Abstract
In this study, we demonstrate the fabrication, characterization, and application of a microchip-based electrokinetic biosensor which exploits streaming current for signal transduction. The sensor chips, fabricated using standard silicon fabrication techniques, are based on Si-glass microfluidics offering precision, robustness, and transparency. A custom-built chip-manifold allowing easy interfacing with standard microfluidic connections is also developed which shows leak-free integration of the microchips up to 6 bar of applied pressure. Within this range of pressure, the devices show linear and highly reproducible values for flow rates and streaming current with RMS noise below 20 pA. The microchips designed for multiplexed measurements were tested with the detection of free proteins (streptavidin) and also transmembrane proteins of small extracellular vesicles (sEVs) to demonstrate the capacity of the microchips to detect various types of bio-analytes. The limit of detection (LOD) for streptavidin was estimated to be 0.5 nM while for the transmembrane protein (CD9), the LOD was found to be 1.2×106 sEVs/mL. The sensitivity (LOD) of the devices was found to be about 4 times better in targeting CD9 transmembrane protein of H1975 extracellular vesicles when compared to commercial silica capillary which was used previously. The improvement in LOD is attributed to the higher surface quality of the sensor in terms of the density of surface charges which may be further exploited for even lower LOD. In addition, optical detection of fluorophore-tagged standard proteins was done through the optical window of the chip manifold and the transparent glass cap of the microchip.
1. Introduction
During the past years, the principle of electrokinetic-biosensor exploiting streaming current/potential method has been applied in the detection of a wide variety of bio-analytes including proteins [1, 2], ligands [3], DNA [4] and extracellular vesicles [5], thereby demonstrating the excellent potential of the method as a generic biosensor. The method relies on the electrostatic and hydrodynamic interaction at the solid-liquid interface inside a microfluidic channel and allows for label-free detection of bio-analytes. A major benefit of the method stems from its high sensitivity to the surface coverage of an analyte, which has been previously studied with respect to its influence on the streaming current [6] and lately, for the determination of concentration of bio-analytes [7]. Besides, the method offers several practical benefits, such as low sample consumption, simple and inexpensive device architecture, and possibility to integrate with standard microfluidic technologies for sample sorting, enrichment, and deliveries. These advantages have obviously attracted further interest in the method aiming for both improved understanding and exploitation of the governing principles. In particular, we have recently demonstrated that the surface charge density and the charge contrast between the sensor surface and analyte play a major role which can be exploited to achieve a far better sensitivity [8, 9]. Further, by designing appropriate charge-labelled detection probe, we demonstrated the possibility to develop an immunosandwich assay in electrokinetic sensing, thereby, extending the application of the method in complex biofluid [8]. A multiplexed detection setup for simultaneous measurement of several bio-analytes has also been reported [10].
Clearly, the detection principle has matured in its ability to analyze bio-analytes, significantly improving in both sensitivity and specificity. An improved understanding of how the physical parameters of an analyte, such as its size and charge, influence the sensor response [11] now provides us new opportunities to design a more sensitive detector. However, the developments so far have mainly been shown using commercial silica capillary, which has limited scope to exploit many of the benefits in practical settings. Instead, an implementation of the detection principle on a microfabricated channels can help to further leverage some of the key advantages of the method. This includes the design of shorter and narrower channels to further reduce sample consumption, improving the quality of surface oxides for better sensitivity and integration of large number of channels for improving the throughput of multiplexing. In addition, such a microfabricated sensor can open new avenues for research, e.g., new material for sensing surface, integration of fluidic actuation, exploring the benefit of nano-engineered surface and etc.
In this study we, for the first time, report on the fabrication and characterization of a microfabricated electrokinetic biosensor and a chip manifold for easy integration of the microchips with standard fluidic connections. The microchannels were fabricated on Si substrate followed by anodic bonding with glass to achieve high surface quality as well as robust and transparent fluidic channels. The channels were characterized to have low surface roughness and uniform cross section along the length. The devices show linear and highly reproducible values for flowrate and streaming current with a standard deviation below 5 pA, as a function of applied pressure. The microchips designed for multiplexed measurements were tested with the detection of free proteins (streptavidin) and also transmembrane proteins of small extracellular vesicles (sEVs) to demonstrate the capacity to detect various types of bio-analytes. The limit of detection (LOD) for streptavidin was estimated to be 0.5 nM while for the transmembrane protein (CD9), the LOD was found to be 1.2×106 sEVs/mL. The sensitivity (LOD) of the devices were improved by about 60% in case of streptavidin and 4-fold for CD9 targeted sEVs, when compared to commercial silica capillary [9]. The improvement in LOD is attributed to a higher density of surface charges which may be further exploited for even lower LOD. In addition, the chip manifold and the microchip are equipped with optical window and a transparent glass to utilize optical detection of the captured targets. The development of highly sensitive and reusable electrokinetic microchip-biosensor is expected to take the sensing principle one step closer to clinical applications in the near future.
2. Materials and Methods
2.1. Reagents
For the study we used ultra-pure deionized water (resistivity: 18 MΩ. cm) locally produced. Phosphate buffered-saline tablets, avidin from egg-white, streptavidin from Streptomyces avidinii were purchased from Sigma Aldrich Sweden AB. For cleaning of the chips, hydrogen peroxide and ammonium hydroxide were used which were also procured from Sigma Aldrich Sweden AB. The capturing probes consist of Poly(L-lysine)-graft-biotinylated PEG (PLL-g-PEG-biotin) were purchased from Nanosoft Polymers. Biotinylated human Anti-CD9 (MEM-61) (catalog no. MA1-19485) antibody and mouse IgG1 isotype control (catalog no. IC002B) antibodies were purchased from Thermofisher Scientific and Bio-Techne, UK, respectively.
2.2. sEVs isolation, collection, and characterization
In this study sEVs were isolated from cell culture media of the non small-cell lung cancer (NSCLC) cell line H1975 (ATCC® CRL-5908™, LGC Standards, Wesel, Germany). The cells were cultured in RPMI-1640 medium with addition of fetal bovine serum (FBS, 10%) and 2 mM L-glutamine (Gibco, Life Technologies, Stockholm, Sweden). sEVs were isolated from 50 mL of cell culture media with two steps of centrifugation to clear out cell debris (200 RCF for 5 min followed by centrifugation of supernatant at 720 RCF for 10 min, Rotina R38 centrifuge, Hettich). Media was concentrated to about 500 µL using Amicon Ultra-15 Centrifugal Filter Unit with a MWCO of 3 kDa (#UFC900324, Merck Chemicals and Life Science AB, Solna, Sweden). sEVs were isolated by size-exclusion chromatography (SEC) on qEVoriginal columns (Izon Science, Oxford, UK) as previously described [12]. Briefly, the samples were added to the column and were eluted in 500 µL fractions by 0.22 µm filtered PBS. Fractions 6-10 were pooled and concentrated to about 500 µL using an Amicon® Ultra-4 Centrifugal Filter Unit (#UFC800324, Merck Chemicals and Life Science AB). The particle size and zeta potential were characterized by nanoparticle tracking analysis (NTA, Zetaview from particle Metrix). For these analyses the sEVs samples were diluted 1:1000 in filtered PBS. The sample was injected in 100 µL portions for three times and the cell temperature was maintained at 24°C. The size analysis was done on eleven positions in the sample cell while for the zeta potential a two-position measurement was performed.
2.3. Surface functionalization
The exposure of the active surface of the sensor to the air as well as the chemical processes during the fabrication decrease the quality of the surface oxide. Hence, prior to the antibody/receptor immobilization, the surface of the microchannels were cleaned using RCA1 (5:1:1 DI-Water: H2O2:NH4OH) cleaning solution at 88°C for 30 minutes. In this process, atomic layers of silicon dioxide on top are etched away by the diluted ammonium hydroxide and a fresh layer of silicon dioxide is created by the hydrogen peroxide solution. Supplementary Information S1 demonstrates the effect of the cleaning process on the recorded streaming current.
After the RCA1 cleaning process, the surface hydroxyl groups on the microchannel are activated leading to a larger negative surface charge density. For the antibody/receptor immobilization, we followed an optimized strategy as reported in our earlier work [8]. In brief, the surface of the microchannels were first coated with a thin layer of PLL-g-PEG-biotin (PPB) by flowing an aqueous solution (0.1 mg/mL) of PPB through the microchannels for 30 minutes. For the detection of streptavidin, PPB layer was directly used where biotin of PPB layer acts as the receptor (Figure 1a). For the control measurements, the surface of the sensor was covered by high concentration of streptavidin (500 nM) using PPB. Then different concentrations of streptavidin were injected into the devices to measure the background signal as the control.
In the case of membrane protein profiling of sEVs, CD9 tetraspanin was targeted. In that case, biotinylated antibodies for CD9 were conjugated to the PPB coated surface using avidin as a linker molecule. For the control measurements, mouse IgG1 isotype control antibody was used instead of the specific capture antibody, i.e. anti-CD9. The concentration of the capture/control antibody was 50 µg/mL in 1xPBS and was immobilized for 1 hour. Prior to the sensing measurements, the microchannels were treated with pluronic (synperonic) F108 solution for 20 minutes in order to suppress non-specific binding.
2.4. The measurement setup and the microchip
The experimental setup, shown in Figure 1b, consists of a high-pressure pure nitrogen gas capsule regulated by an Elveflow OB1 flow controller coupled with a thermal time of flight (TOF) flow sensor (Elveflow, MSF3) to measure the flowrate during the experiments. The working fluids (either the bio-sample or the analyte) are hydraulically pushed by the nitrogen gas into the system through PEEK tubing and microfluidic connections procured from Darwin Microfluidics, France. A continuous train of trapezoidal pressure pulses between 1.5 and 3 bar, having a pulse duration of 30 s was used in the method in order to perform a two-points streaming current measurement. The currents were measured using hollow Platinum tube electrodes placed at the inlet and outlet ports. The resulting streaming current pulses (ΔIs) were measured by a Keithley picoammeter (model no. 2636A) while the pressure pulses (ΔP) were recorded directly by the pressure regulator.
The measured streaming current (Is) by the electrodes was used to calculate the apparent zeta potential (ζ∗) of the microchannel surface using Eq. 1 [13]. where ηand ∈∈0 refer to the dynamic viscosity and permittivity of the measurement buffer, respectively, and L and A refer to the length and cross-sectional area of the microchannels, respectively. More details on the working principle could be found in our previous report [11].
In this study, we performed both real-time and endpoint measurements. The end-point measurements involved recording the initial baseline () and the final baseline after the injection of the sEVs (). The injection of sEVs was done in 1x PBS to maintain the physiological conditions for the sEVs, whereas both the baselines were measured in 0.1x PBS to reduce the electrostatic screening. A Labview program recorded the data, and a custom Python script was used to analyze them afterwards. The noise-sensitive components of the experimental setup were placed in a Faraday cage to minimize the exterior perturbation on the measurements. The heart of the assembled system is a custom-built manifold platform to mount the biosensor chip and to ensure a leak-proof interfacing with standard fluidic connectors even at high pressure (up to 6 bar). For this purpose, an octagonal-shaped PEEK block was machined, on the center of which the microchip was placed. Suitable holes are drilled in the block with the same diameter as the inlet/outlet ports on the biosensor. In order to eliminate the leakage from the system, silicone rubber o-rings, purchased from Apple Rubber Inc., the USA, with 1 mm inner diameter were placed between the biosensor and the platform, and finally two plastic plates were used to sandwich the microchip on the platform. An optical window was designed on the plastic holders in order to perform fluorescence microscopy. Figure 1c shows the cross-sectional view of the chip platform along with the microchip. The silicon and glass substrates were purchased from MicroChemicals Company, Germany. Other required material and devices for the microfabrication were provided by Myfab, the Swedish research infrastructure for microtechnology, nanoscience and materials research.
3. Results
The devices are based on silicon-glass microfluidics where the microchannels of rectangular cross section were etched on Si substrate using reactive ion etching followed by an anodic bonding with thin quartz glass for sealing the channels. Typical dimensions of the fabricated channels are 10 µm × 25 µm in cross section and 3 mm in length. A schematic step by step procedures for the microchip processing and the final product are shown in Figure 2a and b, respectively. For the fabrication, a 100 nm thick layer of aluminum was sputtered on a 4 inch <100> single-side polished silicon wafer. This was followed by a coating of 500 nm thick layer of S1813 photoresist by spin coating and a soft baking at 110°C for 2 minutes. The sensor patterns were generated on the surface through the first lithography. After the development and hard baking of the resist at 120 °C for 10 minutes, the thin aluminum mask was dry etched so that the sensor pattern appears on the substrate. Thereafter, a 10 µm deep reactive ion etching of the silicon wafer was performed that forms the general geometry of the sensor (Figure 2a-1). After washing the aluminum mask off the substrate by wet chemistry, a new 300 nm thick protective aluminum layer was deposited on both sides of the silicon wafer. To create the inlet and outlet ports, the silicon wafer must be etched through. The bottom aluminum protective layer prevents helium leakage during the dry etching in the DRIE machine. A second photomask was used to pattern the inlet and outlet ports on the wafer. Thereafter, a dry etching of the aluminum protective layer (Figure 2a-2) followed by a deep dry etching of the silicon wafer was done (Figure 2a-3). After washing off the protective layers on the silicon wafer using wet chemistry, the substrate underwent a standard RCA surface preparation. Finally, a 300 nm thick layer of dry silicon dioxide was thermally grown on the substrate acting as the active sensing surface.
As the last step of the microfabrication, the substrate underwent a piranha cleaning process prior to the anodic bonding with a borofloat glass wafer (Figure 2a-4). The final product was diced using a high-speed rotary saw into 12 mm × 12 mm devices for the experiments. More details of the fabrication process flow could be found in the supplementary information S2.
Figure 2b shows a fully processed wafer containing 34 chips (left) and also finger-tip size single microchip (12×12 mm) containing four identical channels with a common inlet (at the center) and separate outlets. The geometry of the microchip and surface roughness of the fabricated devices were characterized using white light interferometry (WLI) method. For this purpose, an unbonded device was cleaned using RCA 1 cleaning solution and the structure was characterized by ZYGO optical profiler (Nexview NX2). Furthermore, the roughness elements were analyzed individually at different locations. Figure 2c shows the roughness of the surface. A total number of about 340K data points were analyzed and the centerline average was 9 nm.
3.1. Electrical and fluidic characterization
To shed light on the noise characteristics of the fabricated biosensor, streaming current was measured by flowing 0.1xPBS through a set of cleaned microchannels at different upstream pressures ranging between 1 and 6 bar. The RMS noise was calculated and is presented in Figure 3a as bar plots for different upstream pressures. Clearly, the RMS noise increases with increasing applied pressure and streaming current. The distribution of noise is shown in the inset of Figure 3a. As seen, the RMS noise remains similar (7 pA with standard deviation below 2 pA, n=3) up to 3 bars of applied pressure but then sharply increases, reaching 21 pA at 6 bars. The noise RMS at 3 bar corresponds to 0.42 mV of zeta potential which we defined as the minimum detectable signal (MDS) for the microchip sensor. Figure 3b shows the volumetric flowrate and streaming current measured in a cleaned microchannel as a function of the applied pressure. As expected, both the current and the flow rate show a linear and highly reproducible dependence on the applied pressure. The linear relation between upstream pressure and the measured flowrate with negligible deviation clearly indicates that both the microchip and the chip manifold are leak-proof within the measured pressure range. A simple Poiseuille estimation of the flowrate is also shown in Figure 3b which demonstrates a negligible deviation of the experimental flowrate from the theoretical predictions. The error bars correspond to 3 measurements on different microchips with the same production method.
In order to characterize the flow regime in the microchannel, the Reynolds number was calculated for different applied upstream pressures ranging from 1 to 6 bar (see section S3). Decreasing the size of the microchannels and consequently increasing the surface to volume ratio projects the wall effect on the flow characteristics [14]. A proper model for friction factor calculation could help compare the experimental data and the theoretical predictions to estimate the wall effects. The pressure drop estimation is also an important parameter in fluidic characterization of the system. Therefore, different fluidic components were analyzed, and the estimated pressure drops were calculated (see section S4).
Eventually, a stable and identical flow rate through all the microchannels need to be established as the devices are design for multiplex sensing. For the purpose of this study, a flow sensor was placed at the common inlet and the flowrate was measured at 3 bars. Then, channel 1 (C1) was blocked using microfluidic blockers at the outlet of the chip manifold and the flowrate was measured. The same process was repeated by blocking other two channels consecutively. Since the channels are expected to operate in an identical way, a negligible difference in flowrate is required when the sample is injected through a common inlet. The flowrate measurement data at each microchannel is shown in the section S5. The standard deviation of the flowrates at different microchannels is 96 nL/min while the average flowrate is 13 µL/min at 3 bar of upstream pressure. Blocking microchannels and measuring the flowrate at the others would show a clear picture of the fluidic crosstalk valuable for future multiplex biosensing. In addition, it is vital to know that a possible blockage in one channel during the experiments would not lead to the failure and/or deviation of the sensing results in the other microchannels. The standard deviation of the flowrate at different microchannels demonstrate a negligible fluidic crosstalk and effect of the blockage in other channels.
3.2. Sensing performance
The sensing performances of the devices were first investigated using a well-studied protein-ligand pair, i.e. biotin-streptavidin. Figure 4a shows a set of real-time measurements for two concentrations of streptavidin injected to PPB coated microchannels. For the detection measurement, the zeta potential of the PPB coated surface was first measured (baseline) by injecting 0.1x PBS, until a stable baseline was obtained. Thereafter, streptavidin dissolved in 0.1x PBS was injected and the real-time data was recorded for 1 hour of injection. The results are shown in Figure 4a. As can be seen, the microchip surface reaches an equilibrium in less than 20 minutes for higher concentrations while it takes over 40 minutes at a lower concentration. In order to compare the results with previously used commercial microcapillary-based sensor, a set of end-point streptavidin measurements under identical conditions were performed for both microchip and capillary. For this purpose, a 4 cm long capillary tube with an inner diameter of 25 µm was used. The microcapillaries were cleaned and functionalized using an identical procedure as described earlier. Figure 4b shows a comparison between the signals obtained with a microchip and a microcapillary based system. The measurements were performed for different concentrations of streptavidin ranging from 10 pM to 100 nM in order to evaluate the limit of detection in the two cases. As seen, the signal, measured in terms of apparent zeta potential, proportionally increases from 1.72 mV to 10.25 mV for the microchip. At all of the concentrations, the microchip-based sensor demonstrated higher sensitivity with an enhanced signal over the capillary-based sensor. In addition, comparing the error bars of two systems, microchip-based sensing seems more reproducible in comparison to capillary-based sensing. To further analyze the data and to demonstrate another potential advantage of having a transparent window in the channel design, the fluorescence signal arising from the Atto 565 conjugated streptavidin binding to the channels were also recorded. The images are shown in Figure 4c. As can be seen, the fluorescence intensity scales with the concentrations as expected. In addition, the control channel shows no/negligible fluorescence signal indicating a good specificity of the detection. For comparison of the detection sensitivity between the microchip and capillary-based sensing, we estimated the limit of detection (LOD) by a linear fit of the concentration vs signal (calibration) curve thus obtained (Figure 4b and supplementary information S6). The LOD was determined as the concentration of the target corresponding to the MDS of the sensor. The MDS of the microcapillary and microchips are 0.2 mV and 0.4 mV, respectively. The MDS of the microchip is slightly higher as compared to microcapillary. This observation is because of having four interconnected microchannels in microchips while for the microcapillary, a two-microchannel configuration was used. This observation is inline with the previous work of our group [10]. The LOD estimated for the microchip and microcapillary based systems was 0.5 nM and 0.8 nM, respectively. The indicated data points at the lowest concentration were out of the dynamic range of the sensor that are not considered for LOD calculation.
Beyond protein detection, streaming current-based sensor has also been investigated for the detection and surface protein profiling of sEVs [5, 10], an emerging bio-analyte which is attracting a significantly growing interest both for diagnostics and therapeutics [15]. A major advantage of streaming-current based sensor for sEVs detection arises from its size-dependent sensitivity, as demonstrated in our previous reports [11]. sEVs (diameter: 50-150 nm) being much larger than proteins can be detected with a fM range detection sensitivity using the streaming current-based technique [9]. In order to investigate the performance of the current microchip-based platform, we analyzed the detection sensitivity of sEVs derived from H1975 lung cancer cell line. The sEVs were first characterized using a nanoparticle tracking analyzer (NTA, Zetaview from particle Matrix) for their size and zeta potential measurement as presented in Figure 4d, e. The mean diameter of the sEVs as well as their zeta potential were measured to be 132 nm and -25.83 mV, respectively. The narrow distribution of the sEVs zeta potential demonstrates the consistent signal after the detection of the sEVs across the experiments. The sEVs were then detected by targeting their CD9 transmembrane protein using the microchip. Figure 4f shows the barplots representing the signal as a function of different concentration of sEVs. A calibration plot along with the MDS is shown in the supplementary information (S7). The LOD obtained in this case was 1.2×106 particles/mL, which is about 4 times lower than the results previously obtained with the capillary based system [9]. Clearly, both for streptavidin and sEVs detection, the microchip-based platform significantly outperforms the previously reported capillary-based system.
4. Discussion
The main contribution of the present study is the development of a microchip technology for the streaming current-based detection of biological analytes. While the detection method has been widely studied [2, 5, 9] and innovated during the past years, most of the research has been focused on the fundamental principle and exploitation/optimization of the method aiming to achieve a better sensitivity and specificity [9]. These studies have been done mainly using PDMS microchannel and/or commercial silica microcapillaries. Nevertheless, these research efforts have considerably contributed in the improvement of the method, thereby, extending the applicability of the method for the analysis of various biological targets, e.g., proteins [2], DNA [16] and sEVs [5], as well as exploring different diagnostic opportunities, e.g., the monitoring the efficacy of precision cancer medicine [10], albeit in a laboratory setting. Translation of these technical developments for potential use in clinical setting is an obvious next step which is the primary objective of the research presented here. In this context, the development of a robust and easy-to-use setup is a key requirement along with the development of a microchip that can be mass fabricated. Going further, the entire platform has been characterized for electrical noise, fluidic and electrokinetic characteristics as well as sensitivity towards some common biological targets. In addition, by comparing with a previously used commercial silica capillary, a clear improvement in detection sensitivity has been demonstrated. Some of these aspects are further discussed in the following section. To the best of our knowledge, such a platform has not been reported until now and thereby constitutes the main novelty of the current work.
The microchip and the detection platform
The fabricated microchips clearly show high reliability in the entire pressure range. This is evident from both the flow rate analysis of different chips and micro-channels (Fig 3b and S6) and the streaming current measurement as a function of the applied pressure (Fig 3b). The measurement of streaming current as a function of pressure shows a highly linear behavior with a standard deviation ∼ ±90 pA in the entire pressure range. Fig. 3b also indicates leak-free integration of the chip with the manifold allowing easy interfacing with the commercial microfluidic connectors. The choice of Si-glass microfluidics offers several other benefits, particularly in the context of streaming current based bio-detection. As reported earlier [9, 17], surface roughness can negatively impact the sensitivity of detection. With the optimization of the Si fabrication process, the roughness can be easily controlled. As shown, in Fig. 2c, the roughness of the fabricated channels stands at a mean value of 9 nm which is comparable to many antibodies. Secondly, the geometry can be precisely controlled with nanoscale precision and the anodically-bonded devices offers unparallel chemical compatibility and bonding strength [18]. Finally, SiO2 surface has been widely studied for the immobilization of biological recognition molecules [19, 20] and there are well established protocols which allows highly reproducible surface functionalization.
Electrical and fluidic characteristics
There is a variety of sources introducing noise in the recorded data such as flow disturbance, electrical measurement system, ambient noise, microscopic fluctuations and impurities, and intrinsic sources. The conductive components of the experimental setup are accommodated inside a Faraday cage and the interior walls of the cage are covered by insulating tapes to avoid short circuit and electromagnetic noise. The RMS of the noise at different upstream pressures recommends low applied pressure for the measurements (between 1 and 3 bar). On the other hand, the two-point streaming current measurement requires two upstream pressures to form the desired waveform. For the upper value of the pressure 3 bar was nominated. However, for the lower value, 1.5 bar was chosen instead of 1 bar to ensure higher data sampling rate. Therefore, the RMS noise stays roughly the same for the range of the experiments. In addition, the low absolute value of streaming current leads to a negligible polarization of the electrodes as demonstrated above by the linear relation between the upstream pressure and the streaming current in the working range of the device.
The fluidic analysis on the other hand shows that the friction factor inside the microchip agrees with the analytical predictions. The pressure drop in the fluidic path due to multiple bends in the microchip manifold, different microfluidic connectors, inlet effects, and the electrovisous effect are studied for the current system (supplementary information S4) [11, 12].
Sensing performance
The sensitivity of the fabricated devices has been evaluated by detecting streptavidin at two and five concentrations for real-time and endpoint measurements, respectively. The results presented in Figure 4a and b, show that the microchip biosensor is suitable for both real-time and endpoint measurements in a broad range (10 pM to 100 nM). The LOD of the developed technique in this study was compared with the commercial silica microcapillaries to show the improvement over our previous data [9]. The calibration curves for both systems show 60% improvement in the limit of detection for the microchip biosensor. In addition, the measurement uncertainty (standard deviation, n=3) for the microchip biosensors are significantly lower as compared to that of the capillary biosensors. On the other hand, the transparent microchips along with the chip manifold facilitates the optical analysis of the target. The optical window on the chip manifold and its appropriate design for the optical microscope sample holder makes it possible to optically analyze the targets in the microchannel. The sensing performance of the platform was also evaluated with the profiling of sEV membrane proteins. As mentioned earlier, an improvement in the detection sensitivity by a factor of 4 was also observed, when compared to commercial microcapillary. The improvement can be attributed to a better surface quality of the thermally grown SiO2 surface. As demonstrated in our previous report, surface charge density and electrostatic charge contrast play a critical role in determining the detection sensitivity. In order to compare the surface charge densities, we cleaned a microchip and a capillary sensor using the protocol explained earlier. Then measured the baseline. Thereafter, the same process was repeated after PPB and CD9 antibody functionalization.
Better surface coverage by the PPB layer would make a more positive surface that can attract more negatively charged targets (streptavidin or sEVs). Therefore, on one hand having a negatively charged surface at the pH of the buffer used for the experiments could enhance the surface coverage by the PPB layer and on the other hand, a too negative surface could repel the negatively charged biomolecules such as the sEVs. To demonstrate the mentioned tradeoff, the surface charge density of the microchip is calculated and compared with the microcapillaries here.
The 1-dimentional solution of the Poisson-Boltzmann equation that governs the length of the EDL on a charged surface, shows the relation between surface electrostatic potential and the surface charge density. The effective surface charge density, on the other hand, could be calculated from the zeta potential of the surface (Eq. (2)) [23].
A better surface quality attracts more capture probes on the surface and makes a more positively charged surface that can act in favor of attracting more negatively charged targets i.e., streptavidin and sEVs. The effective surface charge density of the bare capillary and the microchip are -5.3×10−3 and -5.6×10−3 e/nm2, respectively, showing higher density of the charges on the bare surface of the microchip as compared to the capillary. After PPB functionalization, the effective charge density is -0.08×10−3 e/nm2 for the microchip while it is -0.35×10−3 e/nm2 for the capillary sensors. for the capillary sensors. These numbers after the immobilization of CD9 antibody are -0.45×10−3 e/nm2 and -0.9×10−3 e/nm2 [9] for microchip and capillary, respectively. This indicates a higher surface adsorption of the positively charged PPB in case of the microchip, making the overall surface charge density less negative. This better facilitates the sensing of the negatively charged sEVs and could explain the lower limit of detection discussed before. The limit of detection for streptavidin and sEVs by CD9 surface protein were 60% and 4 time better than that of the microcapillaries when compared under identical conditions, respectively. Finally, the devices were cleaned using the RCA1 cleaning protocol. The results (supporting information S8) show that the baseline could be recovered up to five times after the experiments promising a sustainable detection platform.
5. Conclusion
In this study, a microfluidic device was fabricated out of silicon and was covered by a thin layer of silicon dioxide as the active surface of the biosensor. The hydroxyl groups on the surface carrying negative charge were activated by wet chemistry to attract Ploy-L-Lysine molecules conjugated with PEG and biotin on the surface acting as the capture probes for the biosensing. A robust experimental setup was designed and assembled to facilitate electrical measurements as well as a leak-proof fluidic path for the experiments. The zeta potential of the microchannel inner surface was calculated using streaming current measurements during the experiments. The zeta potential of the surface before and after capturing a specific target is the sensing signal in this study. The measured noise of the fabricated biosensor was characterized by calculating root mean square at different upstream pressures from 1 to 6 bar. Upstream pressures of 1.5 and 3 bar were chosen as the working condition for the experiments. The linear relation between the upstream pressure and the measured streaming current/flowrate was controlled as well as the fluidic crosstalk between the microchannels. Thereafter, the performance of the microchip-based biosensor was compared with commercial microcapillary tubes. The sensitivity of the microchip device was one order of magnitude higher, and they showed reusability after the experiments. Finally, CD9 surface biomarker of the extracellular vesicles from H1975 cell line were captured specifically by the capture probes immobilized on the surface. The calibration curves show a 1.2×106 particle/mL detection limit for the microchip biosensor.
In conclusion, we could demonstrate the successful replacement of microcapillary tubes with microchip-based biosensors working with streaming current measurement. The results showed a faster, more sensitive, more robust, and reusable substitution to the commercial microcapillary tubes for biosensing applications.
Conflict of interest
The authors declare no conflict of interest.
Acknowledgement
This study was supported by grants from the Swedish Research Council (grant no. 2016-05051), the Erling Persson Family Foundation, Stockholm Cancer Society (#171123, #191293, and #201202), the Swedish Cancer Society (CAN 2015/401; CAN 2018/597), Stockholm County Council (#20160287 and #20180404), and funds of Karolinska University Hospital FOUU (#75032). Dr. Örjan Vallin and Dr. Milena De Albuquerque Moreira are acknowledged for their input in the microfabrication steps.