ABSTRACT
Objective There is a need for tools enabling efficient evaluation of amyloid- and tau-PET images suited for both clinical and research settings. The purpose of this study was to assess and validate a semi-automated imaging workflow, called Biomarker Localization, Analysis, Visualization, Extraction, and Registration (BLAzER). We tested BLAzER using two different segmentation platforms, FreeSurfer (FS) and Neuroreader (NR), for regional brain PET quantification in images from participants in the Alzheimer’s Disease Neuroimaging Initiative (ADNI) dataset.
Methods 127 amyloid-PET and 55 tau-PET studies along with corresponding volumetric MRI were obtained from ADNI. The BLAzER workflow utilizes segmentation of MR images by FS or NR, then visualizes and quantifies regional brain PET data using FDA-cleared software (MIM), enabling quality control to ensure optimal registration and detect segmentation errors.
Results BLAzER analysis required only ∼5 min plus segmentation time. BLAzER using FS segmentation showed strong agreement with ADNI for global amyloid-PET standardized uptake value ratios (SUVRs) (r = 0.9922, p < 0.001) and regional tau-PET SUVRs across all Braak staging regions (r > 0.97, p < 0.001) with high inter-operator reproducibility for both (ICC > 0.97) and nearly identical dichotomization as amyloid-positive or -negative (2 discrepant cases out of 127). Comparing FS vs. NR segmentation with BLAzER, the global SUVRs were strongly correlated for global amyloid-PET (r = 0.9841, p < 0.001), but were systematically higher (4% on average) with NR, likely due to more inclusion of white matter, which has high florbetapir binding.
Conclusions BLAzER provides an efficient workflow for regional brain PET quantification. FDA-cleared components and the ability to visualize registration reduce barriers between research and clinical applications.
INTRODUCTION
Positron emission tomography (PET) neuroimaging applications have increased in both research and clinical setting in recent years. PET provides the ability to study functional and molecular processes in the brain in vivo, allowing exploration of an array of normal and pathological states, including neurodegenerative disorders, psychiatric conditions, and neurooncology. Quantitation of PET data is well-established for research applications but less so in routine clinical settings. Recent research demonstrates the utility of quantification to supplement visual assessment for clinical PET, especially in Alzheimer’s disease (AD) using the glucose analogue 2-deoxy-2-[18F]fluoro-D-glucose (FDG) and amyloid PET tracers [1-4]. Quantification of targeted brain regions of interest (ROIs) [5, 6] is enabled using masks generated by automated magnetic resonance imaging (MRI) segmentation algorithms, such as the widely used reference standard in research, FreeSurfer (FS) [7-9], or the FDA-cleared and ISO-certified Neuroreader (NR) [10-13]. By registering the segmentation mask with the PET scan, ROIs can be defined for PET measurements without laborious and potentially imprecise or biased manual delineation methods [6, 14]. Barriers to the use of this approach to quantitation of brain PET, particularly for clinical applications, include computationally intensive software, time-consuming workflows, the need for real-time visualization of the primary and processed data for quality control, and limited availability of FDA-cleared software suitable for this purpose.
Accuracy and precision are key features of effective image analysis methods to ensure that the extracted biomarker data can provide reliable in vivo information of the physiological state or disease process of interest. However, efficiency, ease of use, and availability are also essential features for widespread and routine implementation of quantitative brain PET. Many current techniques require lengthy post-processing time [2, 15] and/or heavy computational workload [16, 17]. Incorrect segmentation of the brain MRI data and misregistration between PET and segmented brain regions are potential sources of error when extracting of regional PET data based on ROIs defined through fully automated segmentation [15]. Additionally, although qualitative and quantitative assessment can be performed separately, the ability to easily visualize images in real-time at different states of processing and registration can provide an important degree of quality control and allow for user input to correct problematic studies. Many of the current fully automated processing workflows that involve only inputs and outputs to optimize speed, can be liable to unidentified errors because there is no opportunity to visualize the processed images [18, 19]. Finally, other barriers to routine implementation of quantitative brain PET include the lack of versatility in existing tools to analyze images acquired or processed by different platforms, radiotracers, and segmentation algorithms. Radiotracers commonly used for brain PET analysis in AD include [11C]Pittsburgh compound B (PiB) [20], [18F]florbetapir [21, 22], [18F]florbetaben [23], and [18F]flutemetamol [24] for amyloid-β (Aβ), [18F]flortaucipir for tau [25-28], and [18F]FDG [29, 30] for glucose metabolism. Multicenter clinical trials and longitudinal studies can benefit from a versatile workflow to standardize PET quantification [31].
Although a number of effective research workflows exist for regional brain PET analysis that combine one or more of the previously discussed features [8, 19, 32-35], quantification still has not been widely implemented in routine clinical brain PET for AD, which relies primarily on visual assessment and components validated for routine clinical use. FDA clearance and ISO-certification are key features of not only NR, but also MIM, software that provides image visualization and quantification capabilities. Combining features of such tools could help lower the barriers between clinical and research workflows. However, to the best of our knowledge, no workflow exists that is efficient and easy for physicians and technologists to use, automatically segments ROIs in a customizable fashion, consists of FDA-cleared components, and allows for both qualitative and quantitative assessment for brain PET data in a clinical environment.
We present here a novel brain Biomarker Localization, Analysis, Extraction, and Registration (BLAzER) workflow for analysis of PET based on segmented brain MRI. We demonstrate that BLAzER works well for both global and regional PET quantification of brain amyloid- and tau-PET, respectively, in Alzheimer’s Disease Neuroimaging Initiative (ADNI) participants spanning normal cognition, mild cognitive impairment (MCI), and AD dementia. Although this report focuses on AD biomarkers, BLAzER can be applied to any PET neuroimaging study that utilizes MR-based segmentation for defining ROIs. Additionally, we compare two different inputs for segmentation, FS and NR, which could be used with BLAzER for research and clinical applications, respectively.
METHODS
Study Population
Images were downloaded from the Alzheimer’s Disease Neuroimaging Initiative (ADNI) database (http://www.loni.ucla.edu/ADNI/). ADNI was launched in 2003 as a public-private partnership, led by principal investigator Michael W. Weiner, MD. The primary goal of ADNI has been to test whether serial magnetic resonance imaging (MRI), PET, other biological markers, and clinical and neuropsychological assessment can be combined to measure the progression of mild cognitive impairment (MCI) and early AD. All studies were approved by local Institutional Review Board at each institution through ADNI. Additional information is available at the ADNI website (www.adni-info.org).
127 amyloid-PET and 55 tau-PET studies were selected from ADNI for a total of 178 unique subjects (4 overlap). Subjects were selected to represent the spectrum of ADNI participants in terms of cognitive status (cognitively normal (CN), mild cognitive impairment (MCI), and Alzheimer’s disease (AD)), and age (cohorts aged 55-59, 60–69, 70–79, and ≥80). Subject selection was performed prior to beginning image review or analysis and was based on the preceding demographic criteria alone to capture the broad range of neuropathology available in ADNI (Table 1). Selection was blinded to the actual quality of the scans and output of the data, and image analysis was performed blind to subjects’ cognitive status or age group. No exclusions were made at time of selection or subsequently during the analysis (i.e. all of the initially selected studies were included in the final analysis).
Image sets used for analysis
All subjects in both the amyloid-PET and tau-PET cohorts had a volumetric brain MRI scan performed within 1 year of the PET study. MRI studies were performed using standard T1-weighted sagittal 3D MPRAGE (magnetization prepared rapid gradient echo) sequences acquired from various 3T scanners with a 1.25 × 1.25-mm in-plane spatial resolution and 1.2-mm slice thickness with 256 x 256 voxel resolution according to ADNI specifications.
Amyloid- and tau-PET studies were acquired with [18F]florbetapir (also called AV-45) and [18F]flortaucipir (also called AV-1451), respectively. PET images were obtained on different scanners of varying resolutions, each with its platform-specific acquisition protocol. Therefore, as part of the ADNI protocol, all raw PET images undergo pre-processing for quality control and standardization purposes at the University of Michigan [36]. In summary, 4 × 5 min dynamic image frames, acquired 50 to 70 minutes post-injection, were co-registered to the first extracted frame of the raw image file. Then, the 6 five-minute frames were averaged to form a static PET image and reoriented into a standard 160 × 160 × 96 voxel image grid with 1.5-mm cubic voxels. Finally, each image set was filtered with a scanner-specific filter function to produce images of uniform isotropic resolution of 8 mm FWHM, which is the approximate resolution of the lowest resolution scanners used in ADNI. Only the fully pre-processed, standardized, co-registered, and averaged PET images were used for this study. Further details on the ADNI acquisition protocol are available on the website (http://adni.loni.usc.edu/methods).
Image Analysis
MRI Segmentation
Volumetric MRI images were segmented by either FreeSurfer v6.0.0 (Boston, MA) or Neuroreader (Brainreader, Horsens, Denmark). FS, an open-source software, has been validated as a tool to measure brain volumes in various neurological diseases when compared to either manual delineation [6] or other algorithms [14]. FS uses a complex algorithm with a series of normalization and motion-correction approaches prior to even starting its intensity-based approach to parcellate the brain into not only the larger cortical structures, but also subregions to allow targeted measurement of specific ROIs [8, 9]. FS segmentation produces detailed ROIs, which allow for spatial delineation of brain regions beyond the main cortical and subcortical brain structures. Customization of brain regions is particularly useful in analysis of tau-PET studies in Alzheimer’s disease where pathology follows characteristic Braak staging regions [37].
NR’s separation of gray matter (GM), white matter (WM), and cerebral spinal fluid (CSF) uses an intensity-based approach along with a template, similar to other available approaches [38, 39] such as the widely used Automatic Anatomic Labeling (AAL) template [19]. Similar to AAL, NR is highly efficient, with segmentation of an individual case processed in approximately 15 minutes compared to 8 to 12 hours for FS. However, segmented brains from NR provide larger cortical and subcortical brain regions compared to FS, limiting customization beyond anatomical boundaries defined by their atlas.
FS studies were loaded on a supercomputer environment (Cheaha, Birmingham, AL) in order to run all 184 volumetric MRI input files simultaneously through parallel computing. Parcellated, segmented brain images were converted from MGZ to DICOM format using 3DSlicer v4.6 (Boston, MA). Alternatively, NR automatically produced segmented brains in DICOM format for the same MPRAGE scans. Each of the segmentation output files along with source volumetric MRI and pre-processed PET scans were then visualized and quantified using an automated workflow on a multi-modal imaging software – MIM v6.6.13 (MIM Software Inc., Cleveland, OH).
PET Analysis
MIM is a commercial, FDA-cleared software package designed to help researchers and clinicians quantitatively and qualitatively process multi-modal imaging data. Through a graphical user interface (GUI), MIM allows the user to design customized, automated workflows for processing cases without need for advanced computational skills. Workflows were developed that could use segmentation data from FS or NR as inputs for ROI definition on the PET data sets. For BLAzER, we selected the segmented brain dataset, brain PET, and volumetric brain MRI images for analysis, then ran a user-defined and customizable workflow that registered the images, defined ROIs within the original volumetric MR and PET based on the segmentation, and delivered the quantification in an array of potential outputs while allowing for visual inspection. Workflows were generated and customized to an individual segmentation method and PET radiotracer through a GUI interface where steps in the protocol were simply selected from a drop-down list similar to Porcupine [40], such that users were not limited by their computational skillset to analyze large, multi-modal image datasets. Once the original protocol had been generated, anatomical tags for brain regions from another segmentation method or radiotracer were incorporated by replacing with the new identifiers with minimal user input, allowing clinicians and researchers alike to implement their customized analyses through BLAzER with minimal training.
In summary, in the BLAzER workflow, we 1) segmented the MRI to generate a 3D brain mask using either FS or NR; 2) transferred the segmentation to MIM and selected corresponding volumetric MRI and PET to analyze; 3) automatically co-registered segmented brain mask to volumetric MRI; 4) utilized the segmented brain to delineate the various brain regions based on the pixel intensities for each region defined by FS or NR; 5) performed quality control through visualization and corrected registration, if necessary; and 6) fused PET scan to MRI/brain mask template to extract, visualize, and quantify data (Fig. 1). The workflow suspended the automated process before transferring the contours to allow the user to verify accurate image registration. The original MPRAGE was used as a quality control measure to assure that the rigid registration method used to align the images had been performed properly and the ROIs are correct. If needed, the user could manually correct the registration and/or ROIs at this step. This review step was employed for all cases but is optional and could be omitted to provide a more automated process.
Anatomical Definitions of Brain Regions
In BLAzER, all SUVRs extracted are automatically weighted by the volumes of the individual subregions comprising each ROI. Amyloid-PET ROIs were normalized to the entire cerebellum to extract SUVRs for all regions that comprised the cerebral cortex: frontal lobe, temporal lobe, parietal lobe, and cingulate (BLAzER-FS, as defined by ADNI) or frontal, temporal, parietal, and occipital lobes (BLAzER-NR, as defined by Neuroreader). BLAzER-FS matched ADNI-defined cortical regions exactly while BLAzER-NR included the entire cortex. For head-to-head comparisons between BLAzER and ADNI, BLAzER-FS delineation was based on ADNI-defined brain subregions. However, for comparisons between BLAzER-FS and BLAzER-NR, BLAzER-FS anatomical regions were modified to match the NR-based regions of entire cortex. We denote this distinction as BLAzER-FS*. Specific FreeSurfer subregions are detailed below as well as listed in Supplementary Table 1.
Specifically, BLAzER-FS subregions were defined as follows: caudal middle frontal, lateral orbitofrontal, medial orbitofrontal, pars opercularis, pars orbitalis, pars triangularis, rostral middle frontal, superior frontal, and frontal pole (frontal); caudal anterior cingulate, isthmus cingulate, posterior cingulate, and rostral anterior cingulate (cingulate); inferior parietal, precuneus, superior parietal, and supramarginal (parietal); middle temporal and superior temporal (temporal). In contrast, BLAzER-FS* regions were defined as follows: caudal middle frontal, lateral orbitofrontal, medial orbitofrontal, pars opercularis, pars orbitalis, pars triangularis, rostral middle frontal, superior frontal, frontal pole, paracentral gyrus, precentral gyrus, caudal anterior cingulate, posterior cingulate, and rostral anterior cingulate, and insula (frontal); inferior parietal, precuneus, superior parietal, supramarginal, postcentral gyrus, and isthmus cingulate (parietal); bankssts, entorhinal, fusiform, inferior temporal, middle temporal, parahippocampal, superior temporal, temporal pole, and transverse temporal (temporal); cuneus, lateral occipital, lingual, and pericalcarine (occipital).
Tau-PET SUVRs were calculated similarly to the amyloid-PET defined ROIs with two main differences. Tau-PET was normalized to the cerebellar gray matter instead of the entire cerebellum [37] and ADNI-defined ROIs for BLAzER-FS workflow followed the pathologic Braak staging regions [41]: entorhinal cortex (Braak 1); hippocampus (Braak 2); parahippocampal, fusiform, lingual, and amygdala (Braak 3); middle temporal, caudal anterior cingulate, rostral anterior cingulate, posterior cingulate, isthmus cingulate, insula, inferior temporal, and temporal pole (Braak 4); superior frontal, lateral orbitofrontal, medial orbitofrontal, frontal pole, caudal middle frontal, rostral middle frontal, pars opercularis, pars orbitalis, pars triangularis, caudate, putamen, lateral occipital, parietal supramarginal, inferior parietal, superior temporal, pallidum, superior parietal, precuneus, superior temporal sulcus, nucleus accumbens, and transverse temporal (Braak 5); pericalcarine, postcentral, cuneus, precentral, and paracentral (Braak 6). As NR’s anatomical atlas only includes major cortical and subcortical regions, we defined BLAzER-NR and BLAzER-FS* tau-PET regions the same as for amyloid-PET: frontal, temporal, parietal, and occipital cortices.
The published ADNI data was used as the reference standard. The reported values in the ADNI data set were based on brain segmentations performed using FS v5.3.0. SUV and SUVR data was extracted using SPM5 (version) after co-registeristration of PET and MR data. Key differences between BLAzER and the ADNI method are summarized in Table 2.
Statistical Analysis
Statistical analyses were performed using MedCalc v17.7.2 (MedCalc Software, Mariakerke, Belgium) and Matlab vR2016b (MathWorks, Natick, MA) to compare the BLAzER method with the reference standard from ADNI. SUVRs and volumes were reported as mean ± SD. Statistical significance was defined as p<0.002 to account for multiple comparisons. Univariate regression was used to validate measurements between the two methods using Pearson’s correlation coefficient (poor agreement=0; slight=0.01-0.20; fair=0.21-0.40; moderate=0.41-0.60; good=0.61-0.80, and excellent=0.81-1.00 agreement). Similar comparisons were conducted using Bland Altman to determine limits of agreement and percent differences between the two analysis tools. Intra- and inter-observer variability and agreement were evaluated for global cortical SUVR using the intraclass correlation coefficient (ICC) in a two-way random model (ICC<0.40 = poor; ICC≥0.40 to 0.75 = fair to good; ICC>0.75 = excellent agreement). Finally, amyloid-PET cases were dichotomized into positive or negative to determine the ability of BLAzER to properly classify cases based on SUVR cutoffs. Dichotomization was performed by 1) directly applying ADNI’s autopsy-derived 1.11 cutoff [34, 42] and 2) deriving BLAzER-specific cutoff by using a linear regression to convert “ADNI units” to BLAzER units based on the slope and y-intercept.
RESULTS
Efficiency, quality control, and reproducibility
The BLAzER workflow enabled rapid image analysis. The most time-consuming step was segmentation. NR was considerably faster than FS on a per-case basis (10–20 min/case vs. 8–12 hours/case, respectively), yet slower than FS for total processing time for the 182 subject cohort (45.5 hours vs. 12 hours, respectively) due to the ability to run FS cases in parallel on a supercomputer environment. After segmentation, FS cases required ∼2 min for manual DICOM conversion whereas NR cases were returned in DICOM format. Once the segementation was obtained, the remainder of BLAzER processing took about 5 min per case. Thus, a case could be fully processed, from completion of image acquisition to full regional quantification, in as little as 20 min (when using NR segmentation).
Although the workflow could operate with full automation, the ability to visualize registration provided for quality control and avoiding registration errors. We routinely ran MIM’s “Run Rigid Assisted Alignment” tool serially to fix minor errors until it provided no further adjustment, which took only a few seconds (Fig. 1). Only 1 of the 182 scans required manual correction of registration, which was a case with severe brain atrophy. This highlights the usefulness of the visualization step, however, as the incorrect registration would have been missed in a fully automated workflow. We also examined inter-rater reliability by having two independent operators process each scan. Global florbetapir- and regional flortaucipir-PET measurements showed excellent reproducibility between two users across all brain regions (ICC > 0.97, Table 3).
MRI Volumetric Measurements
As BLAzER’s regional PET extraction depends on accurate MRI-based anatomic segmentation and registration, we first compared BLAzER measurements of regional MRI volumes to the ADNI reference standard. Global cortical volume measured by the BLAzER workflow was highly correlated to that measured by ADNI (r = 0.9749, p < 0.001, Fig. 2A) with small systematic difference (1.61%) and tight 95% confidence interval (CI) (Fig. 2B). Regional comparisons across the frontal, cingulate, parietal, and temporal lobes for CN, MCI, and AD subjects showed similar results (r > 0.92, p < 0.0001, Supplementary Table 2).
Amyloid-PET SUVR and Dichotomization
BLAzER showed strong agreement with the ADNI reference standard when measuring diffuse cortical binding of amyloid-PET. Global SUVRs were similar between BLAzER and ADNI (r = 0.9922, p < 0.0001, Fig. 3A). Additionally, the slope of linear regression and y-intercept (y = 1.0012x + 0.01816, R2 = 0.9844, Table 4) showed a near one-to-one correspondence between BLAzER and ADNI. There was a slight systematic difference in global SUVR, which was on average 1.6% higher with BLAzER than in the ADNI dataset (Fig. 3B). This may be to do a systematic reduction in the reference region mean, which was observed in the cerebellum. Regional comparisons across the frontal, cingulate, parietal, and temporal lobes for CN, MCI, and AD subjects showed similar results (r > 0.94, p < 0.0001, Supplementary Table 3).
We also the examined the dichotomous classification of individuals as amyloid-positive vs. –negative based on SUVR cutoffs. Because BLAzER SUVRs were slightly higher than ADNI on average, a few more of the 127 subjects were classified as amyloid-positive when BLAzER SUVRs were dichotomized using the ADNI cutoff of 1.11 with all 4 of these cases lying near the cutoff (BLAzER vs. ADNI range: [1.113 to 1.125] vs. [1.060 to 1.098], respectively, Table 4). However, translating 1.11 “ADNI units” into 1.13 “BLAzER units” by linear regression (Equations 1 and 2) fixed the dichotomization of these 4 cases but 2 different cases were discrepant out of the 127 total at this cutoff. These 2 cases had SURVs of 1.113 vs. 1.125 and 1.113 vs. 1.115 for BLAzER-FS vs. ADNI, respectively.
1.
2.SUVR BLAzER= SUVR ADNI * slope BLAzER + intercept BLAzER
Regional brain tau-PET comparison
BLAzER showed strong agreement with the ADNI reference standard when looking at regional binding of tau-PET with flortaucipir. Despite slightly lower correlation coefficients and wider 95% confidence intervals than with amyloid-PET (as expected because of the smaller regions defined for analysis), BLAzER correlated strongly with the ADNI reference standard across all six Braak stage regions (r > 0.97, p < 0.0001, Fig. 4). Further demonstrating regional accuracy, BLAzER showed tight 95% CI (<10%) and small differences (–1.5 to 3.6%) across all regions, (Fig. 5), and all cognitive statuses (r > 0.91, p < 0.0001, Supplementary Table 3).
Amyloid and tau PET quantitation based on NR segmentation
Next we compared the performance of a different segmentation platform, comparing BLAzER-FS with BLAzER-NR. To enable apples-to-apples comparison, we first adjusted FS’s anatomical subregions to match NR’s composite regions (FS*, see methods). BLAzER-NR highly correlated with BLAzER-FS* at a global level for amyloid-PET (r = 0.9841, p < 0.001, Fig. 6A). However, BLAzER-NR values had higher SUVR than BLAzER-FS* as shown by the systematic difference (4.0%) and shifted 95% CI (–0.7% to 8.7%, Fig. 6B). This is likely due to BLAzER-NR segmentation including more PET signal from white matter, which has high florbetapir binding, due to a slightly thicker definition of cortical regions with NR compared to FS. Dichotomization results paralleled these findings (Table 5). Directly applying ADNI’s cutoff of 1.11 classified led to 12 out of 127 cases being discrepant between BLAzER-NR and BLAzER-FS* (range of SUVRs: [1.092 to 1.180] vs. [1.049 to 1.111], respectively). After applying regression analysis for both BLAzER-NR to ADNI (y = 0.9845x + 0.0.05658, R2 = 0.9587) and BLAzER-FS to ADNI (y = 0.8947 × + 0.1158, R2 = 0.9482), the new SUVR cutoff obtained was 1.15 for BLAzER-NR while BLAzER-FS* remained unchanged. The adjusted cutoff reduced the number of discrepant cases to 7 out of 127 (NR vs. FS range: [1.0923 to 1.180] vs. [1.088 to 1.120], respectively). This discrepancy reflects the differences between the segmentation provided by NR and FS which in turn affects SUVRs.
Regional analysis for both amyloid- and tau-PET across different cognitive statuses and brain regions revealed excellent correlations between BLAzER-NR and BLAzER-FS* (r > 0.85, p < 0.0001, Supplementary Table 4) although not as high as the correlations between BLAzER-FS and the ANDI reference values, again reflecting differences in the segmentation between the NR and FS platforms.
DISCUSSION
Measurement of the regional brain distribution of PET tracers is a cornerstone of data analysis in molecular neuroimaging and is growing in importance for clinical applications, particularly in the evaluation of patients with cognitive impairment. A number of imaging processing workflows exist that perform quantification of regional brain PET data [8, 19, 32-35]. However, there are few if any analysis tools that perform well for both clinical and research applications, utilize FDA-cleared components, define ROIs and extract regional PET data automatically, provide both quantitative and real-time visual assessment, and work quickly and efficiently without need for advanced coding skills. Our evaluation of BLAzER with validation through comparison to the ADNI database and head-to-head comparison of two segmentation methods demonstrates an approach that addresses this unmet need.
Quantitative brain PET data from BLAzER correlated very closely with the results in ADNI for florbetapir- and flortaucipir-PET analyzed using FreeSurfer-defined regional volumes. We also demonstrated that the regional brain volumes provided by FS and NR were in close agreement with ROIs defined through BLAzER with less than 2% error between the original volumes and those output by BLAzER. BLAzER and ADNI produced nearly identical dichotomous classification of amyloid-PET participants as positive or negative, especially when using regression analysis to account for systematic differences. It is important to acknowledge that our analysis was completed before the most recent ADNI update, which includes a new partial volume corrected set of values based on modified reference regions. Thus, any partial volume effect (PVE), or inaccuracies that result from low PET spatial resolution [43], affected both BLAzER and our reference ADNI set similarly. In addition to PVE, off-target-binding in non-specific regions can be problematic for accurate PET quantification. Recent evidence has shown non-specific flortaucipir uptake in the basal ganglia [44, 45]. In the most recent ADNI updates, the caudate and putamen have been removed from the Braak 5 staging regions for flortaucipir. However, we kept our results consistent by comparing BLAzER to the original dataset that included the caudate and putamen. In future work, BLAzER-FS regions can be modified to account for PVE and off-target-binding if desired.
We observed slightly larger systematic differences and CIs for flortaucipir-PET compared to florbetapir-PET. As pathological cortical tau distribution is more regionally localized and restricted than pathological cortical amyloid which tends to be diffuse throughout the cerebral cortex [28], differences in segmentation and image alignment are expected to have a greater effect on PET measures of regional brain tau compared to regional brain amyloid.
Potential sources of the small differences between BLAzER and the ADNI data set include different FS versions, source volumetric MR images, and registration methods. For our study, FS v6.0.0 was used while v5.3.0 was used for the ADNI data set, which could affect anatomical volume measurements and brain segmentation which in turn would affect ROI definition for PET quantitation [46]. Additionally, the ADNI dataset used two volumetric MR scan input files for FS to produce the segmented brain while we chose to use only one volumetric brain MR to reflect our standard research and clinical workflows. Within ADNI, each subject undergoes two volumetric MR scans during a single scanning session with some subject datasets having slightly different imaging parameters. In order to align the PET and MR images, ADNI utilized SPM to register the images automatically while we used MIM. BLAzER utilizes a rigid-registration algorithm that allows only translation and rotation. Additionally, we utilize a three-step system of registration where the segmentation is first registered with its original, template volumetric MR and then with the PET after the rigid-registration quality control-check. We believe that this visual assessment and option for manual correction minimizes sources of error that can arise from fully automated registration methods (Fig. 1) [2]. Additionally, we believe that our results are generalizable across the ADNI population due to our selection of individuals that represent the range of cognitive statuses and age cohorts from ADNI.
Our implementation of BLAzER for research studies utilizes a high-throughput, supercomputing cluster for segmentation (by FS) along with a commercially available multimodality image viewing and analysis software program (MIM) with customized workflows to use and process the segmentation data. This workflow can be adapted to other segmentation algorithms by customizing the visualization software workflow with minimal change in the user experience. This workflow allows the user to register and process brain segmentation maps with brain PET and brain MRI in a single viewing environment for quality control, manual correction if needed, data extraction, and real-time visual assessment. Utilizing parallel computing strategies with FS allows segmentation of over 100 studies in approximately 12 hours which is suitable for use with clinical trials and large imaging datasets. We chose MIM due to its efficiency and relative ease of use compared to existing tools that are either dependent on relatively heavy computational skills [16, 17] or require significant image processing input [2, 15]. MIM’s GUI-interface is most similar to Porcupine [40], enabling analysis of large, multi-modal image datasets by users without a sophisticated computational skillset. Similar to other existing workflows [2, 47-49], BLAzER allows for automatic delineation of PET ROIs. However, an important advantage of BLAzER is that it can accept brain segmentation data sets from multiple sources, including FDA-cleared algorithms such as NR. This potential integration of a completely FDA-cleared workflow helps overcome barriers between clinical and research applications of BLAzER. While the present work focuses on the use of this workflow for amyloid- and tau-PET, it is just as applicable for the analysis of other brain PET imaging studies. This flexibility combined with visualization software familiar to physicians and technologists facilitates implementation of quantitative PET neuroimaging in routine clinical practice with a wide range of segmentation algorithms and PET tracers. As BLAzER can accommodate any DICOM dataset, other neuroimaging modalities, such as SPECT for Parkinson’s disease or functional MRI and diffusion tensor imaging for schizophrenia, could be processed along with PET neuroimaging biomarkers of neuroinflammation [50-52] by adapting the BLAzER workflow. Although not used in this data study, the BLAzER workflow can be used for analysis of dynamic PET data sets.
Another major strength of the BLAzER workflow is the ability to use any segmentation map in DICOM format for defining PET ROIs and extracting PET measurements. However, in contrast to our inter-workflow comparison between BLAzER-FS and ADNI, our intra-workflow comparison between BLAzER-FS* and BLAzER-NR demonstrated subtle differences between the segmentation methods, yet still showed excellent correlations for both global and regional comparisons. Our results showed that NR provides slightly higher SUVRs when compared to FS across large cortical structures, where dichotomous classification revealed 7 discrepant individuals out of 127 total cases even after regression adjustment. These findings, which contrast with the nearly identical dichotomization between BLAzER-FS and ADNI comparison, emphasize the effect that differences in segmentation can have on amyloid dichotomization. Despite our best efforts to match BLAzER-NR’s cortical anatomical boundaries with rearrangement of BLAzER-FS* regions, BLAzER-NR and BLAzER-FS* have slightly different anatomical definitions of brain subregions across cortical structures not only from each other but also from ADNI (unlike BLAzER-FS), which noticeably affects correlation and dichotomization. Additionally, the methods utilize different segmentation algorithms in separating gray from white matter, leading to slightly thicker cortical ROIs with NR than with FS. As florbetapir binds non-specifically to white matter even in cognitively normal individuals [53], we believe that a greater contribution of PET signal from white matter in BLAzER-NR contributes to the systematically larger SUVRs when compared to BLAzER-FS*. Users can choose segmentation software that best fits their clinical and/or research needs. Based on our observed results, the selection of the appropriate segmentation input depends on the application. For rapid analysis of large major cortical or subcortical structures in a clinical setting, NR and similar algorithms might be preferred. In contrast, FS provides more versatility for brain segmentation due its ability to provide smaller ROIs compromised of multiple subregions which may be more appropriate for research and analysis of PET tracers with spatially-restricted distributions.
In conclusion, BLAzER is a streamlined image processing workflow for efficient registration, visualization, and extraction of brain PET data. We successfully validated the accuracy and reproducibility of BLAzER using ADNI as the reference standard for amyloid-PET and tau-PET analyses. We also showed how two different segmentation inputs, NR and FS, can be used with BLAzER and that the cortical ROIs extracted by BLAzER remain consistent with the original MR volumes. This versatile workflow can reduce barriers to quantitative brain PET and MR analysis for a wide range of research and clinical applications.
Availability of data and materials
Data used in the preparation of this article were obtained from the ADNI database (adni.loni.usc.edu), which is easily available for download from the Laboratory of Neuroimaging (LONI) website to the research public.
Funding Sources
This work was supported by National Institutes of Health grants RF1AG059009 and T32GM008361, the Alzheimer’s Drug Discovery Foundation, and the Department of Radiology at the University of Alabama at Birmingham. Data collection and sharing for this project was funded by the Alzheimer’s Disease Neuroimaging Initiative (ADNI) (NIH U01AG024904) and DOD ADNI (Department of Defense award number W81XWH-12-2-0012). ADNI is funded by the National Institute on Aging, the National Institute of Biomedical Imaging and Bioengineering, and through generous contributions from the following: AbbVie, Alzheimer’s Association; Alzheimer’s Drug Discovery Foundation; Araclon Biotech; BioClinica, Inc.; Biogen; Bristol-Myers Squibb Company; CereSpir, Inc.; Cogstate; Eisai Inc.; Elan Pharmaceuticals, Inc.; Eli Lilly and Company; EuroImmun; F. Hoffmann-La Roche Ltd and its affiliated company Genentech, Inc.; Fujirebio; GE Healthcare; IXICO Ltd.; Janssen Alzheimer Immunotherapy Research & Development, LLC.; Johnson & Johnson Pharmaceutical Research & Development LLC.; Lumosity; Lundbeck; Merck & Co., Inc.; Meso Scale Diagnostics, LLC.; NeuroRx Research; Neurotrack Technologies; Novartis Pharmaceuticals Corporation; Pfizer Inc.; Piramal Imaging; Servier; Takeda Pharmaceutical Company; and Transition Therapeutics. The Canadian Institutes of Health Research is providing funds to support ADNI clinical sites in Canada. Private sector contributions are facilitated by the Foundation for the National Institutes of Health (www.fnih.org). The grantee organization is the Northern California Institute for Research and Education, and the study is coordinated by the Alzheimer’s Therapeutic Research Institute at the University of Southern California. ADNI data are disseminated by the Laboratory for Neuro Imaging at the University of Southern California.
Conflicts of Interests
Jonathan McConathy has declared a relationship with Eli Lilly and Avid to whom he provides consulting and from which he receives research support. The other authors have declared no conflicts of interest, financial or otherwise.
Author Contributions
FR, EDR, and JM designed the research studies. FR and SG performed image analysis. FR, SG, CM, RK, SL, EDR, and JM interpreted data. CM, RK, SL, EDR, and JM provided methodology expertise. ADNI provided imaging studies. FR, EDR, and JM wrote the manuscript. All authors reviewed, edited, and approved the final version of the manuscript.
ACKNOWLEDGMENTS
Footnotes
↵** NOTES Data used in preparation of this article were obtained from the Alzheimer’s Disease Neuroimaging Initiative (ADNI) database (adni.loni.usc.edu). As such, the investigators within the ADNI contributed to the design and implementation of ADNI and/or provided data but did not participate in analysis or writing of this report. A complete listing of ADNI investigators can be found at http://adni.loni.usc.edu/wp-content/themes/freshnews-devv2/documents/policy/ADNI_Acknowledgement_List%205-29-18.pdf
References
- [1].↵
- [2].↵
- [3].
- [4].↵
- [5].↵
- [6].↵
- [7].↵
- [8].↵
- [9].↵
- [10].↵
- [11].
- [12].
- [13].↵
- [14].↵
- [15].↵
- [16].↵
- [17].↵
- [18].↵
- [19].↵
- [20].↵
- [21].↵
- [22].↵
- [23].↵
- [24].↵
- [25].↵
- [26].
- [27].
- [28].↵
- [29].↵
- [30].↵
- [31].↵
- [32].↵
- [33].
- [34].↵
- [35].↵
- [36].↵
- [37].↵
- [38].↵
- [39].↵
- [40].↵
- [41].↵
- [42].↵
- [43].↵
- [44].↵
- [45].↵
- [46].↵
- [47].↵
- [48].
- [49].↵
- [50].↵
- [51].
- [52].↵
- [53].↵