RF-induced heating in tissue near bilateral DBS implants during MRI at 1.5 T and 3T: The role of surgical lead management

Neuroimage. 2019 Jan 1:184:566-576. doi: 10.1016/j.neuroimage.2018.09.034. Epub 2018 Sep 19.

Abstract

Access to MRI is limited for patients with deep brain stimulation (DBS) implants due to safety hazards, including radiofrequency (RF) heating of tissue surrounding the leads. Computational models provide an exquisite tool to explore the multi-variate problem of RF heating and help better understand the interaction of electromagnetic fields and biological tissues. This paper presents a computational approach to assess RF-induced heating, in terms of specific absorption rate (SAR) in the tissue, around the tip of bilateral DBS leads during MRI at 64MHz/1.5 T and 127 MHz/3T. Patient-specific realistic lead models were constructed from post-operative CT images of nine patients operated for sub-thalamic nucleus DBS. Finite element method was applied to calculate the SAR at the tip of left and right DBS contact electrodes. Both transmit head coils and transmit body coils were analyzed. We found a substantial difference between the SAR and temperature rise at the tip of right and left DBS leads, with the lead contralateral to the implanted pulse generator (IPG) exhibiting up to 7 times higher SAR in simulations, and up to 10 times higher temperature rise during measurements. The orientation of incident electric field with respect to lead trajectories was explored and a metric to predict local SAR amplification was introduced. Modification of the lead trajectory was shown to substantially reduce the heating in phantom experiments using both conductive wires and commercially available DBS leads. Finally, the surgical feasibility of implementing the modified trajectories was demonstrated in a patient operated for bilateral DBS.

Keywords: Computational modeling and simulations; Deep brain stimulation (DBS); Finite element method (FEM); MRI safety; Magnetic resonance imaging (MRI); Medical implants; Neuromodulation; Neurostimulation; Specific absorption rate (SAR).

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Computer Simulation
  • Deep Brain Stimulation* / instrumentation
  • Electrodes, Implanted*
  • Electromagnetic Fields
  • Hot Temperature*
  • Humans
  • Magnetic Resonance Imaging / adverse effects*
  • Magnetic Resonance Imaging / methods
  • Models, Theoretical*
  • Phantoms, Imaging
  • Radio Waves