ABSTRACT
2-photon fluorescence microscopy has been used extensively to probe the structure and functions of cells in living biological tissue. 2-photon excitation generates fluorescence from the focal plane, but also from outside the focal plane, with out-of-focus fluorescence increasing as the focus is pushed deeper into tissue. It has been suggested that the 2-photon depth limit, beyond which results become inaccurate, is where in- and out-of-focus fluorescence are equal. We found the depth limit of 2-photon excitation in mice with GCaMP6 indicator expression in all layers of visual cortex, by comparing near-simultaneous 2- and 3-photon excitation. 2-photon results were accurate only superficial to 450 μm, matching the depth at which in-and out-of-focus fluorescence were equal. The expected depth limit is deeper in tissue with fewer fluorophores outside the plane of interest. Our results, from tissue with a largely homogenous distribution of fluorophores, establish a superficial bound on the 2-photon depth limit in the mouse visual cortex.
INTRODUCTION
2-photon excitation permits fluorescence imaging with cellular and subcellular resolution hundreds of micrometers into biological tissue. Generally, the maximal imaging depth (depth limit) of 2-photon excitation is determined by fluorescence from outside the focal plane. As the focal plane is pushed deeper into scattering tissue, illumination intensity at the tissue surface must be increased to maintain intensity in the focal plane, resulting in an increase in out-of-focus fluorescence with increasing depth (Ying et al., 1999; Theer et al., 2003). In a seminal study, Theer and Denk (2006) explored 2-photon excitation analytically and defined the fundamental imaging depth limit by calculating the depth at which the detected fluorescence generated by ballistic and scattered excitation light outside the focal plane equals that from fluorophores excited in the ballistic focus. The ratio of in- and out-of-focus fluorescence is a complex function of numerous factors, including numerical aperture, laser pulse duration, scattering anisotropy, and fluorophore distribution, but the calculations of Theer and Denk (2006) suggest that the depth limit is at ~3 scattering length constants under typical imaging conditions. In cortical grey matter, 3 scattering length constants corresponds to ~600 μm below the tissue surface.
3-photon excitation permits deeper imaging than 2-photon excitation, in part because 3-photon excitation generates fluorescence almost exclusively from the focal plane (Horton et al., 2013; Kobat et al., 2009; Kobat et al., 2011; Ouzounov et al., 2017; Yildirim et al., 2019). In the absence of out-of-focus fluorescence one expects the functional properties of neurons measured with 2- and 3-photon excitation to be identical, but this expectation has not been tested directly and the impact of out-offocus fluorescence has not been measured. 3-photon excitation offers the opportunity to estimate in- and out-of-focus fluorescence and thereby test the predictions of earlier analyses. We implemented near-simultaneous 2- and 3-photon excitation to compare results 200 to 650 μm below the surface of the brain in transgenic mice with dense GCaMP6 expression throughout neocortex. Our results show that 2- and 3-photon excitation produce equivalent results in superficial layers but not in deep in cortex, and indicate that the depth limit of 2-photon excitation corresponds to the plane where in- and out-offocus fluorescence are equal, consistent with Theer and Denk (2006).
RESULTS
As an illumination source for 3-photon excitation, we used a 40 Watt Coherent Monaco laser source and Opera-F optical parametric amplifier, providing 2 μJ, 50 fs pulses at 1 MHz. We configured a MIMMS 2-photon microscope for 3-photon excitation, exchanging the scan and tube lens to increase transmission through the microscope at 1300 nm and added a compressor to compensate for pulse dispersion between the laser source and sample. Through a cranial window over visual cortex, we were routinely able to image neurons >1 mm below the pial surface of cortex in GCaMP6 mice (supplementary figure 1). Fluorescence intensity followed a cubic relationship with illumination intensity, consistent with fluorescence being driven by the absorption of 3 photons.
In mice expressing GCaMP broadly in cortical pyramidal neurons, loss of contrast was noticeable in 2-photon images from hundreds of micrometers below the brain surface (figure 1A) where contrast was preserved by 3-photon excitation (figure 1B). To compare 2- and 3-photon excitation more directly, we implemented near-simultaneous 2- and 3-photon excitation. We used 2 laser sources, combining the beams immediately before the scanning galvanometers (figure 2A). With a fast Pockels cell on each laser line acting as a shutter, we alternated 2- and 3-photon excitation, line-by-line (figure 2B). The line duration was 0.5 ms, resulting in 0.5 ms separation of 2- and 3-photon images.
In superficial cortex, 2- and 3-photon results were similar. The same neurons were visible in near-simultaneous 2- and 3-photon images and changes in fluorescence were coincident in 2- and 3-photon image pairs (supplementary movie 1); the results of motion correction and segmentation on 2- and 3- photon movies were similar (standard deviations of motion correction distributions <2 μm at <350 μm, figure 3C); there were 50-90 neurons identified in each image (figure 3D); >80% of neurons in 3-photon images matched a neuron in the corresponding 2-photon image (figure 3E); and traces extracted from matching neurons in 2- and 3-photon movies were strongly correlated, with Pearson correlation coefficients of ~0.8-0.9 (figure 3F), consistent with a previous study (Wang et al., 2017).
The similarity of 2- and 3-photon results declined with depth. In 3-photon images, image contrast, motion correction, and number of neurons changed little with depth. In 2-photon images, contrast declined incrementally with depth, to near zero at 650 μm (figure 3B). Lateral motion correction from 2-photon movies increased with depth: the standard deviation of motion correction was <3 μm at <400 μm; at 650 μm, the standard deviation of lateral motion correction was <3 μm for 3-photon excitation and ~25 μm for 2-photon excitation (figure 3C, supplementary figure 1). The segmentation routine identified few neurons in deep locations (figure 3D), and the overlap between matching neurons in 2- and 3-photon images and the correlation coefficient between the resulting traces both declined at >350-400 μm (figure 3E and F).
To determine how the decline in image quality with depth affects the functional properties of cortical neurons measured with 2-photon excitation, we examined the apparent responses of cortical neurons to visual stimuli. We presented sinusoidal gratings drifting in 12 directions, and calculated the direction preference of each neuron from extracted fluorescence traces, comparing results from 2- and 3-photon excitation. For superficial neurons, visually-evoked changes in 2- and 3-photon fluorescence were almost identical, trial-by-trial (figure 3G) and the resulting preferred direction of each neuron was closely matched (figure 3H), with 83% (305 of 368) of neurons ≤350 μm from the brain surface exhibiting identical preferred directions with 2- and 3-photon excitation. Visually-evoked changes in 2-photon fluorescence were suppressed in deeper neurons (figure 3G, supplementary figure 3) and the percentage of neurons with matching 2- and 3-photon direction preference declined (figure 3H & J, supplementary figure 4). At 600 μm, the number of neurons with matching preference was above chance (1/12 = 8.3%), but <<50 %.
2-and 3-photon excitation produce equivalent results from superficial depths, but the results become less similar >400 μm below the brain surface. Increasing out-of-focus fluorescence and the resulting decline in image contrast are the likely cause. From the ratio of contrast in 2- and 3-photon images, we estimated the percentage of fluorescence that originated from the focal plane during 2-photon excitation. As expected, the percentage of 2-photon fluorescence originating from the focal plane decreased with increasing depth (figure 4A, supplementary figure 6). In- and out-of-focus fluorescence were equal at ~400-450 μm, the depth beyond which the results of 2-photon excitation are inaccurate. Hence our results support the depth limit corresponding to the depth at which in- and out-of-focus fluorescence are equal.
We compared our measurements of in- and out-of-focus fluorescence with predictions from theoretical modeling of focused light propagation in scattering tissue (Ying, et al. 1999; Theer & Denk, 2006;). According to this model, 50% in-focus fluorescence occurs ~3 scattering lengths below the brain surface, ~450 μm for a scattering length of 150 μm and 600-700 μm for a scattering length of 200 μm (figure 4A, black and grey lines, respectively). 3-photon excitation is almost free of out-of-focus fluorescence at these depths.
2-photon excitation will support imaging >450 μm below the brain surface if there are few fluorophore molecules outside the focal plane. Unfortunately, out-of-focus fluorescence arises from fluorophores throughout the tissue above and, to a lesser extent, below the focal plane (figure 4B; Theer & Denk, 2006). 3-photon excitation generates little out-of-focus fluorescence, but again most arises from locations immediately superficial to the focal plane. Hence a substantial reduction in out-of-focus fluorescence, and increase in depth limit, would likely occur only in tissues with few fluorophore molecules throughout the entire depth of tissue above and below the focal plane.
DISCUSSION
We compared the results of 2- and 3-photon excitation of GCaMP6s in excitatory neurons in mouse visual cortex. Results from superficial cortex were similar, an expected result that confirms that the cellular signals reported by GCaMP6s are independent of the mechanism of excitation, and that 3P imaging has not been compromised by saturation or phototoxic effects (Yildirim, et al. 2019). With increasing depth from ~250-650 μm, 2-photon image contrast declined and 3-photon image contrast was preserved. Many measures (estimated motion, number of neurons segmented, matching of segmented neurons, correlation traces, similarity of fluorescence changes, similarity of preferred direction) were robust to changes in 2-photon image contrast to ~400 μm, but deteriorated between 400 and 550 μm on average, some abruptly, compromising measurement of fluorescence changes and direction tuning.
In our experiments, we used a mouse line with GCaMP6s expression in excitatory neurons through all layers of cortex. From the perspective of out-of-focus fluorescence, we expect these mice to be a worst-case scenario for 2-photon excitation. In these mice, our results place the depth limit at ~450 μm below the brain surface, shallower than the depth predicted Theer and Denk (2006) and by our calculations. We expect aberrations to reduce the depth at which in- and out-of-focus fluorescence are equal. Aberrations are present in any imaging system, but not included in our calculations or those of Theer & Denk (2006). Slight compression of cortex is common in cranial window preparations (e.g. de Vries et al., 2018) and might further reduce the depth limit by reducing the scattering length of cortical tissue. Hence one expects the depth limit of 2-photon excitation to be shallower than suggested by calculations. Our measurements indicate that the depth limit can be as shallow as 2-3 scattering lengths or ~450 μm.
Our results drive two predictions that we have not tested directly. Firstly, we expect that 2-photon excitation will be adequate for characterization of functional properties such as direction tuning in neurons ≤450 μm from the brain surface in nearly all GCaMP6s mouse lines. Secondly, we expect 2- and 3-photon results to be comparable at >450 μm in many preparations. We observed substantial mouse-to-mouse variability at 500-650 μm, suggesting that 2-photon excitation might be a viable tool to >450 μm in a small subset of our mice. In other mouse lines and tissues, 2-photon excitation at >450 μm will provide more accurate functional measurements in preparations with less out-of-focus fluorescence, including tissues with sparser expression of GCaMP6s and tissues labeled with indicators with low resting fluorescence, such as jGCaMP7c (Dana et al., 2018). In such tissues, there are several factors that might limit 2-photon excitation. Out-of-focus fluorescence, though reduced, will still occur and may equal in-focus fluorescence at a location deeper than 450 μm. Aberrations might prove limiting, enabling adaptive optics to extend the depth limit (Ji et al., 2010; Ji et al., 2012). A third possibility is maintenance of image quality to a depth at which the thermal limit of brain tissue is met (Podgorski & Ranganathan, 2016).
In summary, we have established that 2- and 3-photon excitation are equivalent ≤~450 μm below the brain surface in mice with GCaMP6s expression throughout cortical layers. Tentatively, we suggest the depth limit of 2-photon excitation is 450 μm or deeper in nearly all mouse lines, since few if any mice express a higher proportion of fluorophore molecules outside the focal plane than mice with expression throughout the cortical layers. In tissues with and tissues without extensive fluorophore expression outside the focal plane, 3-photon excitation enables measurement of cellular activity beyond the depth limit of 2-photon excitation.
METHODS
Basic 3-photon microscope
Our 3-photon microscope was built around a Coherent Monaco / Opera-F laser source (≤2 nJ, 50 fs pulses at 1 MHz; Coherent Inc., Santa Clara) and a modified MIMMS microscope manufactured by Sutter Instrument (Novato, CA). We replaced the scan and tube lenses (respectively, Thorlabs SL50-3P and a Plössl pair of achromatic doublets, Thorlabs AC254-400-C) to improve transmission at 1300 nm. The primary dichroic mirror was FF735-DI02 (Semrock, Rochester NY). We used an Olympus 25x/1.05 objective (75% transmission at 1300 nm) or Nikon 16×/0.8 objective (50% transmission at 1300 nm) and image acquisition was controlled by ScanImage (Vidrio Technologies LLC) with acquisition gating for low rep rate lasers.
We estimated group delay dispersion (GDD) through the microscope at ~15,000 fs2, approximately half of which was attributable to the Pockels cell (360-40-03-LTA, Conoptics Inc, Danbury CT). To compensate, we built a 4-pass pulse compressor using a single SF-11 glass prism (Thorlabs PS-853) and a two hollow roof prism mirrors (Thorlabs HRS1015-P01 and HR1015-P01). Compression was tuned by maximizing brightness with a fluorescein sample.
400-500 mW of 1300 nm illumination was available after the objective, corresponding to transmission from laser source to sample of ~20%. The maximum field of view of 3-photon excitation was 360 x 360 μm. Images were acquired with dual linear galvanometers at a frame rate of ~8 Hz.
Illumination intensity
Photodamage is often a concern in light microscopy. Photodamage can result from linear processes, principally heating (resulting from the absorption of infrared light by water in brain tissue) and from non-linear processes. Non-linear processes are of particular concern with high-energy pulsed sources such as those used for 2- and 3-photon fluorescence microscopy. Heating-related photodamage often occurs with >250 mW of prolonged illumination at 800-1040 nm (Podgorski & Ranganathan, 2016) and the molar extinction coefficient of water at 1300 nm is ~2x that at 900 nm (Curcio & Petty, 1951; Hale & Querry, 1973; Bertie & Lan, 1996), suggesting that heating-related tissue damage may occur at >~100-150 mW of prolonged illumination at 1300 nm. To avoid damage, we used illumination intensities <100 mW. Typically, we could image through the depth of neocortex using <30 mW illumination while maintaining a signal-to-noise ratio comparable to typical 2-photon experiments. We rarely observed signs of photodamage, even in mice subjected to 2 hours of continuous 3-photon imaging per day for 5 days.
Near-simultaneous 2- and 3-photon excitation
For 2-photon excitation, we used a Coherent Chameleon Ultra II laser source at 920 nm. For near-simultaneous 2- and 3-photon excitation, we used a Nikon 16×/0.8 objective (50% transmission at 1300 nm). Time-averaged power available after the objective was 200-250 mW at 1300 nm. To match the focal planes of 2- and 3-photon excitation, to the 2-photon path we added an electrically-tunable lens (EL-10-30-TC, Optotune, Dietikon, Switzerland).
Mice and surgeries
We used Cre-lox transgenic mice to drive GCaMP6s expression in excitatory neurons throughout cortical layers and areas, crossing Emx1-IRES-Cre (B6.129S2-Emxltml(cre)Krj/J, JAX stock number 005628; Gorski et al., 2002) or Slc17a7-IRS2-Cre (B6;129S-Slc17a7tm11(cre)Hze/J, JAX stock number 023527; Harris et al., 2014) and Ai162(TIT2L-GCaMP6s-ICL-tTA2 reporter mice (JAX stock number 031562, Daigle et al., 2018).
A chronic cranial window was implanted over visual cortex as described previously (Goldey et al., 2014; de Vries et al., 2018). Briefly, under 0.5-2% isoflurane anesthesia, a head restraint bar was attached to the skull using C & B Metabond (Parkell) and a 5 mm diameter craniotomy was opened over the left visual cortex at coordinates 2.7 mm lateral, 1.3 mm anterior to lambda. A durotomy was performed and the craniotomy was sealed with a stack of three #1 coverslips, attached to each other using optical adhesive, and attached to the skull with Metabond.
Visual stimuli
Visual stimuli were full-field sinusoidal gratings of 6 orientations, each drifting perpendicular to its orientation (12 directions), at spatial frequencies of 0.04 and 0.08 cycles per degree and a temporal frequency of 1 Hz. Each grating was presented 8 times in random order, each for 2 seconds with 1 second of grey screen between presentations. 0 degrees corresponds to a grating drifting horizontally in the nasal-to-temporal direction and 90 degrees to a downward-drifting grating. The visual stimulus display and its calibration were as described previously (de Vries et al., 2018). Briefly, stimuli were displayed on an LCD monitor, 15 cm from the right eye, gamma-corrected and of mean luminance of 50 cd/m2. Spherical warping was employed to ensure the apparent size, speed, and spatial frequency were constant across the monitor.
Image analysis
Image analysis was performed using custom routines in Python 3. For comparison of 2- and 3-photon excitation, images were first separated into 2- and 3-photon movies. Dark current, the mean of several images acquired with no laser illumination, was measured in each movie and subtracted. Image brightness (figure 4A) was measured in digitizer units. To avoid artifacts, each movie was normalized to the same mean brightness.
Image contrast (figure 4B) was expressed on a scale from 0 (no contrast) to 1. Contrast was calculated locally (in 22.5 x 22.5 pixel blocks) from the temporal mean projection of a movie, the final value being the mean of all the blocks. Contrast in each block was defined as 1 – minimum brightness / maximum brightness.
Each movie was motion-corrected and putative neuronal somata identified by segmentation. Soma and neuropil fluorescence traces were extracted and neuropil fluorescence was subtracted from the corresponding soma trace (r = 1). Motion correction, segmentation and trace extraction were performed using Suite2p (Pachitariu et al., 2017) with default settings except for maxregshift which was set to 0.2 to permit ≤~70 μm motion correction in each transverse axis. Motion correction (figure 4C) was the mean of x- and y-corrections applied by Suite2p. Neuron count (figure 4D) was the number of putative somata returned by Suite2p, with manual editing to assist the sorting of somatic from non-somatic regions of interest. % match (figure 4E) was the percentage of putative neurons segmented in the 3- photon image that were also segmented in the corresponding 2-photon image, assessed manually by comparing images of segmented regions. Pearson correlation coefficient (figure 4F) was calculated from the neuropil-subtracted fluorescence traces using scipy.stats.pearsonr. To ensure that the correlation coefficient calculation was from matching regions of interest, traces were extracted from 2-photon and 3-photon movies using regions of interest segmented from 3-photon movies.
To compare 2- and 3-photon measurements of responses to drifting gratings, we used two measures: mean fluorescence response and preferred direction. Again, these measures were applied to traces extracted from 2- and 3-photon movies using regions of interest segmented from 3-photon movies. For each measure, we first calculated the mean response of each neuron (from 8 presentations). For the mean fluorescence response, we plot 2- vs 3-photon amplitudes of the mean change in fluorescence for each grating. Hence in the mean fluorescence response plots (figure 5A), each neuron is represented by 24 data points (12 directions x 2 temporal frequencies). The direction preference plots (figure 5B), report the grating direction that evoked the largest change in fluorescence and each neuron is therefore represented by a single data point.
Modeling in- and out-of-focus fluorescence
To estimate the out-of-focus fluorescence generated by excitation light focusing through a homogeneous volume of fluorescent and scattering tissue, we modeled the intensity of ballistic and scattered light, Ib (z, ρ) and Is (z, ρ) respectively, in a plane transverse to the optical axis defined by the polar radius, ρ, and depth z below the surface of the brain. We calculated the out-of-focus, 2-photon-excited fluorescence (Foof) numerically, following Theer & Denk (2006). where, V is the out-of-focus illuminated volume of tissue, C2p is a modality-specific scaling factor incorporating contributions from fluorophore concentration and excitation efficiency, and assumed to be constant over the volume.
We neglected possible depth dependence of fluorescence collection and detection, non-conservative attenuation due to bulk absorption of near-IR light, and the time dependence of excitation by ultrashort pulses that becomes a significant factor for pulse widths < ~50 fs (Theer & Denk, 2006; but see also Leray et al., 2007).
Previous models (Xu & Webb, 1996; Theer & Denk, 2006) neglected the difference in distances traveled through tissue by on-axis and marginal rays. The difference in distance can be substantial for high-numerical aperture objectives, but of marginal importance when the focal plane is many multiples of the scattering length below the tissue surface. Here, we calculated fluorescence with the focal plane 1-4 scattering lengths below the tissue surface and therefore account for the dependence on propagation angle relative to the optical axis by incorporating a radially varying propagation distance, where z0 is the focal plane depth. This factor modifies the intensity profiles of Ib and of Is.
Foof can be decomposed into individual contributions from ballistic, scattered, and cross-term interaction excitation, for 2-photon excitation: where, Voof is the out-of-focus volume denoting the range δ is the exclusion depth of in-focus light around z0.
In our calculations, δ was a fifth of the scattering length, or 40 μm, which we assume to be larger than the depth of focus and therefore underestimates the magnitude of the background; wavelength was 900 nm; numerical aperture 0.8; and anisotropy factor 0.9.
To calculate ballistic and scattered light intensities, we considered a Gaussian beam propagating from the surface (z = 0) of a scattering medium of scattering length to a ballistic focus located at z = z0. We introduced a direction dependent propagation length , and calculate the ballistic intensity profile at depth z and radial distance ω according to where, is the width is the Rayleigh length determined by the NA-derived focusing half-angle.
As in Theer & Denk (2006), we calculated the intensity distribution of scattered light from a beam spread function derived for small-angle scattering (McLean, Freeman & Walker, 1998). We integrated over temporal and angular coordinates to obtain the normalized spatial distribution function
The spreading parameter is derived from the anisotropy factor g and the function h(z,p) accounts for the diffusive spreading of scattered light with increasing depth from an initial on-axis ray, with total power increasing with depth according to 1 – exp [-az], modeling the transfer ofenergy from the ballistic to the scattered field.
Integrating over the initial surface distribution, the intensity distribution of scattered light at depth z is where .
is the propagation distance from the surface
is the Gaussian beam width at the surface.
Proportion of fluorescence originating from the focal plane
Calculation of the ratio of in- and out-of-focus fluorescence was based on image contrast. We subdivided the 256×512 pixel images of the motion-corrected, time-averaged 2-photon and 3-photon movies into 32×32 pixel subregions. Within each subregion, we determined the minimum pixel value and pixel value mean, and respectively, where denotes the time-averaged fluorescence in each pixel with the minimum and mean functions over the 32×32 = 1024 pixels. For each subregion in each imaging modality (2P and 3P), we then calculated a contrast parameter, , for the j-th subregion in the k = {2,3} (2P,3P) modality.
To calculate in- and out-of-focus fluorescence, we made three assumptions. Firstly, we assumed the time-averaged fluorescence in each pixel reflects the sum of the in-focus and out-of-focus fluorescence . Secondly, we assumed 3-photon excitation generates no out-of-focus fluorescence so that for 3-photon excitation. Thirdly, we assumed in-focus fluorescence is proportional to a modality-independent concentration factor, C, with a modality-dependent proportionality constant, so that Fik = ρkC.
As a measure of the percentage of fluorescence that originates from the focal plane, we calculated an empirical contrast ratio (ECR):
The ECR calculated in each subregion was averaged over the subregions to determine the time-averaged ECR for a given imaging depth.
We calculated the theoretical contrast ratio via a signal-to-background ratio calculation. We modeled the total in-focus fluorescence, Ft, according to where Pz0 is the total, scattering attenuated, ballistic power estimated at the focal plane according to . The signal-to-background ratio was defined as the ratio of total in-focus to out-of-focus fluorescence, given by which ranges from 0 at very large depths to ∞ in the background-free case. We defined the contrast ratio, to range from 0 to 1.