PT - JOURNAL ARTICLE AU - Mikhail Repin AU - Sergey Pampou AU - David J. Brenner AU - Guy Garty TI - The use of a centrifuge-free RABiT-II system for high-throughput micronucleus analysis AID - 10.1101/627596 DP - 2019 Jan 01 TA - bioRxiv PG - 627596 4099 - http://biorxiv.org/content/early/2019/05/04/627596.short 4100 - http://biorxiv.org/content/early/2019/05/04/627596.full AB - The cytokinesis-block micronucleus (CBMN) assay is considered as the most suitable biodosimetry method for automation. Previously, we automated this assay on a commercial robotic biotech high-throughput system (RABiT-II) adopting both a traditional and an accelerated micronucleus protocol, both using centrifugation steps for lymphocyte harvesting and washing, after whole blood culturing. Here we describe further development of our accelerated CBMN assay protocol for using on High Throughput/High Content Screening (HTS/HCS) robotic systems without a centrifuge. This opens the way for implementation of the CBMN assay on a wider range of commercial automated HTS/HCS systems and thus increases the potential capacity of dose estimates following a mass-casualty radiological event.