elicits eCAPs (“neural threshold”) is significantly lower than what elicits a physiological response (“physiological threshold”, PT) (25 μA ±1.8 vs. 70 μA ±5.2, respectively; Mean ±SEM). Changes in BP occur at the lowest stimulus intensities (80 μA ±7), followed by changes in HR (105 μA ±8.4) and finally in breathing (310 μA ±32.5). PT is lower with than without electrode insulation (60 μA ±12, vs. 700 μA ±123). PT and electrode impedance are correlated in long-term (r=0.47; p<0.001) but not in acute implants (r=-0.34; p NS); both PT and impedance increase with implant age (Pearson correlation r=0.44; p<0.001 and r=0.64; p<0.001, respectively). PT is lowest when animals are awake (210 μA ±33; Mean ±SEM), followed by ketamine-xylazine (630 μA ±154), and isoflurane (1075 μA ±131). The sequence of physiological responses with increasing VNS intensity is similar in both anesthetized and awake states. Implant age, electrical impedance and the type of anesthesia affect VNS threshold and should be accounted for when determining stimulation dose.
Introduction
In neuromodulation, electrical energy is delivered with the intent of altering the function of the central or peripheral nervous system. It can be delivered either by means of invasive (e.g., deep brain stimulation and peripheral nerve stimulation)1–4 or non-invasive stimulation (e.g., transcranial stimulation and transauricular vagus nerve stimulation)5–7. Usually, neurostimulation is initiated at a low intensity, which is increased gradually until a neurophysiological, behavioral, or physiological response is observed. The minimum intensity to elicit such a response is defined as the stimulation threshold. At the cellular level, the stimulation threshold corresponds to the electrical energy level that is just enough to activate neural elements, typically neurons and their axonal and dendritic processes, with the lowest excitability threshold8, 9. At higher intensities, elements with higher excitability thresholds get recruited, leading to increasingly stronger or additional responses8, 10.
The type of threshold responses and the way they are measured depends on the site of neurostimulation. Examples of threshold responses are the motor evoked potential (MEP), a neurophysiological response to cortical stimulation11, reduction in pain, a behavioral response to spinal cord stimulation12, and a change in heart rate, a physiological response to cervical vagus nerve stimulation (VNS)13. These examples underscore that stimulation threshold responses depend on which components of the nervous system are activated or inhibited, but also on the end-organs through which that activation, or inhibition, manifests itself. For example, motor cortex stimulation would not elicit the same MEP in subjects taking muscle relaxants and in subjects not taking any medication14. Similarly, the behavioral response to spinal cord stimulation depends on cognitive state15. Therefore, the physiological state of the neural circuits and end-organs is important when the stimulation threshold is determined.
Knowledge of threshold intensities is essential in neuromodulation studies and therapies. The threshold for a given stimulation modality varies widely across subjects16, 17. Mechanisms that determine threshold can be grouped into 3 categories: those related to the neural interface, those related to the generated electrical field, and those related to the physiological status of the end-organ. Factors that determine the neural interface include surgical technique2, 11, foreign body response to a long-term implanted electrode1, 8,19, and degradation of the electrode itself20. Factors influencing the generated electrical field include electrode geometry21,22, electrode orientation relative to the stimulated neural elements17, 21, 23, and stimulation parameters11, 21. Finally, the state of a subject or the end-organ at the time of stimulation affects the stimulation threshold24. For example, general anesthesia has distinct effects on the nervous system and multiple end-organs, altering stimulation thresholds through several mechanisms24. Factors affecting stimulation thresholds in brain stimulation2, 11,21, spinal cord stimulation23, and somatic peripheral nerve stimulation, such as tibial or ulnar nerve18, 20, 24 are relatively well-studied. There is much less data on factors affecting stimulation thresholds in vagus nerve stimulation (VNS), an established neuromodulation therapy in epilepsy and depression1, 25, 26, and currently under study in a variety of cardiovascular, immune, and metabolic disorders27–30.
In this study, we examined several factors affecting threshold intensity in cervical VNS in rats, and more specifically the presence of insulation at the neural interface, the level of ex-vivo and in-vivo electrode impedance, the age of the implant, and the conditions of anesthesia. To address these questions, we developed acute and long-term implants of vagus nerve stimulation and recording, and a methodology to determine thresholds in individual animals. We found that the threshold for evoking nerve compound action potentials with VNS (“neural threshold”, NT) is lower than that for eliciting a reproducible physiological response (“physiological threshold”, PT). The PT has a weak correlation with ex-vivo or in-vivo electrode impedance on an individual measurement basis, even though they both increased with implant age. Anesthesia caused a significant increase in the threshold of VNS, compared to the conscious state: isoflurane has the largest effect, followed by ketamine-xylazine.
Methods
Animal preparation
Male Sprague-Dawley rats (n=67) weighing 250-400gm were used in this study. Institutional Animal Care and Use Committee approved all animal experiments. lsoflurane was a primary anesthetic agent (induction 4%, maintenance 1-2%). Ketamine (80 mg/kg IP) and xylazine (6 mg/kg IP) were used in a few experiments. After induction, rat was placed in a supine position, and hairs were removed from surgical sites. Alcohol and Betadine were used to disinfect surgical sites. Adequate level of anesthesia was assessed by pedal reflex and breathing rate.
Instrumentation of physiological sensors
Physiological signals were monitored during experiments. Nasal temperature sensor was inserted inside nostril to detect temperature changes during inhalation and exhalation; breathing rate was calculated using nasal temperature sensor (fig. 1). Electrocardiogram (ECG) was recorded using 3-needle electrodes inserted intradermally (fig. 1). Left femoral artery was isolated, and polyethylene-tube 50 was inserted to monitor systemic arterial pressure (SAP) (fig. 1). Body temperature was measure using rectal probe and maintained at 37-37.5°c using heating pad.
Electrodes
The “Flex” electrodes used for this study were designed for acute and long-term implantation and made using microfabrication processes31 (fig. 1B). Some of the acute experiments were conducted using commercially available CorTec bipolar electrodes (CorTec, Germany). Additional microfabrication details of electrodes are described in supplementary file.
Surgical procedure for acute implantation
A midline skin incision of 2-3 cm was given on the neck. After the incision, a submandibular gland can be visualized (fig. 2A). This gland was separated to visualize sternohyoid (midline) and sternocleidomastoid (oblique) muscles (fig. 2B, 2C). The sternocleidomastoid muscle was then retracted using magnetic retractors (Fine Science Tools, CA) to reach the carotid bundle (common carotid artery and vagus nerve; fig. 2C). A third muscle named omohyoid can be seen connecting sternohyoid and sternocleidomastoid muscles; carotid bundle can be visualized rostral and caudal to the omohyoid muscle (fig. 2C, 2D). A dissecting microscope (Carl Zeiss Microscopy, NY) was used to separate vagus nerve from carotid bundle (fig. 2D). The rostral end of vagus nerve was isolated, and in a few experiments, the caudal end was also isolated. The middle part of the nerve embedded under omohyoid muscle was left intact to minimize nerve manipulation. A bipolar or tripolar electrode was placed under the nerve (fig. 2E). In-case of two electrodes, the distance between them was always between 5-6 mm. A silicone elastomer Kwiksil (World Precision Instruments, FL) was used to insulate the nerve-electrode interface from the surrounding fluid.
Surgical procedure for long-term implantation
A 1 cm skin incision was given on the skull while the rat was in a prone position. Scalp tissue was cleared using hydrogen peroxide between bregma and lambda. A drilling machine using a bit of shaft diameter 2.3 mm (19007-09, FST, CA) was used to drill two holes on the right and left side of the skull, between coronal and lambdoid sutures. Self-tapping bone screws (19010-10, FST, CA) were screwed on these holes using a screwdriver (Stanley, CT). Rat was then placed in a left lateral recumbent position. A subcutaneous tunnel was then made between skull and neck using a hemostat, through which electrode leads were tunneled. Rat was then placed in a supine position, and electrode was placed around the left vagus nerve, as explained above. A small ligating clip (LT-100, Johnson & Johnson, NJ) was used to close Flex electrode. Electrode leads were sutured on a surrounding muscle to minimize nerve manipulation during physical movement. Skin was closed using 4.0 nylon-suture. Rat was then again placed in a prone position to fix the head-connector on the skull. A layer of acrylic dental cement (A-M systems, WA) was applied on the exposed area of skull, head connector was then fixed between the two skull screws (fig. 1A). Subcutaneous antibiotics (Enfloroxacin 5 mg/kg) and Meloxicam 3 mg/kg were given once a day for three days post-surgery. Few rats were purchased pre-implanted with a telemetry device (HD-S21, Data Science International, MN) to record the effect of VNS on systemic arterial pressure and heart rate on freely moving rats.
Stimulation and Recording
All biological signals were first pre-processed using Bio Amplifier for ECG signals, Bridge Amplifier for pressure signals, and T-type Pods for nasal temperature sensor and rectal probe. These amplifiers and pods were linked to Powerlab (ADlnstruments Inc, CO). All the raw signals captured by Powerlab were displayed on LabChart (ADI, CO). The biological signals from telemetry implanted rats were captured on PhysioTel Hardware (OSI, MN), and were displayed on LabChart. The stimulator (STG4008, Multi-Channel Systems, DE) was connected to the rostral electrode, and a recording amplifier (RHS2000 series, lntan Technologies, CA) was connected to the caudal electrode. Physiological threshold (PT) was defined as intensity needed to elicit a 5-10% decrease in heart rate (HR). Neural threshold (NT) was defined as intensity needed to elicit a visible eCAP on a recording electrode. Stimulation parameters used for determining thresholds were constant current, rectangular pulses with pulse widths (PW) of 100 μs, frequency (F) of 30 Hz, and duration of 10 s. Stimulus triggers were registered on LabChart as well as on lntan software and were used to synchronize the recordings. Parameters used in this study are PW of 100 μs, F= 30 Hz, and duration 10 s. Bipolar stimulation was used in terminal and survival experiments.
Analysis
Analysis of physiological and neurophysiological signals
Physiological and neural recording data were processed using custom scripts on MATLAB (Mathworks). The ECG signal was first pre-processed with a high-pass filter at 0.1 Hz to remove direct current (DC) voltage drift, followed by QRS peak detection to calculate heart rate (HR) based on individual R-R intervals. For nasal temperature sensor signal, a high-pass filter at 0.1 Hz was applied to eliminate the temperature drift. Both positive and negative peaks were detected to compute breathing cycles for accurate detection of each breath.
The change in breathing rate (ΔBR) and heart rate (ΔHR) for stimulation trains were calculated as follow: where BRpre (or HRpre) corresponds to the mean BR (or HR) during the 1Os long train before the stimulation, and BRVNS (or HRVNS) to the mean BR (or HR) during the stimulation. Neural signal processing and elimination of non-neural signals captured from a recording nerve electrode. Briefly, non-neural components in eCAP recordings include ECG-artifact, stimulus-artifact, and EMG-artifact from laryngeal muscle contraction. To improve nerve signal quality and resolve short-latency fiber responses, a stimulus artifact suppression algorithm was employed.
Statistical analysis
A paired t-test was used to compare the physiological and neural thresholds. A paired t-test was used to compare physiological thresholds of animals under isoflurane, ketamine-xylazine, and awake state. Pearson-correlation coefficient was used to find correlation between physiological and neural thresholds, and between physiological threshold and electrode impedance. Comparisons were deemed statistically significant for p<0.05 for all analyses. All statistical analyses were conducted on MATLAB (Mathworks).
Results
Neural threshold is significantly lower than physiological threshold
We sought to investigate the relationship between the minimum stimulus intensity that elicits a neural response (“neural threshold”, NT) and the minimum intensity that elicits a vagally-mediated physiological response (“physiological threshold”, PT). We determined NT by measuring stimulus-evoked compound action potentials (eCAPs) (fig. 3A). We determined PT by measuring stimulus-elicited change in heart rate (HR). eCAP responses comprise components with different latencies, corresponding to activation of different fiber types, recruited at different intensities: short latency A-fibers are engaged at low intensities, whereas intermediate latency B-fibers are engaged at higher intensities (fig. 3A). This fiber recruitment pattern correlates well with the physiological responses seen during VNS: at low-intensities, A-fibers mediate an increase in SAP and moderate drop in BR, at intermediate-intensities B-fibers mediate drop in HR, and secondarily in SAP, and at high intensities C-fibers mediate apnea responses (fig. 3A, 4).
In 19 animals, NT range between 14 and 40 μA and PT between 42 and 120 μA. Mean NT (25 μA± 1.83) is significantly smaller than mean PT (70 μA± 5.24) (p<0.01, paired t-test) (fig. 3B). On a single animal basis, NT is always lower than PT (fig. 3C), and NT and PT were correlated: animals with relatively low NT values have relatively low PT values, and vice-versa (Pearson-correlation coefficient: r=0.7, p<0.01). NT and PT are not significantly different between right and left VNS (Suppl. fig. S1). In a separate group of 4 rats, neural and physiological responses were compared by using 2 different stimulation frequencies, 1 and 30 Hz; neural responses are similar in both frequencies, whereas physiological responses are significantly different (table I). These results indicate that there is a significant, almost 3-fold, variability in both NT and PT across animals and that PT is 2-4 times greater than NT. These results also suggest that stimulation frequency of VNS affects the magnitude of the physiological response, but not of the neural response.
Acute cardiopulmonary responses to VNS depend on stimulus intensity
Different acute cardiopulmonary responses are elicited at different VNS intensities, including changes in breathing rate, usually in the form of slower breathing, and in heart rate, usually in the form of bradycardia (fig. 3A). Generally, intensity is a major determinant of the pattern of physiological responses to VNS. In animals anesthetized with isoflurane, at low VNS intensities, we routinely observed slower breathing and increase SAP, both of which returned to baseline after the end of stimulation (fig. 4A). At intermediate intensities, slower breathing and bradycardia are typically seen; often, the bradycardic response is associated with a drop in SAP; again, parameters returned to baseline after the end of stimulation (fig. 4B). At higher intensities, we observed apnea during stimulation, more intense bradycardia, and a larger drop in SAP; even though breathing and heart rhythm returned to baseline after the end of stimulation, a rebound increased in SAP is commonly seen (fig. 4C). Thresholds for different responses under anesthesia are summarized in table. II. Heart rate responses to VNS in awake, freely behaving animals are comparable to those in anesthetized animals, but SAP responses are often more complex and variable (fig. 4D-F). Overall trends for low, intermediate, and high intensities of anesthetized and awake groups are summarized in table. III. These results indicate that acute cardiopulmonary responses to VNS depend on the stimulus intensity.
Physiological threshold is lower with electrode insulation
In many acute experiments, we observed an increase in PT when moderate amounts of fluid accumulated around the stimulation electrode. We compared PT between experiments in the same animals with and without application of a silicone elastomer for additional insulation (Suppl. fig. S2). In the same 7 animals, PT with application of insulating elastomer is significantly lower than without: 60 μA ±12.3, vs. 700 μA ±122.6 (Suppl. table I). In one animal, we were able to reduce PT by dabbing the fluid from the area around the stimulating electrode with an absorbing cotton tip (Suppl. table I). These results indicate that additional insulation of the stimulating electrode prevents current shunting by accumulated fluid, resulting in lower PT values.
Physiological threshold increases with increasing implant age
To determine whether implant age affects physiological responses to VNS, we chronically implanted stimulating electrodes on the cervical vagus and determined PT under anesthesia (isoflurane) in multiple stimulation sessions across several weeks until no physiological response was observed for two consecutive stimulation sessions. In general, the intensity to elicit the same physiological response is increased with increasing age; for example, to elicit a drop in HR by 8-10% 1 day after implantation intensity required is 100 μA, whereas at 60 days later, it is 1000 μA (fig. 5A, 5B). In 14 rats, PT is gradually increased significantly during the first 2 weeks, and only moderately afterward (fig. 5C). Mean PT at the day of implantation is 190 μA± 28.5 (n=14), 650 μA ±112.8 (n=12) at 2 weeks post-implantation, 670 μA ±100 (n=9) at 4 weeks, 630 μA ±160.7 (n=4) at 8 weeks, and 950 μA ± 550 (n=2) at 12 weeks. Additionally, significant amount of fibrosis is observed around the neural interface during the autopsy; the fibrotic response seems to get worse with the age of the implant (Suppl. fig. S3). These results indicate that implant age is positively correlated with PT, with the greatest increases happening during the first two weeks post-implantation.
Physiological threshold is affected by in-vivo/ex-vivo electrical impedance
To determine whether the level of the electrical impedance of the stimulating electrode, often used to infer the functional status of an electrode or the tissue-electrode interface, affects PT, we measured in-vivo/ex-vivo electrical impedance and PT in the same stimulation sessions, in acute and long-term nerve implants. In 53 acute experiments, mean ex-vivo impedance at 1KHz before implantation is 1.6 kΩ ±0.14, and mean PT is 80 μA ±5.84. The correlation between ex-vivo impedance and PT is non-significant (r = 0.21, p NS) (Suppl. fig. S4). In 12 animals, we compared the ex-vivo and in-vivo impedances of electrodes, measured immediately before and after implantation, respectively; ex-vivo impedance is always lower than the in-vivo impedance, and the two impedance values are highly correlated (r = 0.83) (Suppl. fig. S4). In 14 rats implanted with long-term electrodes, PT progressively increases with implant age (fig. 5C; r = 0.44, p<0.001). Electrode impedance also increases with age (fig. 5D) (r = 0.64, p<0.001). On a single measurement basis, there is a weak but statistically significant correlation between PT and impedance (fig. 5E; r = 0.47, p<0.001). In 3 out of 14 long-term implants, impedances were low but no physiological response was observed (fig 5E). These results indicate that in-vivo/ex-vivo electrical impedance of the stimulating electrode is a poor predictor of PT in acute implants. In-vivo impedance is a predictor in most, but not all, long-term implants.
Physiological threshold is higher under anesthesia compared to wakefulness
To determine the effects of anesthesia on PT in rats chronically implanted with vagus electrodes, we measured PT under anesthesia with isoflurane, with ketamine-xylazine, and during awake conditions. In these experiments, the animal was first anesthetized with isoflurane, PT was measured, and isoflurane was turned off; after the animal recovered from isoflurane, PT was determined again, either in an awake state (first animal group; fig. 6A, 6B) or after a second round of anesthesia with ketamine-xylazine (second animal group). In the first group of 5 animals (23 experiments), PT measured under anesthesia with isoflurane is 1020 μA ±122.3, and in awake conditions, it is 210 μA ±33.5 (p<0.001, Paired t-test) (fig. 6C). In the second group of 5 animals (7 experiments), PT measured under isoflurane is 1215 μA ±207.5, and under ketamine-xylazine is 630 μA ±153.8 (p<0.001, Paired t-test) (fig. 6C). To determine whether these differences persisted at supra-threshold intensities, we measured the stimulus-elicited change in HR (ΔHR) to the same suprathreshold intensities, under isoflurane vs. awake, or under isoflurane vs. ketamine-xylazine. In the first group, the mean change in heart rate (ΔHR) under isoflurane is −2 % ±0.7, and in awake conditions, it is −26 % ±3.3 (p<0.01, paired t-test) (fig 6D). In the second group, HR under isoflurane is −1 % ±1.01 and under ketamine-xylazine it is −13 % ±3.98 (p<0.01, paired t-test) (fig. 6E). These results indicate that PT are the smallest when the animals are awake, followed by anesthesia with ketamine-xylazine and finally with isoflurane.
Discussion
In animal studies of vagus nerve stimulation (VNS), determining stimulation threshold is commonly one of the first steps in an experiment, as it defines the minimal effective “dose”. Expressing stimulation dose in units of threshold removes variability between experimental conditions or between animals (fig. 7). The selection of the type of threshold depends on the study outcome: “neural threshold” (NT), the minimal intensity that elicits nerve activity, is appropriate when nerve action potentials are studied32, whereas “physiological threshold” (PT), the minimal intensity that produces a physiological response, is more suitable when the physiological effects of VNS are of interest33, 34.
In our study, we determined both NT and several types of PT to cervical VNS in laboratory rats. PT was determined with respect to several different physiological responses, which followed a sequence as stimulus intensity was increased: increase in blood pressure (BP) and drop in breathing rate (BR) typically occur at lowest intensities, drop in heart rate (HR), and a resulting drop in BP (under isoflurane) or increase in BP (awake conditions), at intermediate intensities, and finally apnea, along with drop in HR and BP, at high intensities (table. III). Interestingly, the same sequence of physiological effects with increasing intensity is seen in both anesthetized and awake animals (fig. 4), something reported for the first time, to the best of our knowledge. This means that anesthesia does not alter the sequence of physiological effects of VNS as stimulus intensity increases.
This sequence of physiological effects can be explained by the stimulus intensity-dependent engagement of vagal fibers by VNS, which follows a size principle: larger fibers have relatively low thresholds and smaller fibers have high thresholds35. At low intensities, activation of larger A-delta fibers from lung stretch receptors may be responsible for the drop in breathing rate, as part of the Herring-Breuer reflex36–38. At those intensities, afferent A-delta fibers from aortic baroreceptors39 are also activated. Isolated activation of those fibers is known to elicit a drop in BP40, in contrast to the increase in BP we documented at low intensities. It is likely that in our case, activation of A-delta fibers happens concurrently with co-activation of efferent sympathetic fibers, of which there is a small number in the vagus41, resulting in vasoconstriction and increase in BP. It is also likely that engagement of afferent A-delta fibers initiates vagal-sympathetic reflexes resulting in an increase in the sympathetic tone to vessels, resulting in vasoconstriction, and to the heart, resulting in tachycardia34. Afferent fiber activation happening at these low intensities could also centrally suppress the parasympathetic tone to the heart33, 34. Such vagal-sympathetic reflexes are mediated centrally by direct connections from the vagal sensory nucleus of the solitary tract (NTS) to the dorsal motor nucleus (DMN) of the vagus and sympathetic ganglia34, 42, 43. As stimulus intensity increases further and cardio-inhibitory B-fibers become more engaged, their cardioinhibitory action at the heart dominates, and drops in HR and BP ensues33, 34 (fig. 4). Interestingly, in awake conditions, a similar drop in HR is observed but with an increase in SAP (table. III). This difference in response between anesthetized and awake conditions is possibly due to the fact that isoflurane suppresses sympathetic activity which resulted in the absence of an increase in SAP under isoflurane44, 45. Finally, activation of smaller C-fibers at highest stimulus intensities in rats produces apnea38, 46, 47.
On a single animal-basis, NT and PT are positively and highly correlated (fig. 3C); this finding may be used to estimate NT by measuring PT, under specific conditions. In general, measuring PT is much faster and easier than determining NT, which involves the step of recording stimulus-evoked compound action potentials (eCAPs). We found that NT ranged between 14-40 μA (single pulses of 100 μs pulse width) and PT between 32-150 μA (trains of 100 μs-long pulses at 30 Hz pulsing frequency). The higher values for PT than for NT are expected, given the additional anatomical elements involved in the determination of PT and have been described before10, 35. This difference can be explained by the neural components engaged in their measurement. For NT, those include the electrode-nerve interface and the level of excitability of the stimulated nerve fibers. Determining PT in addition, involves synaptic transmission at the ganglionic synapse, the level of excitability of the corresponding ganglionic cells, and the postganglionic synapse or circuit at the end-organ, as well as the state, at the time of stimulation, of the end-organ itself (fig. 8). It follows that, due to synaptic temporal summation, PT should be dependent on pulsing frequency. Indeed, we found that the physiological responses to VNS trains are different between 2 pulsing frequencies (1 Hz vs. 30 Hz), even though the eCAP responses to individual pulses of those same trains are largely similar (table I). We hypothesized that since a physiological response relies both on nerve activation and engagement of the neural circuit controlling the end organ (sinus node of the heart, baroreflex or respiratory center in the brain): (a) NT should be lower than PT, (b) PT, but not NT, should be dependent on pulsing frequency.
Adequate insulation of an electrode during in-vivo implantation is required to focus the resulting electrical field to the targeted nerve and to limit current spread and activation of non-targeted excitable tissues in the vicinity of the electrode48–51. In our experiments, PT with extra insulation, in the form of a silicone layer completely surrounding the electrode-nerve interface, is 10 times smaller than PT without that insulation (Suppl. table I). In one animal, “non-insulated” PT is reduced to a similar level as “insulated” PT after dabbing the fluid accumulation around the interface (Suppl. table I), indicating that the higher intensities needed with non-insulated interfaces are likely due to current shunting between the electrode contacts. In long-term implants, the insulating material must be biocompatible52, or else, tissue reaction may compromise the nerve and the electrode-nerve interface53, hence affecting the performance of the implant.
We found that PT increases with implant age (fig. 5C). Similarly, the intensities required to elicit a given suprathreshold physiological response are significantly higher at a later age compared to an earlier implant age (fig. 5A, 5B). This gradual rise in PT is due to a multifactorial process involving inflammatory and fibrotic response to the presence of the electrode18, 19, nerve injury with loss of nerve fibers54, and degradation of electrode contacts or leads20. Electrode impedance reflects the electrical characteristics of the electrode-nerve interface and has been suggested to be a measure of its effectiveness and integrity55. In our study, in-vivo electrode impedance increases with implant age (fig. 5D), in agreement with other reports56, 57. Both PT and impedance increase with implant age, having a weak but significant correlation on a single measurement basis (fig. 5E). When impedance is high (greater than 200 kΩ), PT is always indeterminable (fig. 5E), indicating a non-functional implant, but measurable impedance does not always guarantee a functional implant57 (fig. 5E). In acute experiments, neither ex-vivo nor in-vivo electrode impedance is correlated with PT (Suppl. fig. S4). That may be because PT is closely related to the spatial relationships between the electrode contacts and the different nerve fiber types that produce the physiological responses, especially at the relatively low threshold intensities used in our study. The anatomical heterogeneity of the fiber composition of the cervical vagus between animals58, the presence or absence of aortic depressor nerve in the carotid sheath along with the vagus59, the exact orientation of the electrode and the resulting electric field in relation to the nerve22, are all significant determinants of PT and their contribution is not captured by measuring in vivo or ex vivo electrode impedance.
The use of general anesthesia is typically a prerequisite for acute rodent experiments. General anesthetic agents provide analgesia and loss of consciousness, primarily by acting on the central nervous system (CNS). Actions on the autonomic nervous system (ANS) have also been described, including actions on ganglionic neurons60, synaptic transmission61–63, and higher anesthesia concentration can also block the axonal action potentials64, 65. Additionally, general anesthesia reduces the receptors’ sensitivity for neurotransmitters66, 67. In our study, we found that the negative chronotropic effect of VNS is maximum in awake state, followed by ketamine-xylazine, and isoflurane (fig. 6C-6E). These differences between awake and anesthetized states are likely because of the effects of general anesthesia on ANS as explained above. Our findings are in agreement with a previous study comparing isoflurane and awake states, which shows that at a given stimulation intensity, the chronotropic response of VNS was more pronounced in awake conditions than under isoflurane (figure 5 of Seagard et al68). We also found that PT is significantly lower in awake state compared to anesthesia (fig. 6C). The increase in PT under general anesthesia is also consistent with other peripheral nerve stimulation studies, which show the stimulation threshold of muscle contraction (end-organ) increases with general anesthesia, possibly as a result of inhibition of nicotinic acetylcholine receptors at the neuromuscular junction69–71. In addition to the effect of anesthesia on ANS, they also affect the cardiovascular system, specifically the heart. In particular, isoflurane has known to depress the cardiovascular system by inhibiting ion-channels on cardiac muscles and pacemaker. The channels it inhibits are the sodium channels responsible for generating action potentials72, the voltage-gated cardiac calcium channel responsible for generating pacemaker potentials73, and the voltage-gated potassium channels that delay the recovery from the action potential74. On the other hand, ketamine is an N-methyl-D-aspartate (NMDA) antagonist, which can also modulate parasympathetic cardiac activity by altering nicotinic cholinergic excitatory mechanisms75. The effects of ketamine on ANS, especially vagal tone, is controversial. One study indicated that ketamine does not affect the vagal tone, such as its effects on ganglion and synaptic potential transmission is limited76, while the other stated that it affects vagal tone by stimulating vagal fibers77. Additionally, xylazine is an alpha-2 adrenergic receptor antagonist that can cause central nervous system (CNS) depression78. We found out that physiological thresholds are significantly lower under ketamine-xylazine compare to isoflurane (fig. 6C); these findings can be explained by the fact that isoflurane has been shown to have a more pronounced effect on ANS and the heart compare to ketamine-xylazine. Hence, a pronounced chronotropic effect of VNS would be expected with ketamine-xylazine compare to isoflurane. Therefore, the discrepancies in physiological thresholds observed in awake and anesthetized conditions are probably due to a combination of anesthetic agent’s effects on the autonomic nervous system, as well as on the heart (end-organ).
Acknowledgement
All authors declare no conflict of interest.
Footnotes
Sources of financial support This work was supported in part by a research grant to SZ from United Therapeutics Corporation.
Authorship statement UA conceived and designed experiments, performed experiments, analyzed and interpreted experimental results, and wrote the manuscript. YCC, MFL and JW performed experiments. TDC, LR and YAA critically reviewed and edited the manuscript. SZ conceived and designed experiments, analyzed and interpreted experimental results, wrote the manuscript, and obtained funding.