1 Abstract
Serial block face scanning electron microscopy (SBFSEM) is an increasingly popular method for investigating the three-dimensional ultrastructure of large biological samples. Prior to imaging, samples are typically chemically fixed, stained with osmium and uranyl acetate, and subsequently embedded in resin. The purpose of staining is to provide image contrast and reduce specimen charging under the electron beam, which is detrimental to the quality of imaging. Obtaining, using, and disposing of uranyl acetate is getting increasingly cumbersome in many countries due to new regulations on the handling of radioactive substances. Therefore, we developed an alternative staining procedure that does not rely on the use of uranium or any other radioactive substance. This procedure provides excellent contrast and efficiently reduces specimen charging.